Trescal (Australia) Pty Ltd

Site

Queensland Laboratory

Accreditation No.

19397

Site No.

1628

Date of Accreditation

25 Aug 2023

Address

28 Staple Street
Seventeen Mile Rocks, QLD 4073
Australia

trescal.com.au

Contact

Mr Anthony Wynn P: +61 1800256838

Availability

Services available to external clients

Scope of Accreditation

Queensland Laboratory

ISO/IEC 17025 (2017)

Calibration

  • The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
Service Product Determinant Technique Procedure Limitations
Dimensional metrology - Engineering equipment and precision instruments Dial gauges Length measurements Comparison with a reference standard

including compliance to  AS 2103

Capability

with Calibration and Measurement Capability of -

(1 + 0.025L) μm for indication errors up to 100 mm where L is the range in mm

Electronic calipers; Vernier calipers Length measurements Comparison with a reference standard

including compliance to  AS 1984, BS 887, and JIS 7507

Capability

with Calibration and Measurement Capability of -

(10 + 0.025L) μm for indication errors up to 500 mm where L is the scale value in mm

Steel rulers and measuring tapes Length measurements Comparison with a reference standard

Capability

with Calibration and Measurement Capability -f -

Steel rules
45 μm up to 2 m in length

Force metrology - Force measuring and testing equipment Load cells; Load measuring rings Force in compression; Force in tension Comparison measurement with reference load cell; Gravimetric measurement with reference mass

Calibration of devices such as load measuring rings

for soil testing machines in tension and compression

to class AA and 

to class A requirements of AS 2193


Electronic load cells and hydrostatic weighing units

including those in batching plants and aircraft weighing kits


Capability

With Calibration and Measurement Capability of -
Tension
0.17% of reading  from 0.01 N to 1000 kN 
0.25% of reading above 1000 kN to 3.3 MN
Class A up to 3.3 MN
Class AA up to 500 kN

Compression
0.17% of reading  from 0.01 N to 500 kN 
0.25% of reading above 500 kN to 10 MN
Class AA up to 500 kN
Class A up to 4.5 MN
Class B up to 10 MN
Force metrology - Force measuring and testing equipmentincluding on-site calibrations Compression and universal machines in compression; Elastic force measuring devices; Hydraulic rams and jacks; Load cells; Load measuring rings; Tension and universal machines in tension Force in compression; Force in tension Comparison measurement with reference load cell Including compliance with  AS 2193; ISO 7500; and ASTM E4

Capability

Including on site calibrations
with Calibration and Measurement Capability of -
Tension and universal machines and force measuring machines in tension
Calibration at forces up to 500 kN to Class AA and 3 MN to Class A requirements to AS 2193
0.17% of reading from 0.01N to 500 kN
0.25% of reading from 500 kN to 3 MN
Compression and universal machines and force measuring systems in compression including self-alignment and restraint on tilt tests for compliance with BS1881 Part 115 
Calibration at forces up to 500 kN to Class AA, 20 MN to Class A requirements of AS 2193
0.17% of reading from 0.01N to 500 kN
0.25% of reading from 500 kN to 10 MN
 


Force metrology - Hardness standards and equipment Indenters; Portable Brinell measuring microscopes Brinell hardness Comparison with a reference standard

Visual examination for Indenters only
Brinell Measuring microscopes are calibrated


Capability

Including on-site calibrations
with Calibration and Measurement Capability of
0.006mm
Force metrology - Hardness standards and equipmentincluding on site calibrations Brinell hardness machines Brinell hardness Comparison with a reference standard

Partial and complete calibration including compliance to AS 1816.2



Capability

Including on-site calibrations
with Calibration and Measurement Capability of
0.8% HB for indirect measurement
0.17% for force measurements
Rockwell hardness machines; Rockwell superficial hardness machines Rockwell hardness Comparison with a reference standard

Partial and complete calibration except indenter dimensions

clause (2.4.1.2b) including compliance to AS 1815.2 and AS 2025

Capability

Including on-site calibrations
with Calibration and Measurement Capability of
Starting at 0.4 Rockwell Units for indirect verification
0.17% for force measurements
Rockwell superficial hardness machines Rockwell hardness Comparison with a reference standard

Partial and complete calibration including compliance to AS 2025

Vickers hardness machines; Vickers low-load hardness machines; Vickers micro-hardness machines Vickers hardness Comparison with a reference standard Direct verification except indenter dimensions including compliance to AS 1817

Capability

including on site calibrations
with Calibration and Measurement Capability of
1% HV for indirect verification
0.17% for force measurements
Vickers micro-hardness machines (less than HV 0.2)
Vickers low-load hardness machines (HV 0.2 to HV 5 and 9.8 N to 49 N)
Vickers hardness machines from 24.5 N to 980 N
Vickers low-load hardness machines Vickers hardness Comparison with a reference standard

(HV 0.2 to HV 5)

Direct verification and indirect verification except indenter dimensions
including compliance to ISO 146

9.8 N to 49 N

Vickers micro-hardness machines Vickers hardness Comparison with a reference standard

(less than HV 0.2)

Direct verification except indenter dimensions including compliance to ISO 146-2

Mass - Determination of mass and calibration of weighing devices Industrial mass standards Mass Gravimetric measurement against reference mass From 1 g up to 20 kg

Capability

with Calibration and Measurement Capability of -

0.9 mg from 1 g up to 100 g;
20 mg from 100 g to 2 kg;
130 mg from 2 kg to 20 kg

Mass - Determination of mass and calibration of weighing devicesincluding on site calibrations Industrial weighing devices; Laboratory weighing devices; Precision laboratory balances Mass Gravimetric measurement against reference mass

Capability

Including on site calibrations

with Calibration and Measurement Capability of -

n the range from 0.01 g and up to 400 kg
2 in 10⁶ or 20 μg (whichever is greater) ) in the range from 0.01g and up to 250 g
3 in 106 in the range from 250 g and up to 500 g
3 mg in the range from 500 g and up to 1.1 kg
2 in 10or 4 mg (whichever is greater) in the range from 1.1 kg and up to 21 kg
4 in 10in the range from 21 kg and up to 31 kg
5 in 10in the range from 31 kg and up to 41 kg
5 in 10in the range from 41 kg and up to 400 kg

Pressure metrology - Pressure and vacuum measuring equipmentincluding on-site calibrations Digital pressure gauges; Mechanical pressure gauges; Pressure control devices; Pressure gauges; Pressure recorders; Pressure transducers; Vacuum gauges Gauge pressure By pressure calibrator; Comparison with dead weight tester AS 1349 and/or MSA test methods 1 and 2

Capability

including on-site calibration

With Calibration and Measurement Capability of-
0.05% of reading from 100 kPa to 70 MPa
0.23% of reading from -100 kPa to -20 kPa and 20 kPa to 100 kPa
0.46% of reading or 80 Pa (whichever is greater) from -20 kPa to 20 kPa


For on site calibrations

Calibration of industrial gauges

0.23% of reading from -100 kPa to -20 kPa and 20 kPa to 70 MPa

0.46% or 80 Pa (whichever is greater) from -20 kPa to 20 kPa


Torque - Torque measuring and testing equipmentincluding on site calibrations Torque wrenches Torque Comparison with a reference standard

Capability

Including on site calibrations

with Calibration and Measurement Capability of-

1.5% of reading from 0.1 Nm to 1.2 Nm

1.0 % of reading above 1.2 Nm to 1350 Nm

The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.

Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.