WIKA Australia Pty Ltd incorporating KD Instruments

ISO/IEC 17025 (2017)

Calibration

  • The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
Service Product Determinant Technique Procedure Limitations
Pressure metrology - Pressure and vacuum measuring equipment Digital pressure gauges; Mechanical pressure gauges; Pressure control devices; Pressure recorders; Pressure transducers; Vacuum gauges Gauge pressure Comparison with dead weight tester; Comparison with reference instrument

MSA Test Method 1, AS 1349 and MSA Test Method 2

DC voltage, DC current and resistance output on transducers

Capability

With Calibration and Measurement Capability of -

± 0.015% of reading or 30 Pa whichever is greater from -100 kPa to 110 MPa

For onsite calibrations

0.15% of reading from 10 kPa to 100 kPa

0.1% of reading from 100 kPa to 110 MPa


Pressure metrology - Pressure and vacuum measuring equipment Digital pressure gauges; Mechanical pressure gauges; Pressure control devices; Pressure recorders; Pressure transducers; Vacuum gauges Absolute pressure Comparison with dead weight tester; Comparison with reference instrument

MSA Test Method 1 

AS 1349 and MSA Test Method 2 
DC voltage, DC current and resistance outputs on transducers

Capability

With Calibration and Measurement Capability of -

0.25 kPa from 1.3 kPa to 110 kPa

0.15% of reading from 110 kPa to 200 kPa

0.1% of reading from 200 kPa to 110 MPa

Temperature metrology - Ancillary temperature measuring equipment Multi-channel thermocouple data recorders; Transmitters Resistance to temperature conversion; Volt to temperature conversion Direct measurement by electrical input Over range of electrical readout

Capability

Including on site calibrations

With Calibration and Measurement Capability of -

0.25 % of full scale reading or 0.5ºC (whichever is greater) Over range of electrical readout

Temperature metrology - Temperature measuring equipment Bimetallic systems; Digital temperature measuring systems; Filled metal systems; Vapour pressure thermometers Temperature Measurement against reference standard

Capability

With Calibration and Measurement Capability of -

Vapour pressure thermometers.  2 °C from 0 °C to 200 °C;

Filled metal systems.   0.5 °C + 1% T from ambient to 600 °C

Bimetallic systems.  0.5 °C + 1% T from 0 °C to 160 °C;

Digital temperature indicator systems (Includes on site calibrations)

 0.2°C + 1 digit from -25 °C to 200 °C

2°C + 1 digit from 200 °C to 1000 °C
Digital temperature measuring systems; Surface probes Temperature Measurement against reference standard

Capability

With Calibration and Measurement Capability of -

Data Loggers
±0.5ºC from -20ºC to 70ºC

Surface Probes
±3.0​​​​ºC from 25ºC to 350ºC


Temperature metrology - Temperature standards and reference equipment Dry block calibrators Temperature Direct measurement using a reference standard

Capability

With Calibration and Measurement Capability of -

1 °C + 1 %T  from 0 °C to 1000 °C

Temperature metrology - Verification of controlled enclosures Baths; Freezers; Furnaces; Incubators; Ovens Gradient; Spatial uniformity; Temperature; Temporal uniformity Direct temperature measurement

In accordance with AS 2853 (except freezers)

Capability

Including on site calibrations

With Calibration and Measurement Capability of -

Ovens and Furnaces.  
1 °C + 1 %T from 0 °C to 1000 °C;
15 °C above 1000 °C to 1200 °C

Incubators. 1 °C + 1 %T from 0 °C to 200 °C; 

Baths.   1 °C + 1 %T from 0 ºC to 550 ºC; 

Industrial freezers  0.5 °C from -20 °C to 0 °C

where T is Temperature in °C

The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.