CATS Pty Ltd

Site

CATS Pty Ltd

Accreditation No.

2124

Site No.

2117

Date of Accreditation

22 Oct 1986

Address

Unit 8/48 May Holman Drive
Bassendean, WA 6054
Australia

calibration.net.au

Contact

Mr Benjamin Arrow P: +61 (08) 61806811

Availability

Services available to external clients

Scope of Accreditation

CATS Pty Ltd

ISO/IEC 17025 (2017)

Calibration

  • The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
Service Product Determinant Technique Procedure Limitations
DC and low frequency electrical metrology - Electrical instrument calibrators Instrument calibrators AC current; AC voltage; DC current; DC voltage; Resistance Comparison with a reference standard

Capability

With Calibration and Measurement Capability of -

DC Voltage
0.01% of reading + 1 μV from 1 mV to 10 V;
0.01% of reading + 10 μV from 10 V to 1000 V
AC Voltage
0.05% of reading from 100 mV to 100 V at 40 Hz to 10 kHz;
0.15% of reading from 100 mV at 10 kHz to 30 kHz;
0.05% of reading from 100 V to  1 kV at 50 Hz to 8 kHz
DC Current

0.05% of reading from 190 μA to 1 A
0.11% of reading above 1 A to 10 A
0.23% of reading above 10 A to 20 A

AC Current
0.07% of reading  from 1 mA to 100 mA at 50 Hz;
0.2% of reading above 100 mA to 10 A at 50 Hz;
0.3% of reading above 10 A to 20 A at 50 Hz
Resistance
1.1% of reading from 0.1 Ω to 1 Ω
0.11% of reading above 1 Ω to 10 Ω
0.005% of reading above 10 Ω to 1 MΩ
0.015% of reading above 1 MΩ to 10 MΩ
0.1% of reading above 10 MΩ to 100 MΩ
1% of reading above 100 MΩ to 1 GΩ
DC and low frequency electrical metrology - Electrical measurement and test equipment Ammeters; Current clamps; Digital multimeters (DMM); LCR meters; Ohm meters; Voltmeters AC current; AC voltage; Capacitance; DC current; DC voltage; Resistance Comparison with a reference standard


Capability

With Calibration and Measurement Capability of -

DC Voltage

0.01% of reading + 1 μV from 1 mV to 10V;
0.01% of reading + 10 μV above 10V to 1000 V;
2% above 1 kV to 5 kV;

AC Voltage
0.05% of reading from 100 mV to 100 V at 50 Hz to 20 kHz;
0.05% of reading from 100 V to 1 kV at 50 Hz to 8 kHz;
DC Current
0.07% of reading from 190 μA to 10 A;
0.23% of reading above 10 A to 20 A
Clamp meters
0.70% of reading from 20 A to 1 000 A

AC Current

1.25% of reading from 0.33 mA to 20 A at 50 Hz

Clamp meters
0.70% of reading from 20 A to 1 000 A at 50 Hz
Resistance
0.90% of reading from 0.1 mΩ to 1 mΩ
0.53% of reading above 1 mΩ to 10 mΩ
0.25% of reading above 10 mΩ to 100 mΩ
0.17% of reading above 100 mΩ to 1 Ω;
0.1% of reading above 1 Ω to 100 Ω
0.02% of reading above 100 Ω to 1000 Ω
0.005% of reading above 1000 Ω to 1 MΩ
0.02% of reading above 1 MΩ to 10 MΩ;
0.12% of reading above 10 MΩ to 100 MΩ
1.5% of reading above 100 MΩ to 1 GΩ
Capacitance
0.6% of reading from 330 pF to 10 nF at 50 Hz to 1 kHz;
1.0% of reading from 10 nF to 3.3 μF at 50 Hz to 300 Hz;
1.0% of reading from 3.3 μF to 33 μF at 50 Hz to 120 Hz;
1.0% of reading from 33 μF to 110 μF at 50 Hz
Insulation resistance test equipment Resistance Comparison with a reference standard

Capability

With Calibration and Measurement capability of -

2% of reading from 1 MΩ to 100 MΩ from 50V to 5 kV

2.3% of reading from 100 MΩ to 1 GΩ from 50 V to 5 kV

DC and low frequency electrical metrology - Electrical standards Conductance boxes; Precision resistors; Resistance boxes Resistance Comparison with a reference standard

Capability

With Calibration and Measurement Capability of -

0.5% of reading from 0.1 mΩ to 10 mΩ
0.25% of reading above 10 mΩ to 100 mΩ
0.03% of reading above 100 mΩ to 10 Ω
0.005% of reading above 10 Ω to 100 kΩ
0.02% of reading above 100 kΩ to 10 MΩ
0.1% of reading above 10 MΩ to 100 MΩ
1% of reading above 100 MΩ to 1 GΩ

Temperature metrology - Non-contact temperature measuring equipment Radiation pyrometers; Thermal imaging systems Temperature Comparison with a reference standard LP-026-03

Capability

with Calibration and Measurement Capability of -
with fixed Emissivity of 0.95
3.9 °C from 50 °C to 100 °C
4.1 °C above 100 °C to 200 °C
4.7 °C above 200 °C to 300 °C
5.6 °C above 300 °C to 400 °C
5.8 °C above 400 °C to 500 °C

Time and frequency metrology - Frequency, time and waveform measuring equipment Frequency analysers; Frequency meters; Oscilloscopes; Stroboscopes; Tachometers; Waveform measuring instruments Frequency; Rotational speed; Time interval Measurement against reference standard

Capability

With Calibration and Measurement Capability of -

Frequency
5 in 106  from 0 Hz to  1 GHz

Rotational speed (optical devices only)
0.05% of reading from 240 rpm to 50 000 rpm
Time interval
2 in 106 from 2 ns to 5 s

The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.

Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.