Trilab Pty Ltd

Site

Calibration Laboratory

Accreditation No.

9926

Site No.

18871

Date of Accreditation

22 Feb 2010

Address

346A Bilsen Road
Geebung, QLD 4034
Australia

trilab.com.au

Contact

Mr Tarin Glazier P: +61 (07) 32655656

Availability

Services available to external clients

Scope of Accreditation

Calibration Laboratory

ISO/IEC 17025 (2017)

Calibration

  • The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
Service Product Determinant Technique Procedure Limitations
Dimensional metrology - Engineering equipment and precision instruments Dial gauges Length measurements Direct measurement against a reference standard

For compliance with AS 2103-1978

Capability

with Calibration and Measurement Capability of -
1.7 μm for linear measurements from 0.1 mm to 25 mm

3.6 μm above 25  mm to 100 mm


Extensometers Length measurements Direct measurement against a reference standard

For compliance with AS 1545

Capability

with Calibration and Measurement Capability of -
(1 + 0.5L) μm where L is the extension in mm from 0.05 mm up to 25 mm

Test sieves Length measurements Direct measurement against a reference standard AS 1152(1993), ISO 3310-1, ISO 3310-2 and ASTM E11

Wire woven and perforated plate sieves with a minimum aperture 32 μm to
AS 1152(1993), ISO 3310-1, ISO 3310-2 and ASTM E11


Capability

with Calibration and Measurement Capability of -
3 μm from 0.032 mm to 16 mm

Vernier calipers Length measurements Direct measurement against a reference standard

For compliance with JIS B 7507

Capability

with Calibration and Measurement Capability of -
17 μm for linear measurements from 25 mm up to 300 mm
Force metrology - Force measuring and testing equipment Falling weight deflectometer Load cell and sensor gains (deflection) Comparison with a reference standard TP-BF-StB-Part-B 8.4

Capability

With Calibration and Measurement Capability of -
0.16% of reading for force from 1 kN to 16 kN
 0.008 mm for deflection from -2.0 mm to +2.0 mm
Force metrology - Force measuring and testing equipmentincluding on-site calibrations Compression and universal machines in compression; Hydraulic rams and jacks; Load cells; Load measuring rings; Tension and universal machines in tension Force in compression; Force in tension Comparison measurement with reference load cell; Gravimetric measurement with reference mass

AS 2193

Capability

including on-site calibrations
with Calibration and Measurement Capability of -

0.15% of reading for Class AA 0.01kN to 100 kN in Tension

0.25% of reading for Class A 0.01 kN to 100 kN in Tension

0.16% of reading for Class AA 0.01 kN to 1.5 MN in Compression
0.25% of reading for Class A 0.01 kN to 3.0 MN in Compression

Force metrology - Hardness standards and equipment Rubber bases Rubber hardness Direct force and length measurement

AS 1289.3.1.1

Capability

with Calibration and Measurement Capability of -

1 IRHD for rubber hardness

Mass - Determination of mass and calibration of weighing devices Industrial weighing devices Mass Gravimetric measurement against reference mass

Capability

including on site calibrations
with Calibration and Measurement Capability of -
1.5 mg above 300 g up to and including 1 kg
6 mg above 1 kg up to and including 5 kg
25 mg above 5 kg up to and including 20 kg
35 mg above 20 kg up to and including 30 kg
150 mg above 30 kg up to and including 60 kg
Precision laboratory balances Mass Gravimetric measurement against reference mass

Capability

including on site calibrations
with Calibration and Measurement Capability of -
40 μg above 1 g up to and including 20 g
100 μg above 20 g up to and including 100 g
200 μg above 100 g up to and including 200 g
300 μg above 200 g up to and including 300 g
Pressure metrology - Pressure and vacuum measuring equipmentincluding on site calibrations Pressure gauges; Pressure transducers Gauge pressure By pressure calibrator
MSA Test Method 1

Pressure gauges:

Including on-site calibration to MSA Test Method 1
In the range of 0 to 60 MPa

Pressure transducers:

Including on-site calibration of self-indicating transducers
 In the range of 50 kPa to 60 MPa

Capability

including on-site calibrations

With Calibration and Measurement Capability of-

0.35% or 0.15 kPa (whichever is greater) from 50 kPa to 60 MPa ( Pressure Transducers)
0.35% or 0.15 kPa (whichever is greater) from 0 MPa to 60 MPa ( Pressure Gauges)

The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.

Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.