Vokes Air Filtration Pty Ltd

ISO/IEC 17025 (2017)

Infrastructure and Asset Integrity

Service Product Determinant Technique Procedure Limitations
Evaluation of air control equipment Air filters Efficiency - Test dust No.1 Photometric AS 1324.2 Section 4
Mass stabilisation Gravimetric AS 1324.2 Section 3
Gravimetric - Test dust No.4 Gravimetric AS 1324.2 Section 5
Initial resistance Pressure differential AS1324.2 Section 2
Air filters; Gas filters Integrity of HEPA filter installations Cold DOP AS  1807.6 for compliance with AS 4260 Section 5.2 - Integrity testing of HEPA filters

Procedures in factory only.  

Biological safety cabinets Integrity of HEPA filter installations Cold DOP AS 1807.6, (2000), AS 1807.7, AS 1807 clause 4.4, NSF ANSI 49 N5.5 for compliance with AS 2252.1, AS 2252.2, NSF ANSI 49
Gas tightness Not applicable AS 1807.25 (2004), AS 1807 clause 4.11 for compliance with AS 2252.2
Illuminance Not applicable AS 1807.15 (2000), AS 1807 clause 4.5, NSF ANSI 49 N5.9 for compliance with AS 2252.1, AS 2252.2, NSF ANSI 49
Ultraviolet intensity Not applicable AS 1807.23 (2000), AS 1807 clause 4.10 for compliance with AS 2252.2
Air velocity and uniformity of air velocity Not applicable AS 1807.1 (2000), AS 1807 clause 4.1, NSF ANSI 49 N5.2 for compliance with AS 2252.1, AS 2252.2, NSF ANSI 49
Inward air velocity Not applicable AS 1807.21 (2000) AS 1807 clause 4.8, NSF ANSI 49 N5.3 for compliance with AS 2252.1, AS 2252.2, NSF ANSI 49
Work zone integrity Cold DOP AS 1807.5 (2000), AS 1807 clause 4.3 for compliance with AS 2252.2
Air barrier containment Cold DOP; Potassium iodide (KI) discus AS 1807.22 (2000), AS 1807.26 (2004), AS 1807 clause 4.9, 4.12 for compliance with AS 2252.2
Sound level Not applicable AS 1807.20 (2000), AS 1807 clause 4.7, NSF ANSI 49 N5.11 for compliance with AS 2252.1, AS 2252.2, NSF ANSI 49
Clean rooms, pharmaceutical isolators and work stations Recovery times Not applicable AS 1807.24 (2000), AS ISO 14644.3 Method B4 for compliance with AS 2252.6 (Clean Workstation), AS ISO 14644.3 (Clean Room)
Air pressure Not applicable

AS 1807.10, AS ISO 14644.3 Method B1


for compliance with AS 2252.6 (Clean Workstation), AS ISO 14644.3 (Clean Room), AS 4273
Particle count in work zone Automatic particle counter AS 1807.8 (2000), AS ISO 14644.1 for compliance with AS 2252.6 (Clean Workstation), AS ISO 14644.3 (Clean Room), AS 4273
Sound level Not applicable AS 1807.16, AS 1807.20 (2000), AS 1807 Clause 4.7 for compliance with AS 2252.6 (Clean Workstation), AS ISO 14644.3 (Clean Room)
Illuminance Not applicable AS 1807.15 (2000), AS 1807 clause 4.5 for compliance with AS 2252.6 (Clean Workstation), AS ISO 14644.3 (Clean Room), AS 4273
Air velocity and uniformity of air velocity Not applicable AS 1807.1 (2000), AS 1807.3 (2000), AS 1807 clause 4.1, AS ISO 14644.3 Method B2 for compliance with AS 2252.6 (Clean Workstation), AS ISO 14644.3 (Clean Room), AS 4273
Integrity of HEPA filter installations Cold DOP AS 1807.6, (2000), AS 1807.7, AS 1807 clause 4.4, AS ISO 14644.3 Method B7 for compliance with AS 2252.6 (Clean Workstation), AS ISO 14644.3 (Clean Room), AS 4273
Work zone integrity Cold DOP AS 1807.5 (2000), AS 1807 clause 4.3 for compliance with AS 2252.6 (Clean Workstation), AS ISO 14644.3 (Clean Room), AS 4273
Cytotoxic drug safety cabinets Air velocity and uniformity of air velocity Not applicable AS 1807.1 (2000), AS 1807 clause 4.1 for compliance with AS 2252.5
Illuminance Not applicable AS 1807.15 (2000), AS 1807 clause 4.5 for compliance with AS 2252.5
Ultraviolet intensity Not applicable AS 1807.23 (2000), AS 1807 clause 4.10 for compliance with AS 2252.5
Sound level Not applicable AS 1807.20 (2000), AS 1807 clause 4.7 for compliance with AS 2252.5
Work zone integrity Cold DOP AS 1807.5 (2000), AS 1807 clause 4.3 for compliance with AS 2252.5
Particle count in work zone Automatic particle counter AS 1807.8 (2000), AS ISO 14644.1 for compliance with AS 2252.5
Integrity of HEPA filter installations Cold DOP AS 1807.6, (2000), AS 1807.7, AS 1807 clause 4.4 for compliance with AS 2252.5
Fume cupboards and recirculating fume cabinets Extraction performance Face velocity AS/NZS 2243.8 Appendix B; AS/NZS 2243.9 Appendix B
Extraction performance Smoke visualisation AS/NZS 2243.8 Appendix A, AS/NZS 2243.9 Appendix A

The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.

Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.