National Association of Testing Authorities, Australia
Scope of Accreditation
Camfil Australia Pty Ltd
Site
Camfil Technical Services – NSW
Accreditation No.
15330
Site No.
24198
Date of Accreditation
01 Jun 2017
Contact
Mr Praseuth Voraboud
P: +61 0429605783
Availability
Services available to external clients
Scope of Accreditation
Camfil Technical Services – NSW
ISO/IEC 17025 (2017)
Infrastructure and Asset Integrity
Service | Product | Determinant | Technique | Procedure | Limitations |
---|---|---|---|---|---|
Evaluation of air control equipment | Biological safety cabinets | Air velocity and uniformity of air velocity | Cold DOP | AS 1807 clause 4.1 | For determining compliance with AS 2252.2 |
Inward air velocity | Cold DOP |
AS 1807 clause 4.8, |
for compliance with: AS 2252.1 (2002), AS 2252.2 (Biological safety cabinets), |
||
Biological safety cabinets; Cytotoxic drug safety cabinets | Particle count in work zone | Automatic particle counter | AS ISO 14644.1 Appendix A |
for compliance with: AS 2252.1 (2002), AS 2252.2 (Biological safety cabinets), AS 2252.5 (Cytotoxic safety cabinets) |
|
Ultraviolet intensity | Not applicable | AS 1807 clause 4.10 |
for compliance with: AS 2252.1 (2002), AS 2252.2 (Biological safety cabinets), AS 2252.5 (Cytotoxic safety cabinets) |
||
Integrity of HEPA filter installations | Cold DOP |
AS 1807.7, AS 1807 clause 4.4 |
for compliance with: AS 2252.1 (2002), AS 2252.2 (Biological safety cabinets), AS 2252.5 (Cytotoxic safety cabinets), AS 4260 clause 5.2 |
||
Air velocity and uniformity of air velocity | Cold DOP |
AS 1807 clause 4.1 |
for compliance with: AS 2252.1 (2002), AS 2252.2 (Biological safety cabinets), AS 2252.5 (Cytotoxic safety cabinets) |
||
Sound level | Not applicable |
AS 1807 clause 4.7 |
for compliance with: AS 2252.1 (2002), AS 2252.2 (Biological safety cabinets), AS 2252.5 (Cytotoxic safety cabinets) |
||
Work zone integrity | Cold DOP |
AS 1807 clause 4.3 |
for compliance with: AS 2252.1 (2002), AS 2252.2 (Biological safety cabinets), AS 2252.5 (Cytotoxic safety cabinets) |
||
Illuminance | Not applicable | AS 1807 clause 4.5 |
for compliance with: AS 2252.1 (2002), AS 2252.2 (Biological safety cabinets), AS 2252.5 (Cytotoxic safety cabinets) |
||
Air barrier containment | Cold DOP |
AS 1807 clause 4.9 |
for compliance with: AS 2252.1 (2002), AS 2252.2 (Biological safety cabinets), AS 2252.5 (Cytotoxic safety cabinets) |
||
Clean rooms | Sound level | Not applicable | AS 1807.16 |
for compliance with: AS ISO 14644.1, AS ISO 14644.3 |
|
Air velocity and uniformity of air velocity | Cold DOP | AS ISO 14644.3 clause 4.2.2, 4.2.3, Methods B.2, B.3 |
for compliance with: AS ISO 14644.1, AS ISO 14644.3 |
||
Recovery times | Not applicable | AS ISO 14644.3 clause 4.2.4, Method B.4 |
for compliance with: AS ISO 14644.1, AS ISO 14644.3 |
||
Clean rooms, pharmaceutical isolators and work stations | Air velocity and uniformity of air velocity | Cold DOP | AS 1807 clause 4.1 |
for compliance with: AS 2252.6 (Clean Workstation), AS 4273 clause 9.4 (Pharmaceutical isolators) |
|
Integrity of HEPA filter installations | Cold DOP |
AS 1807.7, AS 1807 clause 4.4, AS ISO 14644.3 Method B.7 |
for compliance with: AS 2252.6 (Clean Workstation), AS 4260 clause 5.2, AS 4273 clause 9.3 (Pharmaceutical isolators), AS ISO 14644.1 (Clean Room), AS ISO 14644.3 (Clean Room) |
||
Illuminance | Not applicable | AS 1807 clause 4.5 |
for compliance with: AS 2252.6 (Clean Workstation), AS ISO 14644.1 (Clean Room), AS ISO 14644.3 (Clean Room) |
||
Work zone integrity | Cold DOP | AS 1807 clause 4.3 |
for compliance with: AS 2252.6 (Clean Workstation), AS ISO 14644.1 (Clean Room), AS ISO 14644.3 (Clean Room) |
||
Air pressure | Not applicable |
AS ISO 14644.3 clause 4.2.1, Method B.1 |
for compliance with: AS 2252.6 (Clean Workstation), AS 4273 clause 9.5 (Pharmaceutical isolators), AS ISO 14644.1 (Clean Room), AS ISO 14644.3 (Clean Room) |
||
Particle count in work zone | Automatic particle counter | AS ISO 14644.1 Appendix A |
AS 2252.6 (Clean Workstation), AS ISO 14644.1 (Clean Room), AS ISO 14644.3 (Clean Room) |
||
Ultraviolet intensity | Not applicable | AS 1807 clause 4.10 |
for compliance with: AS 2252.6 (Clean Workstation), AS ISO 14644.1 (Clean Room), AS ISO 14644.3 (Clean Room) |
||
Fume cupboards and recirculating fume cabinets | Illuminance | Not applicable | AS 1807 clause 4.7 |
for compliance with: AS/NZS 2243.8, AS/NZS 2243.9 |
|
Extraction performance | Smoke visualisation |
AS/NZS 2243.8 Appendices A AS/NZS 2243.9 Appendices A |
|||
Sound level | Not applicable | AS 1807 clause 4.7 |
for compliance with: AS/NZS 2243.8, AS/NZS 2243.9 |
||
Extraction performance | Face velocity |
AS/NZS 2243.8 Appendix B, |
for compliance with: AS/NZS 2243.8, AS/NZS 2243.9 |
The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.
Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.
Accreditation No.
15330
Site No.
24198
Print date
08 Dec 2023
END OF SCOPE