National Association of Testing Authorities, Australia
Scope of Accreditation
Rencomp Pty Ltd t/a Sci-Tek Systems and Pinnacle CSK
Site
Calibration Service
Accreditation No.
16831
Site No.
22000
Date of Accreditation
09 Jul 2014
Contact
Availability
Services available to external clients
Scope of Accreditation
ISO/IEC 17025 (2017)
Calibration
- The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
Service | Product | Determinant | Technique | Procedure | Limitations | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Temperature metrology - Verification of controlled enclosuresincluding on site calibrations | Medical refrigeration equipment | Spatial uniformity; Temperature; Temporal uniformity | Direct temperature measurement | SOP-Calibration Refrigeration Systems Temperature verification and calibrations as per the methodology in AS3864.2 | Including the following enclosure types: blood fridges (upright, single and multi-door, under bench and cool room) and plasma freezers (ultra-low, upright, chest and freezer room) | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability Including on site calibrations With Calibration and Measurement Capability of- 0.4°C from -80 ºC to 15 ºC Spatial Temperature Distribution1.0 °C from -80 ºC to -40 ºC 1.0 °C above -40 ºC to 15 ºC |
The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.
Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.
Accreditation No.
16831
Site No.
22000
Print date