UPT PTY LTD

Site

UPT Instrument Calibration Laboratory

Accreditation No.

18798

Site No.

22748

Date of Accreditation

06 Aug 2015

Address

Unit 6 and 7, 4 Dwyer Court
Chinchilla , QLD 4413
Australia

upt.com.au

Contact

Mr Philip Spann P: +61 (07) 46628450

Availability

Services available to external clients

Scope of Accreditation

UPT Instrument Calibration Laboratory

ISO/IEC 17025 (2017)

Calibration

  • The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
Service Product Determinant Technique Procedure Limitations
Pressure metrology - Pressure and vacuum measuring equipment Pressure control devices; Pressure gauges; Pressure transducers; Vacuum gauges Gauge pressure By pressure calibrator; Comparison with dead weight tester

Pressure and vacuum gauges

including test gauges to the methods of-
AS 1349 and MSA Test Method 2 - 2008

Pressure transducers 

Self indicating and electrical transducers with dc voltage or dc current output to the methods of MSA Test Method 1 - 2008

Pressure control devices

Calibration of Pressure Switches to UPT method UPT_Q_WMS_005 Section 11.4 



Capability

In the range from -90 kPa to 0 kPa and 20 kPa to 260 MPa

with Calibration and Measurement Capability of -

0.07% of reading or 0.005 kPa (whichever is the greater) from -90 kPa to 0 kPa
0.03% of reading from 20 kPa to 1 MPa
0.015% of reading from 1 MPa to 200 MPa
0.025% of reading from 200 MPa to 260 MPa

Torque - Torque measuring and testing equipment Torque wrenches Torque Comparison with a reference standard ISO 6789-2

Capability

with Calibration and Measurement Capability of -

1.5% of reading in the range from 1.25 Nm to 1500 Nm

The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.

Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.