Environmental Monitoring and Analysis Division – Abu Dhabi

ISO/IEC 17025 (2017)

Calibration

  • The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
Service Product Determinant Technique Procedure Limitations
Chemical metrology - Gas analysers Gas analysers Concentration Direct measurement by certified reference material

in-house methods  EAD-EQ-EMA-11,EAD-EQ-EMA-12,  EAD-EQ-EMA-14, EAD-EQ-EMA-15, EAD-EQ-EMA-16, EAD-EQ-EMA-31,  EAD-EQ-EMA-42,  EAD-EQ-EMA-72,  EAD-EQ-EMA-PM10

and manufacturer's instructions

CO : IR, 0 - 50 ppm

H2S : Fluorescence,  0 - 500 ppb 

NO/NOx : Chemiluminescence,  0 - 500 ppb

SO2: Fluorescence, 0 - 500 ppb

Capability

With calibration and measurement capability - 

Benzene

With least uncertainties of -
1ppb or 5% of reading (whichever is greater)
using in-house method EAD-EQ-EMA-72

CO : IR

With least uncertainties of -

0.7 ppm or 7 % of reading (whichever is greater)

using in-house method EAD-EQ-EMA-12 and EAD-EQ-EMA-31

H2S : Fluorescence

With least uncertainties of -

5 ppb or 5 % of reading (whichever is greater)

using in-house methods EAD-EQ-EMA-11 and EAD-EQ-EMA-14

NO/NOx : Chemiluminescence

With least uncertainties of measurement of -

8 ppb or 8 % of reading (whichever is greater)

using in-house methods EAD-EQ-EMA-15 and EAD-EQ-EMA-16

Ozone (O3):
With least uncertainties of -
5ppb or 5% of reading (whichever is greater)
using in-house method EAD-EQ-EMA-42

SO2: Fluorescence

With least uncertainties of measurement of -

5 ppb or 5 % of reading (whichever is greater)

using in-house methods EAD-EQ-EMA-11 and EAD-EQ-EMA-14

PM2.5, PM10: beta attenuation monitor 0-1000µg/M3
With least uncertainties of measurement of –
18% of reading
 using in-house EAD-EQ-EMA-PM10