National Association of Testing Authorities, Australia
Scope of Accreditation
Flowtech (QLD) Pty Ltd
Site
Brisbane Site
Accreditation No.
21619
Site No.
25060
Date of Accreditation
20 Oct 2021
Contact
Availability
Services available to external clients
Scope of Accreditation
ISO/IEC 17025 (2017)
Infrastructure and Asset Integrity
Service | Product | Determinant | Technique | Procedure | Limitations |
---|---|---|---|---|---|
Evaluation of air control equipment | Biological safety cabinets | Air barrier containment | Cold DOP | AS 1807.22 (2000), AS 1807 clause 4.9 | for compliance with AS 2252.1, AS 2252.2 |
Ultraviolet intensity | Not applicable | AS 1807.23 (2000), AS 1807 clause 4.10 | for compliance with AS 2252.1, AS 2252.2 | ||
Inward air velocity | Not applicable | AS 1807.21 (2000) AS 1807 clause 4.8 | for compliance with AS 2252.1, AS 2252.2 | ||
Integrity of HEPA filter installations | Cold DOP | AS 1807.6, (2000), AS 1807.7, AS 1807 clause 4.4 | for compliance with AS 2252.1, AS 2252.2 | ||
Illuminance | Not applicable | AS 1807.15 (2000), AS 1807 clause 4.5 | for compliance with AS 2252.1, AS 2252.2 | ||
Sound level | Not applicable | AS 1807.20 (2000), AS 1807 clause 4.7 | for compliance with AS 2252.1, AS 2252.2 | ||
Air velocity and uniformity of air velocity | Not applicable | AS 1807.1 (2000), AS 1807.3, AS 1807 clause 4.1 | for compliance with AS 2252.1, AS 2252.2 | ||
Work zone integrity | Cold DOP | AS 1807.5 (2000), AS 1807 clause 4.3 | for compliance with AS 2252.1, AS 2252.2 | ||
Clean rooms, pharmaceutical isolators and work stations | Integrity of HEPA filter installations | Not applicable | AS ISO 14644.3 Annex B7.2, B7.3, AS/NZS 14644.3 Annex B.6.2, B 6.3, B13.2.2 (2009) | for compliance with AS 2252.6 (Clean Workstation), AS ISO 14644.3 (Clean Room), AS 4273 | |
Relative humidity (RH) | Not applicable | AS ISO 14644.3 Annex B6, AS/NZS 14644.3 Annex B9 (2009) | for compliance with AS 2252.6 (Clean Workstation), AS ISO 14644.3 (Clean Room), AS 4273 | ||
Air velocity and uniformity of air velocity | Not applicable | AS 1807.1 (2000), AS 1807.3 (2000), AS 1807 clause 4.1, AS ISO 14644.3 Annex B2 | for compliance with AS 2252.6 (Clean Workstation), AS ISO 14644.3 (Clean Room), AS 4273 | ||
Sound level | Not applicable | AS 1807.16, AS 1807.20 (2000), AS 1807 clause 4.7 | for compliance with AS 2252.6 (Clean Workstation), AS ISO 14644.3 (Clean Room) | ||
Particle count in work zone | Automatic particle counter | AS 1807.8 (2000), AS ISO 14644.1 Appendix A | for compliance with AS 2252.6 (Clean Workstation), AS ISO 14644.1 (Clean Room) | ||
Recovery times | Not applicable | AS 1807.24 (2000), AS ISO 14644.3 Annex B4, AS/NZS 14644.3 Annex B.12.3.1 (2009) | for compliance with AS 2252.6 (Clean Workstation), AS ISO 14644.3 (Clean Room) | ||
Illuminance | Not applicable | AS 1807.15 (2000), AS 1807 clause 4.5 | for compliance with AS 2252.6 (Clean Workstation), AS ISO 14644.3 (Clean Room), AS 4273 | ||
Temperature | Not applicable | AS ISO 14644.3 Annex B5, AS/NZS 14644.3 Annex B8 (2009) | for compliance with AS 2252.6 (Clean Workstation), AS ISO 14644.3 (Clean Room), AS 4273 | ||
Integrity of HEPA filter installations | Cold DOP | AS 1807.6, (2000), AS 1807.7, AS 1807 clause 4.4, AS ISO 14644.3 Annex B7, AS/NZS 14644.3 Annex B6 (2009) | for compliance with AS 2252.6 (Clean Workstation), AS ISO 14644.3 (Clean Room), AS 4273 | ||
Air pressure | Not applicable | AS 1807.10 (2000), AS ISO 14644.3 Annex B1 | for compliance with AS 2252.6 (Clean Workstation), AS ISO 14644.3 (Clean Room), AS 4273 | ||
Fume cupboards and recirculating fume cabinets | Extraction performance | Smoke visualisation | AS/NZS 2243.8 Appendices A, AS/NZS 2243.9 Appendices A | for compliance with AS/NZS 2243.8, AS/NZS 2243.9 | |
Extraction performance | Face velocity | AS/NZS 2243.8 Appendix B; AS/NZS 2243.9 Appendix B | for compliance with AS/NZS 2243.8, AS/NZS 2243.9 | ||
Operating theatres and treatment rooms | Air velocity and uniformity of air velocity | Not applicable | in-house method FTNSOP-22 | Table height velocity measurement |
ISO/IEC 17025 (2017)
Manufactured Goods
Service | Product | Determinant | Technique | Procedure | Limitations |
---|---|---|---|---|---|
Performance evaluation of personal protective equipment (PPE) | Medical face masks | Synthetic blood penetration | Not applicable | AS 4381 Table 2, ASTM F1862/F1862M | Single-use face masks (target plate method only) |
Breathability | Not applicable | AS 4381 Table 2, EN 14683 (2014) clause 5.2.3 and Annex C | |||
Bacterial filtration efficiency (BFE) | Aerosolised bacterial impaction | AS 4381 Table 2, ASTM F2101, EN 14683 (2014) Clause 5.2.2 and Annex B, EN 14683 Clause 5.2.2 and Annex B | Single-use face masks | ||
Respiratory protective devices (respirators) | Inhalation resistance | Not applicable | AS/NZS 1716 Appendix G | Excluding exhalation resistance |
The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.
Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.
Accreditation No.
21619
Site No.
25060
Print date