ASG Exical Pty Ltd as trustee for the ASG Exical Unit Trust

Site

ASG-X-Technologies

Accreditation No.

15464

Site No.

16363

Date of Accreditation

01 Aug 2006

Address

7/1 Trade Park Drive
Tullamarine, VIC 3043
Australia

asgxtech.com.au

Contact

Ms Saeedeh Saadati P: +61 1800256838

Availability

Services available to external clients

Scope of Accreditation

ASG-X-Technologies

ISO/IEC 17025 (2017)

Calibration

  • The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
Service Product Determinant Technique Procedure Limitations
DC and low frequency electrical metrology - Electrical instrument calibrators Instrument calibrators AC current; AC voltage; DC current; DC voltage; Resistance Comparison with a reference standard

Capability

with Calibration and Measurement Capability of 

DC Voltage
0.0017% of reading + 1.4 μV from 0 V to 0.1 V
0.0011% of reading + 11 μV above 0.1 V to 1 V
0.001% of reading + 50 μV above 1 V to 10 V
0.0014% of reading + 0.8 mV above 10 V to 100 V
0.002% of reading + 16 mV above 100 V to 300 V
AC Voltage 
0.02% of reading from 0.3V to 10 V and from 10 Hz to 20 kHz
0.05% of reading from 0.3V to 10 V and from 20 kHz to 50 kHz
0.12% of reading from 0.3 V to 10.0 V and from 50 kHz to 100 kHz
0.04% of reading from 10 V to 100 V at 10 Hz
0.03% of reading from 10 V to 100 V and from 10 Hz to 20 kHz
0.06% of reading from 10 V to 100 V and from 20 kHz to 50 kHz
0.18% of reading from 10 V to 100 V and from 50 kHz to 100 kHz
0.07% of reading from 100 V to 700 V at 40 Hz
0.09% of reading from 100 V to 700 V and from 40 Hz to 10 kHz

DC Current
0.004% of reading + 0.002 μA from 0 μA to 100 μA
0.003% of reading + 0.14 μA above 100 μA to 10 mA
0.005% of reading + 1.9 μA above 10 mA to 100 mA
0.014% of reading + 30 μA above 100 mA to 1 A
0.14% of reading + 0.75 mA above 1 A to 3A
0.18% of reading + 0.22 mA above 3 A to 10 A

AC Current
0.6% of reading from 10 μA to 1 A at 10 Hz
0.3% of reading from 10 μA to 100 μA and from 10 Hz to 1 kHz
0.1% of reading from 0.1 mA to 100 mA and from 10 Hz to 1 kHz
0.2% of reading from 100 mA to 1 A and from 10 Hz to 1 kHz
2.4% of reading from 1 A to 10 A and from 40 Hz to 5 kHz
 

 Resistance
0.0033% of reading + 0.2 mΩ from 0 Ω to 10 Ω
0.003% of reading + 2 mΩ above 10 Ω to 100 Ω
0.002% of reading + 1.5 Ω above 100 Ω to 100 kΩ
0.003% of reading + 24 Ω above 100 kΩ to 1 MΩ
0.01% of reading + 730 Ω above 1 MΩ to 10 MΩ
0.085% of reading + 0.06 MΩ above 10 MΩ to 100 MΩ

DC and low frequency electrical metrology - Electrical measurement and test equipment Ammeters; Digital multimeters (DMM); Ohm meters; Voltmeters AC current; AC voltage; DC current; DC voltage; Resistance Comparison with a reference standard

Capability

with Calibration and Measurement Capability of -

DC Voltage
68 µV/V + 4.5 µV from 0 V to 320 mV
82 µV/V + 0.1 µV above  320 mV to 3.2 V
86 µV/V above 3.2 V to 32 V
87 µV/V above 32 V to 320 V
140 µV/V above 320 V to 1000 V
AC Voltage
350 µV/V + 56 µV from 300 mV to 320 mV and 10 Hz to 3 kHz
400 µV/V + 29 µV from 300 mV to 320 mV and 3 kHz to 10 kHz
650 µV/V + 35 µV from 300 mV to 320 mV and 10 kHz to 30 kHz        
890 µV/V + 110 µV from 300 mV to 320 mV and 30 kHz to 50 kHz
2200 µV/V + 230µV from 300 mV to 320 mV and 50 kHz to 100 kHz            
270 µV/V + 860 µV from 320 mV to 3.2 V and 10 Hz to 3 kHz
400 µV/V + 270 µV from 320 mV to 3.2 V and 3 kHz to 10 kHz
600 µV/V + 510 µV from 320 mV to 3.2 V and 10 kHz to 30 kHz        
890 µV/V + 980 µV from 320 mV to 3.2 V and 30 kHz to 50 kHz
2000 µV/V + 2.6 mV from 320 mV to 3.2 V and 50 kHz to 100 kHz
270 µV/V + 8. 6mV from 3.2 V to 32 V and 10 Hz to 3 kHz
590 µV/V + 3.1 mV from 3.2 V to 32 V and 3 kHz to 10 kHz
790 µV/V + 5.1 mV from 3.2 V to 32 V and 10 kHz to 30 kHz
1500 µV/V + 9.8 mV from 3.2 V to 32 V and 30 kHz to 50 kHz
3500 µV/V + 32 mV from 3.2 V to 32 V and 50 kHz to 100 kHz
530 µV/V + 19 mV from 32 V to 105 V and 10 Hz to 10 kHz
810 µV/V + 16 mV from 32 V to 105 V and 10 kHz to 30 kHz
1500 µV/V + 33 mV from 32 V to 105 V and 30 kHz to 50 kHz
3500 µV/V + 110 mV from 32 V to 105 V and 50 kHz to 100 kHz
430 µV/V + 53 mV from 105 V to 320 V and 40 Hz to 100 Hz
510 µV/V + 21 mV from 105 V to 320 V and 100 Hz to 1 kHz
850 µV/V + 15 mV from 105 V to 320 V and 1 kHz to 10 kHz
1200 µV/V from 320 V to 750 V and 40 Hz to 10 kHz

DC Current
420 µA/A + 17 nA from 0 µA to 320 µA
140 µA/A + 110 nA above 320 µA to 3.2 mA
180 µA/A + 9 nA above 3.2 mA to 32 mA
250 µA/A + 2.3 µA above 32 mA to 320 mA
9700 µA/A + 3.1 mA above 320 mA to 3.2 A
1800 µA/A + 22 mA above 3.2 A to 10 A
AC Current
940 µA/A + 2.1 µA from 3.2 mA to 32 mA and 40 Hz to 1 kHz
2100 µA/A + 6.8 µA from 32 mA to 320 mA and 40 Hz to 1 kHz
1800 µA/A + 240 µA from 320 mA to 1 A and 40 Hz to 1 kHz
790 µA/A + 1.3 mA from 1 A to 3.2 A and 45 Hz to 1 kHz
2000 µA/A + 3.6 mA from 3.3 A to 10A and 45 Hz to 1 kHz


Resistance
12 mΩ from 0 Ω to 40 Ω
200 µΩ/Ω + 3.9 mΩ  above 40 Ω to 400 Ω
170 µΩ/Ω + 15 mΩ above 400 Ω to 4 kΩ
230 µΩ/Ω + 220 mΩ above 4 kΩ to 40 kΩ
230 µΩ/Ω + 10 Ω above 40 kΩ to 400 kΩ
560 µΩ/Ω + 200 Ω above 400 kΩ to 4 MΩ
890 µΩ/Ω + 2 kΩ above 4 MΩ to 40 MΩ
2500 µΩ/Ω + 70 kΩ above 40 MΩ to 100 MΩ





DC and low frequency electrical metrology - Electrical standards Conductance boxes; Precision resistors; Resistance boxes Resistance Comparison with a reference standard

Capability

with Calibration and Measurement Capability of -

0.0033% of reading + 0.2 mΩ from 0 Ω to 10 Ω 

0.003% of reading + 2 mΩ above 10 Ω to 100 Ω 

0.002% of reading + 1.5 Ω above 100 Ω to 100 kΩ 

0.003% of reading + 24 Ω above 100 kΩ to 1 MΩ 

0.01% of reading + 730 Ω above 1 MΩ to 10 MΩ 

0.085% of reading + 0.06 MΩ above 10 MΩ to 100 MΩ 

Pressure metrology - Pressure and vacuum measuring equipmentincluding on site calibrations Digital pressure gauges; Mechanical pressure gauges; Pressure recorders; Pressure transducers; Vacuum gauges Gauge pressure; High vacuum Comparison with dead weight tester; Comparison with reference instrument


Electrical transducers with dc voltage, dc current or resistance outputs

Capability

Gauge pressure calibration

with Calibration and Measurement Capability of-

0.016% reading or 0.016 kPa (whichever is greater) from  -95 kPa to 100 kPa

0.016% reading above 100 kPa to 120 MPa


On-site calibration

with Calibration and Measurement Capability of-

0.13% of reading or 0.13 kPa (whichever is greater) from -95 kPa to 120 MPa



Pressure metrology - Pressure standards Pressure calibrators - Non-dead weight Gauge pressure Comparison with dead weight tester; Comparison with reference instrument

Capability

with Calibration and Measurement Capability of -
0.016% reading or 0.016 kPa (whichever is greater) from -95 kPa to 100 kPa
0.016% of reading above 100 kPa to 120 MPa

Time and frequency metrology - Frequency, time and waveform measuring equipment Counters; Stroboscopes; Tachometers; Time interval meters Frequency; Rotational speed; Time interval Comparison with a reference standard

Capability

with Calibration and Measurement Capability of -

Counters

0.051% of reading from 1 Hz to 40 Hz
0.015% of reading from above 40 Hz to 10 MHz

Tachometers
0.045% of reading from 60 rpm to 2400 rpm
0.015% of reading from 2400 rpm to 90,000 rpm
Stroboscopes
0.045% of reading from 1 Hz to 40 Hz
0.015% of reading above 40 Hz to 1.5 kHz
Time interval meters
0.015 % of reading from 100 ns to 1 s


The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.

Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.