Scope of Accreditation

Chemical Laboratory

ISO/IEC 17025 (2017)

Calibration

  • The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
Service Product Determinant Technique Procedure Limitations
Chemical metrology - Gas analysers Gas analysers; Mine safety and confined space equipment Concentration Direct measurement by certified reference material in-house method WOS0001

Capability

Calibration in the laboratory or at a customer site (above ground)
With Calibration and Measurement Capability of-
Carbon monoxide - CO (Infra-red): 2.3% of reading over the range 20-5000 part per 106 (ppm)
Carbon monoxide - CO (Electrochemical): 2.3% of reading over the range 1-1000 part per 106 (ppm)
Carbon dioxide - CO2 (Infra-red): 2.3% of reading over the range 0.001-100 %vol/vol
Methane - CH4 (Infra-red): 2.3% of reading over the range 0.01-100 %vol/vol
Nitric oxide - NO (Infra-red): 2.3% of reading over the range 0.0075-100 %vol/vol
Oxygen - O(Paramagnetic): 2.3% of reading over the range 0.01-100 %vol/vol
Oxygen - O2 (Electrochemical): 2.3% of reading over the range 0.01-25 %vol/vol
Sulfur dioxide - SO2 (Infra-red): 2.3% of reading over the range 0.004-100 %vol/vol
Propane - C3H8 (infra-red): 2.3% of reading over the range 0.01-100 %vol/vol 