ISO/IEC 17025 (2017)
Calibration
- The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
Service | Product | Determinant | Technique | Procedure | Limitations | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Pressure metrology - Pressure and vacuum measuring equipment | Digital pressure gauges; Mechanical pressure gauges; Pressure control devices; Pressure recorders; Pressure transducers; Vacuum gauges | Gauge pressure | Comparison with dead weight tester; Comparison with reference instrument | MSA Test Method 1, AS 1349 and MSA Test Method 2 | DC voltage, DC current and resistance output on transducers | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability With Calibration and Measurement Capability of - ± 0.015% of reading or 30 Pa whichever is greater from -100 kPa to 110 MPaFor onsite calibrations 0.15% of reading from 10 kPa to 100 kPa 0.1% of reading from 100 kPa to 110 MPa | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Pressure metrology - Pressure and vacuum measuring equipment | Digital pressure gauges; Mechanical pressure gauges; Pressure control devices; Pressure recorders; Pressure transducers; Vacuum gauges | Absolute pressure | Comparison with dead weight tester; Comparison with reference instrument | MSA Test Method 1 AS 1349 and MSA Test Method 2 | DC voltage, DC current and resistance outputs on transducers | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability With Calibration and Measurement Capability of - 0.25 kPa from 1.3 kPa to 110 kPa 0.15% of reading from 110 kPa to 200 kPa 0.1% of reading from 200 kPa to 110 MPa | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Temperature metrology - Ancillary temperature measuring equipment | Multi-channel thermocouple data recorders; Transmitters | Resistance to temperature conversion; Volt to temperature conversion | Direct measurement by electrical input | Over range of electrical readout | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability Including on site calibrations With Calibration and Measurement Capability of - 0.25 % of full scale reading or 0.5ºC (whichever is greater) Over range of electrical readout | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Temperature metrology - Temperature measuring equipment | Bimetallic systems; Digital temperature measuring systems; Filled metal systems; Vapour pressure thermometers | Temperature | Measurement against reference standard | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability With Calibration and Measurement Capability of - Vapour pressure thermometers. 2 °C from 0 °C to 200 °C; Filled metal systems. 0.5 °C + 1% T from ambient to 600 °C; Bimetallic systems. 0.5 °C + 1% T from 0 °C to 160 °C;Digital temperature indicator systems (Includes on site calibrations) 0.2°C + 1 digit from -25 °C to 200 °C 2°C + 1 digit from 200 °C to 1000 °C | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Digital temperature measuring systems; Surface probes | Temperature | Measurement against reference standard | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability With Calibration and Measurement Capability of -Data Loggers ±0.5ºC from -20ºC to 70ºC Surface Probes ±3.0ºC from 25ºC to 350ºC | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Temperature metrology - Temperature standards and reference equipment | Dry block calibrators | Temperature | Direct measurement using a reference standard | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability With Calibration and Measurement Capability of - 1 °C + 1 %T from 0 °C to 1000 °C | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Temperature metrology - Verification of controlled enclosures | Baths; Freezers; Furnaces; Incubators; Ovens | Gradient; Spatial uniformity; Temperature; Temporal uniformity | Direct temperature measurement | In accordance with AS 2853 (except freezers) | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability Including on site calibrations With Calibration and Measurement Capability of - Ovens and Furnaces. Incubators. 1 °C + 1 %T from 0 °C to 200 °C; Baths. 1 °C + 1 %T from 0 ºC to 550 ºC; Industrial freezers 0.5 °C from -20 °C to 0 °C where T is Temperature in °C |
The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.