National Association of Testing Authorities, Australia
Scope of Accreditation
Trescal (Australia) Pty Ltd
Site
Queensland Laboratory
Accreditation No.
19397
Site No.
1628
Date of Accreditation
25 Aug 2023
Contact
Availability
Services available to external clients
Scope of Accreditation
ISO/IEC 17025 (2017)
Calibration
- The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
Service | Product | Determinant | Technique | Procedure | Limitations | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Dimensional metrology - Engineering equipment and precision instruments | Dial gauges | Length measurements | Comparison with a reference standard | including compliance to AS 2103 | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability with Calibration and Measurement Capability of - (1 + 0.025L) μm for indication errors up to 100 mm where L is the range in mm | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Electronic calipers; Vernier calipers | Length measurements | Comparison with a reference standard | including compliance to AS 1984, BS 887, and JIS 7507 | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability with Calibration and Measurement Capability of - (10 + 0.025L) μm for indication errors up to 500 mm where L is the scale value in mm | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Steel rulers and measuring tapes | Length measurements | Comparison with a reference standard | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability with Calibration and Measurement Capability -f - Steel rules | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Force metrology - Force measuring and testing equipmentincluding on-site calibrations | Compression and universal machines in compression; Elastic force measuring devices; Hydraulic rams and jacks; Load cells; Load measuring rings; Tension and universal machines in tension | Force in compression; Force in tension | Comparison measurement with reference load cell | Including compliance with AS 2193; ISO 7500; and ASTM E4 | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability Including on site calibrationswith Calibration and Measurement Capability of - Tension and universal machines and force measuring machines in tension Calibration at forces up to 500 kN to Class AA and 3 MN to Class A requirements to AS 2193 0.17% of reading from 0.01 N to 500 kN 0.25% of reading from 500 kN to 3 MN Compression and universal machines and force measuring systems in compression including self-alignment and restraint on tilt tests for compliance with BS1881 Part 115 Calibration at forces up to 500 kN to Class AA, 20 MN to Class A requirements of AS 2193 0.17% of reading from 0.01 N to 500 kN 0.25% of reading from 500 kN to 20 MN | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Force metrology - Hardness standards and equipmentIncluding on-site calibrations | Portable Brinell measuring microscopes | Length measurements | Direct length measurement | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability Including on-site calibrationswith Calibration and Measurement Capability of 0.006 mm | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Force metrology - Hardness standards and equipmentincluding on site calibrations | Brinell hardness machines | Brinell hardness | Comparison with a reference standard | Partial and complete calibration including compliance to AS 1816.2 | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability Including on-site calibrationswith Calibration and Measurement Capability of 0.8% HB for indirect measurement 0.25% for force measurements | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Rockwell hardness machines; Rockwell superficial hardness machines | Rockwell hardness | Comparison with a reference standard | Partial and complete calibration except indenter dimensions including compliance to AS 1815.2 | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability Including on-site calibrationswith Calibration and Measurement Capability of Starting at 0.4 Rockwell Units for indirect verification 0.25% for force measurements | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Vickers hardness machines; Vickers low-load hardness machines; Vickers micro-hardness machines | Vickers hardness | Comparison with a reference standard | Direct verification except indenter dimensions including compliance to AS 1817 and ISO 146-2 | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability including on site calibrationswith Calibration and Measurement Capability of 1% HV for indirect verification 0.25% for force measurements Vickers micro-hardness machines (less than HV 0.2) Vickers low-load hardness machines (HV 0.2 to HV 5 and 9.8 N to 49 N) Vickers hardness machines from 24.5 N to 980 N | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Mass - Determination of mass and calibration of weighing devices | Industrial mass standards | Mass | Gravimetric measurement against reference mass | From 1 g up to 20 kg | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability with Calibration and Measurement Capability of - 0.9 mg from 1 g up to 100 g; | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Mass - Determination of mass and calibration of weighing devicesincluding on site calibrations | Industrial weighing devices; Laboratory weighing devices; Precision laboratory balances | Mass | Gravimetric measurement against reference mass | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability Including on site calibrations with Calibration and Measurement Capability of - n the range from 0.01 g and up to 400 kg | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Pressure metrology - Pressure and vacuum measuring equipmentincluding on-site calibrations | Digital pressure gauges; Mechanical pressure gauges; Pressure control devices; Pressure gauges; Pressure recorders; Pressure transducers; Vacuum gauges | Gauge pressure | By pressure calibrator; Comparison with dead weight tester | AS 1349 and/or MSA test methods 1 and 2 | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability including on-site calibration With Calibration and Measurement Capability of- For on site calibrations Calibration of industrial gauges 0.23% of reading from -100 kPa to -20 kPa and 20 kPa to 70 MPa 0.46% or 80 Pa (whichever is greater) from -20 kPa to 20 kPa | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Torque - Torque measuring and testing equipmentincluding on site calibrations | Torque wrenches | Torque | Comparison with a reference standard | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability Including on site calibrations with Calibration and Measurement Capability of- 1.5% of reading from 0.1 Nm to 1.2 Nm 1.0 % of reading above 1.2 Nm to 1350 Nm |
The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.
Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.
Accreditation No.
19397
Site No.
1628
Print date