Draeger Australia Pty Ltd

Site

Draeger Services National Service Centre

Accreditation No.

13694

Site No.

13687

Date of Accreditation

08 Mar 2007

Address

8 ACACIA PLACE
Notting Hill, VIC 3168
Australia

draeger.com

Contact

Mr David Gladman P: +61 0402927176

Availability

Services available to external clients

Scope of Accreditation

Draeger Services National Service Centre

ISO/IEC 17025 (2017)

Calibration

  • The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
Service Product Determinant Technique Procedure Limitations
Chemical metrology - Breath alcohol instruments Drager 7110; Drager 9510 Alcohol content Direct measurement against a reference standard

in-house method DSP-SE-0020

Capability

With Calibration and Measurement Capability of -
0.0029 g/210L for above 0.000 to less than 0.080 g/210L,
 3.6% of reading from 0.080 to 0.500 g/210L,
 6.3% of reading above 0.500 g/210L
Chemical metrology - Gas analysers Gas analysers; Mine safety and confined space equipment Concentration Direct measurement by certified reference material

in-house method DSP SE0010

Capability

With Calibration and Measurement Capability of - 
Carbon monoxide: Electrochemical; 3.2% of reading in the range of 1 to 405 part per 106 vol/vol (ppm)

Carbon dioxide: Electrochemical; 25% of reading in the range of 0.2 to 5 %vol/vol

Carbon dioxide: Infra-red; 3.1% of the reading in the range of 0.01 to 100 %vol/vol

Methane: Catalytic combustion; 4.2% of reading in the range of 0.1 to 5 %vol/vol

Methane: Infra-red; 5.9% of reading in the range of 0.05 to 100 %vol/vol

Nitrogen oxide: Electrochemical; 5.2% of reading in the range of 0.3 to 200 part per 106 vol/vol (ppm)

Nitrogen dioxide : Electrochemical; 4.7% of reading in the range of 0.04 to 50 part per 106 vol/vol (ppm)

Oxygen: Electrochemical; 3.0 of reading in the range of 0.1 to 25 %vol/vol

Hydrogen sulfide: Electrochemical; 4.8% of reading in the range of 0.4 to 200 part per 106 vol/vol (ppm)

Sulfur dioxide: Electrochemical; 4.6% of reading in the range of 0.1 to 100 part per 106 vol/vol (ppm)

Mobile department

ISO/IEC 17025 (2017)

Calibration

  • The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
Service Product Determinant Technique Procedure Limitations
Chemical metrology - Gas analysers Gas analysers; Mine safety and confined space equipment Concentration Direct measurement by certified reference material

in-house method DSP SE0010

Capability

With Calibration and Measurement Capability of -
Carbon monoxide: Electrochemical; 3.2% of reading in the range of 1 to 405 part per 106 vol/vol (ppm)
Carbon dioxide: Electrochemical; 25% of reading in the range of 0.2 to 5 %vol/vol
Carbon dioxide: Infra-red; 3.1% of the reading in the range of 0.01 to 100 %vol/vol
Methane: Catalytic combustion; 4.2% of reading in the range of 0.1 to 5 %vol/vol
Methane: Infra-red; 5.9% of reading in the range of 0.05 to 100 %vol/vol
Nitrogen oxide : Electrochemical; 5.2% of reading in the range of 0.3 to 200 part per 106 vol/vol (ppm)
Nitrogen dioxide : Electrochemical; 4.7% of reading in the range of 0.04 to 50 part per 106 vol/vol (ppm)
Oxygen: Electrochemical; 3.0 of reading in the range of 0.1 to 25 %vol/vol
Hydrogen sulfide: Electrochemical; 4.8% of reading in the range of 0.4 to 200 part per 106 vol/vol (ppm)
Sulfur dioxide: Electrochemical; 4.6% of reading in the range of 0.1 to 100 part per 106 vol/vol (ppm)