Environment Agency - Abu Dhabi (EAD)

Site

Environmental Monitoring and Analysis Division – Abu Dhabi

Accreditation No.

19099

Site No.

21539

Date of Accreditation

28 Nov 2013

Address

Envirozone L.L.C, M-41 Workshop 2, Industrial City Abu Dhabi (ICAD), Mussafah
0
United Arab Emirates

adairquality.ae

Contact

Mr Sultan Ahmed Mallick P: 971 503277547

Availability

Services not available to external clients

Scope of Accreditation

Environmental Monitoring and Analysis Division – Abu Dhabi

ISO/IEC 17025 (2017)

Calibration

  • The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
Service Product Determinant Technique Procedure Limitations
Chemical metrology - Gas analysers Gas analysers Concentration Direct measurement by certified reference material

in-house methods  EAD-EQ-EMA-11,EAD-EQ-EMA-12,  EAD-EQ-EMA-14, EAD-EQ-EMA-15, EAD-EQ-EMA-16, EAD-EQ-EMA-31,  EAD-EQ-EMA-42,  EAD-EQ-EMA-72,  EAD-EQ-EMA-PM10

and manufacturer's instructions

CO : IR, 0 - 50 ppm

H2S : Fluorescence,  0 - 500 ppb 

NO/NOx : Chemiluminescence,  0 - 500 ppb

SO2: Fluorescence, 0 - 500 ppb

Capability

With calibration and measurement capability - 

Benzene

With least uncertainties of -
1ppb or 5% of reading (whichever is greater)
using in-house method EAD-EQ-EMA-72

CO : IR

With least uncertainties of -

0.7 ppm or 7 % of reading (whichever is greater)

using in-house method EAD-EQ-EMA-12 and EAD-EQ-EMA-31

H2S : Fluorescence

With least uncertainties of -

5 ppb or 5 % of reading (whichever is greater)

using in-house methods EAD-EQ-EMA-11 and EAD-EQ-EMA-14

NO/NOx : Chemiluminescence

With least uncertainties of measurement of -

8 ppb or 8 % of reading (whichever is greater)

using in-house methods EAD-EQ-EMA-15 and EAD-EQ-EMA-16

Ozone (O3):
With least uncertainties of -
5ppb or 5% of reading (whichever is greater)
using in-house method EAD-EQ-EMA-42

SO2: Fluorescence

With least uncertainties of measurement of -

5 ppb or 5 % of reading (whichever is greater)

using in-house methods EAD-EQ-EMA-11 and EAD-EQ-EMA-14

PM2.5, PM10: beta attenuation monitor 0-1000µg/M3
With least uncertainties of measurement of –
18% of reading
 using in-house EAD-EQ-EMA-PM10

The only data displayed is that deemed relevant and necessary for the clear description of the Activities and Services covered by the Scope of Accreditation (SoA).

Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.

The Product column may also describe an Item or Material.

The Limitations column may also describe the Range.