ABB Australia Pty Limited

Site

Flowmeter Calibration Laboratory

Accreditation No.

1251

Site No.

1244

Date of Accreditation

01 Jun 2017

Address

Bapaume Road
Moorebank, NSW 2170
Australia

new.abb.com/au

Contact

Mr Michael Freeman P: +61 0409115769

Availability

Services available to external clients

Scope of Accreditation

Flowmeter Calibration Laboratory

ISO/IEC 17025 (2017)

Calibration

  • The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
Service Product Determinant Technique Procedure Limitations

Suspended

Flow metrology - Flow measuring devices and systems
Liquid meters Flow rate; Minimum volume Gravimetric

Calibration of flow metering systems on water by gravimetric method at flow rates from 0.1 L/s up to 750 L/s

Capability

0.15% from 0.1 L/s to 2 L/s with a minimum 300L test volume
0.15% from 2 L/s to 100 L/s with a minimum 4000L test volume
0.25% from 100 L/s to 750 L/s with a minimum 16000L test volume

The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.

Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.