DC and low frequency electrical metrology - Electrical measurement and test equipmentIncluding onsite calibrations | Ammeters; Digital multimeters (DMM) | DC current | Direct measurement against a reference standard | | |
Capability with Calibration and Measurement Capability of - Including onsite calibrations 0.01% + 0.04 μA from 0 mA to 3.3 mA 0.008% + 0.2 μA from 3.3 mA to 33 mA 0.008% + 2.54 μA from 33 mA to 330 mA 0.025% + 34 μA from 330 mA to 2.2 A 0.05% + 254 μA from 2.2 A to 11 A |
| Digital multimeters (DMM); Ohm meters | Resistance | Direct measurement against a reference standard | | |
Capability with Calibration and Measurement Capability of - Including onsite calibrations 93 μΩ/Ω + 4.7 mΩ up to 11 Ω 93 μΩ/Ω + 7.7 mΩ from 11 Ω to 33 Ω 70 μΩ/Ω + 7.7 mΩ from 33 Ω to 330 Ω 70 μΩ/Ω + 46 mΩ from 330 Ω to 3.3 kΩ 70 μΩ/Ω + 0.46 Ω from 3.3 kΩ to 33 kΩ 85 μΩ/Ω + 4.6 Ω from 33 kΩ to 110 kΩ 93 μΩ/Ω + 4.6 Ω from 110 kΩ to 330 kΩ 0.012% + 43 Ω from 330 kΩ to 3.3 MΩ 0.046% + 0.43 kΩ from 3.3 MΩ to 11 MΩ 0.077% + 0.43 kΩ from 11 MΩ to 33 MΩ 0.39% + 4.3 kΩ from 33 MΩ to 110 MΩ |
| Digital multimeters (DMM); Voltmeters | AC voltage; DC voltage | Direct measurement against a reference standard | | |
Capability with Calibration and Measurement Capability of - Including onsite calibrations D.C. voltmeters 46 μV/V + 2.3 μV up to 330 mV 38 μV/V + 3.9 μV from 330 mV to 3.3 V 38 μV/V + 39 μV from 3.3 V to 33 V 42 μV/V + 0.49 mV from 33 V to 330 V 42 μ V/V + 1.2 mV from 330 V to 900 V A.C. voltmeters From 10 mV up to 33 mV 0.27% + 15 μV from 10 Hz to 45 Hz 0.12% + 15 μV from 45 Hz to 10 kHz 0.15% + 15 μV from 10 kHz to 20 kHz 0.19% + 15 μV from 20 kHz to 50 kHz 0.27% + 25 μV from 50 kHz to 100 kHz 0.77% + 46 μV from 100 kHz to 450 kHz From 33 mV to 330 mV 0.19% + 38 μV from 10 to 45 Hz 0.04% + 15 μV from 45 Hz to 10 kHz 0.08% + 15 μV from 10 kHz to 20 kHz 0.12% + 31 μV from 20 kHz to 50 kHz 0.18% + 131 μV from 50 kHz to 100 kHz 0.54% + 254 μV from 100 kHz to 450 kHz From 0.33 V up to 3.3 V 0.12% + 192 μV from 10 Hz to 45 Hz 0.02% + 46 μV from 45 Hz to 10 kHz 0.06% + 46 μV from 10 kHz to 20 kHz 0.11% + 231 μV from 20 kHz to 50 kHz 0.18% + 1308 μV from 50 kHz to 100 kHz 0.04% + 2538 μV from 100 kHz to 450 kHz From 3.3 V to 33 V 0.12% + 1923 μV from 10 Hz to 45 Hz 0.03% + 462 μV from 45 Hz to 10 kHz 0.06% + 2000 μV from 10 kHz to 20 kHz 0.15% + 3846 μV from 20 kHz to 50 kHz 0.18% + 13077 μV from 50 kHz to 450 kHz From 33 V to 330 V 0.04% + 5 mV from 45 Hz to 1 kHz 0.06% + 12 mV from 1 kHz to 10 kHz 0.07% + 25 mV from 10 kHz to 20 kHz From 330 V to 1020 V 0.04% + 62 mV from 45 Hz to 1 kHz 0.15% + 77 mV from 1 kHz to 5 kHz 0.15% + 385 mV from 5 kHz to 10 kHz |
High frequency electrical metrology - Communications, electromagnetic field strength and EMC test equipmentIncluding onsite calibrations | Amplifiers; Attenuators; Communications systems; Data transmission measuring equipment; Electrical noise and interference measuring equipment; Impedance and reflection measuring equipment; Power measuring equipment; Radio transmission measuring equipment; Spectrum analysers; Waveguide and coaxial components | Reflection coefficient | Comparison with a reference standard | | |
Capability Determination of magnitude of reflection coefficient Including onsite calibrations For One Port reflection coefficients in type 50 ohm coaxial system± with Calibration and Measurement Capability of- (0.001 to 0.4) linear 0.012 in the range 10 MHz to 50 MHz 0.006 in the range 50 MHz to 3 GHz 0.008 in the range 3 GHz to 9 GHz 0.010 in the range 9 GHz to 14 GHz 0.011 in the range 14 GHz to 18 GHz 0.014 in the range 18 GHz to 26.5 GHz 0.016 in the range 26.5 GHz to 40 GHz 0.018 in the range 40 GHz to 48 GHz 0.019 in the range 48 GHz to 50 GHz (0.4 to 1.0) linear 0.017 in the range 10 MHz to 10 GHz 0.020 in the range 10 GHz to 18 GHz 0.022 in the range 18 GHz to 26.5 GHz 0.026 in the range 26.5 GHz to 40 GHz 0.029 in the range 40 GHz to 50 GHz For Two Ports reflection coefficients in type 50 ohm coaxial system with Calibration and Measurement Capability of - (0.001 to 0.4) linear 0.015 in the range 10 MHz to 9 GHz 0.021 in the range 9 GHz to 14 GHz 0.022 in the range 14 GHz to 18 GHz 0.026 in the range 18 GHz to 26.5 GHz 0.030 in the range 26.5 GHz to 40 GHz 0.034 in the range 40 GHz to 48 GHz 0.036 in the range 48 GHz to 50 GHz (0.4 to 1.0) linear 0.017 in the range 10 MHz to 10 GHz 0.021 in the range 10 GHz to 18 GHz 0.026 in the range 18 GHz to 26.5 GHz 0.033 in the range 26.5 GHz to 40 GHz 0.038 in the range 40 GHz to 50 GHz Determination of phase of reflection coefficient Including onsite calibrations with Calibration and Measurement Capability of – sin⁻¹[linear uncertainty/linear]º (phase uncertainty is ±180º when linear uncertainty > linear) |
| Amplifiers; Attenuators; Communications systems; Data transmission measuring equipment; Impedance and reflection measuring equipment; Radio transmission measuring equipment; Waveguide and coaxial components | Attenuation; Insertion loss; Transmission coefficient | Comparison with a reference standard; Comparison with a reference standard by differential measurement | | |
Capability For attenuation in N type 50 ohm coaxial system Including onsite calibrations with Calibration and Measurement Capability of - 0.1 dB at frequencies from 10 MHz to 18 GHz For attenuation range from 0 dB to 20 dB in a 50 ohm coaxial system Including onsite calibrations with Calibration and Measurement Capability of - 0.11 dB at frequencies from 10 MHz to 2 GHz 0.08 dB at frequencies from 2 GHz to 10 GHz 0.09 dB at frequencies from 10 GHz to 18 GHz 0.10 dB at frequencies from 18 GHz to 26.5 GHz 0.11 dB at frequencies from 26.5 GHz to 40 GHz 0.13 dB at frequencies from 40 GHz to 50 GHz For attenuation range from 20 dB to 40 dB in a 50 ohm coaxial system Including onsite calibrations with Calibration and Measurement Capability of - 0.15 dB at frequencies from 10 MHz to 2 GHz 0.11 dB at frequencies from 2 GHz to 10 GHz 0.12 dB at frequencies from 10 GHz to 18 GHz 0.13 dB at frequencies from 18 GHz to 26.5 GHz 0.14 dB at frequencies from 26.5 GHz to 40 GHz 0.16 dB at frequencies from 40 GHz to 50 GHz |
| Communications systems; Data transmission measuring equipment; Power measuring equipment; Spectrum analysers | Power | Comparison with a reference standard | | |
Capability By Power Sensors Including onsite calibrations with Calibration and Measurement Capability of - 2.5% from DC to 10 MHz 2% from 10 MHz to 18 GHz 3.5% from 18 GHz to 26.5 GHz 4% from 26.5 GHz to 40 GHz 5% from 40 GHz to 50 GHz
Rohde & Schwarz power sensors using NRVC/NRPC system Including onsite calibrations with Calibration and Measurement Capability of - 0.5% for linearity from -30 dBm to 33 dBm
Rohde & Schwarz power sensors using NRVC/NRPC system Determination of absolute power calibration factor Including onsite calibrations at the level of 0 dBm with Calibration and Measurement Capability of - 0.62% for 0.1 MHz ≤ f < 100 MHz 0.81% for 100 MHz ≤ f ≤ 8 GHz 1.05% for 8 GHz < f < 13 GHz 1.3% for 13 GHz ≤ f ≤ 18 GHz 1.76% for 18 GHz < f < 26.5 GHz 2.11% for 26.5 GHz ≤ f < 33 GHz 2.38% for 33 GHz ≤ f < 40 GHz 3.51% for 40 GHz ≤ f ≤ 50 GHz at the level of -10 dBm, with Calibration and Measurement Capability of - 0.62% for 0.1 MHz ≤ f < 100 MHz 0.90% for 100 MHz ≤ f < 8 GHz 1.22% for 8GHz ≤ f < 13 GHz 1.46% for 13 GHz ≤ f ≤ 18 GHz at the level of -20 dBm with Calibration and Measurement Capability of - 1.0% for 0.1 MHz ≤ f < 100 MHz 1.22% for 100 MHz ≤ f < 8 GHz 1.55% for 8 GHz ≤ f < 13 GHz 2.09% for 13 GHz ≤ f < 18 GHz 1.88% for 18 GHz ≤ f < 26.5 GHz 2.14% for 26.5 GHz ≤ f ≤ 33 GHz at the level of -30 dBm with Calibration and Measurement Capability of - 1.22% for 0.1 MHz ≤ f < 100 MHz 1.39% for 100 MHz ≤ f < 8 GHz 1.78% for 8 GHz ≤ f < 13 GHz 1.97% for 13 GHz ≤ f ≤ 18 GHz |
| Electrical noise and interference measuring equipment | Pulse characteristics | Comparison with a reference standard | | Pulse Response for quasi-peak/peak measuring receivers to meet the requirements with AS/NZS CISPR 16.1.1 Band A, B, C/D, E Based on capability of Schwarzbeck Mes-Electronik |
Capability with Calibration and Measurement Capability of - Including onsite calibrations 0.31 dB for Band A 0.32 dB for Band B 0.34 dB for Band C/D 0.86 dB for Band E |
Time and frequency metrology - Frequency and time standards | Signal sources | Frequency; Modulation; Power | Comparison with a reference standard; Direct measurement; Measurement against reference standard | | |
Capability with calibration and Measurement Capability of - Frequency characteristics with calibration and Measurement Capability of - 1.4 in 10¹⁰ up to 50 GHz Including onsite calibrations with calibration and Measurement Capability of - 1.97 in 109 up to 50 GHz Power measurements Including onsite calibrations with calibration and Measurement Capability of - 2.5% from DC to 10 MHz 2% from 10 MHz to 18 GHz 3.5% from 18 GHz to 26.5 GHz 4% from 26.5 GHz to 40 GHz 5% from 40 GHz to 50 GHz Relative power measurement at a single Frequency From DC to 18 GHz Including onsite calibrations with calibration and Measurement Capability of - 1% for dynamic range +15 dBm to − 20 dBm 2% for dynamic range +15 dBm to − 55 dBm Amplitude modulation Including onsite calibrations with least uncertainties of measurement of - 1% for carrier frequencies from 50 kHz to 1.36 GHz, modulation frequencies from 10 Hz to 200 kHz and modulation index from 0 to 95%
Frequency modulation Including onsite calibrations with least uncertainties of measurement of - 0.5% for carrier frequencies from 50 kHz to 1.36 GHz, modulating frequencies from 10 Hz to 50 kHz and deviations up to 700 kHz Phase modulation 2% for carrier frequencies from 50 kHz to 1.36 GHz and modulation frequencies from 300 Hz to 100 kHz and deviations up to 700 rad |
Time and frequency metrology - Frequency, time and waveform measuring equipmentIncluding onsite calibrations | Counters; Frequency analysers; Waveform measuring instruments | Frequency; Time interval | Comparison with a reference standard; Direct measurement; Measurement against reference standard | | |
Capability with Calibration and Measurement capability of - 1.4 in 10¹⁰ up to 50 GHz Including onsite calibrations with calibration and Measurement Capability of - 1.97 in 109 up to 50 GHz |