Scope of Accreditation

Camfil Technical Services – Queensland

ISO/IEC 17025 (2017)

Infrastructure and Asset Integrity

Service Product Determinant Technique Procedure Limitations
Evaluation of air control equipment Biological safety cabinets Inward air velocity Cold DOP AS 1807 clause 4.8,
NSF ANSI 49 N5.3
for compliance with:
AS 2252.1 (2002), AS 2252.2 (Biological safety cabinets),
NSF ANSI 49
Biological safety cabinets; Cytotoxic drug safety cabinets Work zone integrity Cold DOP

AS 1807 clause 4.3

for compliance with:
AS 2252.1 (2002), AS 2252.2 (Biological safety cabinets),
AS 2252.5 (Cytotoxic safety cabinets)
Sound level Not applicable AS 1807 clause 4.7,
NSF ANSI 49 N5.11
for compliance with:
AS 2252.1 (2002), AS 2252.2 (Biological safety cabinets),
AS 2252.5 (Cytotoxic safety cabinets),
NSF ANSI 49
Air barrier containment Cold DOP AS 1807 clause 4.9
for compliance with:
AS 2252.1 (2002), AS 2252.2 (Biological safety cabinets),
 AS 2252.5 (Cytotoxic safety cabinets)
Air velocity and uniformity of air velocity Cold DOP AS 1807 clause 4.1,
NSF ANSI 49 N5.2
for compliance with:
AS 2252.1 (2002), AS 2252.2 (Biological safety cabinets),
AS 2252.5 (Cytotoxic safety cabinets),
NSF ANSI 49
Integrity of HEPA filter installations Cold DOP AS 1807.7,
AS 1807 clause 4.4,
NSF ANSI 49 N5.5
for compliance with:
AS 2252.1 (2002), AS 2252.2 (Biological safety cabinets),
AS 2252.5 (Cytotoxic safety cabinets),
AS 4260 clause 5.2,
NSF ANSI 49
Ultraviolet intensity Not applicable AS 1807 clause 4.10 for compliance with:
AS 2252.1 (2002), AS 2252.2 (Biological safety cabinets),
AS 2252.5 (Cytotoxic safety cabinets)
Particle count in work zone Automatic particle counter AS ISO 14644.1 Appendix A for compliance with:
AS 2252.1 (2002), AS 2252.2 (Biological safety cabinets),
AS 2252.5 (Cytotoxic safety cabinets)
Illuminance Not applicable AS 1807 clause 4.5,
NSF ANSI 49 N5.9
for compliance with:
AS 2252.1 (2002), AS 2252.2 (Biological safety cabinets),
AS 2252.5 (Cytotoxic safety cabinets),
NSF ANSI 49
Air flow Not applicable NSF ANSI 49 N5.4 for compliance with:
NSF ANSI 49
Clean rooms Recovery times Not applicable AS ISO 14644.3 clause 4.2.4, Method B.4 for compliance with:
AS ISO 14644.1,
AS ISO 14644.3
Sound level Not applicable AS 1807.16 for compliance with:
AS ISO 14644.1,
 AS ISO 14644.3
Air velocity and uniformity of air velocity Cold DOP AS ISO 14644.3 clause 4.2.2, 4.2.3, Methods B.2, B.3 for compliance with:
AS ISO 14644.1,
AS ISO 14644.3
Clean rooms, pharmaceutical isolators and work stations Ultraviolet intensity Not applicable AS 1807 clause 4.10 for compliance with:
AS 2252.6 (Clean Workstation),
AS ISO 14644.1 (Clean Room),
AS ISO 14644.3 (Clean Room)
Air velocity and uniformity of air velocity Cold DOP AS 1807 clause 4.1 for compliance with:
AS 2252.6 (Clean Workstation),
AS 4273 clause 9.4 (Pharmaceutical isolators)
Integrity of HEPA filter installations Cold DOP AS 1807.7,
AS 1807 clause 4.4,
AS ISO 14644.3 Method B.7
for compliance with:
AS 2252.6 (Clean Workstation),
AS 4260 clause 5.2,
AS 4273 clause 9.3 (Pharmaceutical isolators),
AS ISO 14644.1 (Clean Room),
AS ISO 14644.3 (Clean Room)
Air pressure Not applicable

AS ISO 14644.3 clause 4.2.1, Method B.1

for compliance with:
AS 2252.6 (Clean Workstation),
AS 4273 clause 9.5 (Pharmaceutical isolators),
AS ISO 14644.1 (Clean Room),
AS ISO 14644.3 (Clean Room)
Work zone integrity Cold DOP AS 1807 clause 4.3 for compliance with:
AS 2252.6 (Clean Workstation),
AS ISO 14644.1 (Clean Room),
AS ISO 14644.3 (Clean Room)
Particle count in work zone Automatic particle counter AS ISO 14644.1 Appendix A for compliance with:
AS 2252.6 (Clean Workstation),
AS ISO 14644.1 (Clean Room),
AS ISO 14644.3 (Clean Room)
Illuminance Not applicable AS 1807 clause 4.5 for compliance with:
AS 2252.6 (Clean Workstation),
AS ISO 14644.1 (Clean Room),
AS ISO 14644.3 (Clean Room)
Fume cupboards and recirculating fume cabinets Extraction performance Face velocity

AS/NZS 2243.8 Appendix B,
AS/NZS 2243.9 Appendix B

for compliance with:
AS/NZS 2243.8,
AS/NZS 2243.9
Extraction performance Smoke visualisation AS/NZS 2243.8 Appendix A,
AS/NZS 2243.9 Appendix A
for compliance with:
AS/NZS 2243.8,
AS/NZS 2243.9
Sound level Not applicable AS 1807 clause 4.7 for compliance with:
AS/NZS 2243.8,
AS/NZS 2243.9
Illuminance Not applicable AS 1807 clause 4.5 for compliance with:
AS/NZS 2243.8,
AS/NZS 2243.9