Brisbane Laboratory

ISO/IEC 17025 (2017)

Calibration

  • The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
Service Product Determinant Technique Procedure Limitations
DC and low frequency electrical metrology - Electrical measurement and test equipment Electrical energy meters; Electricity meters; Varmeters; Watt meters AC current; AC voltage; Active energy; Active power; Apparent energy; Apparent power; Frequency; Phase angle; Reactive energy; Reactive power Comparison with a reference standard

Accuracy requirements of single phase and polyphase electricity meters by the methods of

AS/IEC 62053.21 clause 8

AS/IEC 62053.22 clause 8

AS/IEC 62053.23 clause 8

NMI NITP 14 clause 7

Functional tests for no load and starting current by the methods of

NMI NITP 14 clause 7.2

Capability

with Calibration and Measurement Capability of 

Watthours and Watts at 50 Hz
0.03% of reading for nominal voltages of 63.5 V, 110 V, 230 V, 240 V and currents from 10 mA to 20 mA and unity power factor
(0.03/cos φ)% of reading for nominal voltages of 63.5V, 110V, 230 V, 240 V and currents from 20 mA to 20 A
(0.03/cos φ)% of reading for nominal voltage of 230 V & 240 V and currents from 20 A to 100 A

Varhours and Vars at 50 Hz
(0.05/sin φ)% of reading for nominal voltages of 63.5 V, 110 V, 230 V, 240 V and currents from 20 mA to 20 A
(0.05/sin φ)% of reading for nominal voltages of 230 V and 240 V and currents from 20 A to 100 A
VAhours and VA at 50 Hz
0.023% of reading for nominal voltages of 63.5 V, 110 V, 230 V and 240 V and currents from 50 mA to 20 A
0.023% of reading for nominal voltages of 230 V and 240 V and currents from 20 A to 100 A
A.C. voltage at 50 Hz
0.02% of reading for voltages from 57.7 V to 264 V
A.C. current at 50 Hz
0.03% of reading for currents from 10 mA to 100 A
Phase angle at 50 Hz
0.03º for phase angle from -180º to 0º to +180º
Frequency 
0.001% of reading for frequencies from 45 Hz to 65 Hz 