IPD Group Ltd

Site

IPD Services (Bendigo)

Accreditation No.

20292

Site No.

15952

Date of Accreditation

01 Jun 2017

Address

10 Maynard Drive
Epsom, VIC 3551
Australia

ipd.com.au

Contact

Mr Jeremy McGuckian P: +61 (03) 54456222

Availability

Services available to external clients

Scope of Accreditation

IPD Services (Bendigo)

ISO/IEC 17025 (2017)

Calibration

  • The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
Service Product Determinant Technique Procedure Limitations
DC and low frequency electrical metrology - Electrical measurement and test equipment Ammeters; Current clamps; Data recorders; Digital multimeters (DMM) AC current Comparison with a reference standard

Capability

With a Calibration and Measurement Capability of -

At 0.25 μA to 1000 A at 50 Hz

0.5% reading + 0.25 μA up to 0.33 mA
0.4% reading + 0.3 μA from 0.33 mA to 3.3 mA
0.36% reading + 3 μA from 3.3 mA to 33 mA
0.36% reading + 30 μA from 33 mA to 330 mA

0.4% reading + 300 μA from 330 mA to 2.2 A
0.24% reading + 2 mA from 2.2 A to 11 A
0.8% reading from 11 A to 100 A
2.0% reading from 100 A to 1000 A

Ammeters; Current clamps; Digital multimeters (DMM) DC current Comparison with a reference standard

Capability

With a Calibration and Measurement Capability of -

At 0.05 μA to 1000 A
0.052% reading + 0.05 μA up to 3.3 mA
0.04% reading + 0.25 μA from 3.3 mA to 33 mA
0.04% reading + 3.3 μA from 33 mA to 330 mA
0.12% reading + 44 μA from 330 mA to 2.2 A
0.24% reading + 330 μA from 2.2 A to 11 A
0.8% reading from 11 A to 100 A
2.0% reading from 100 A to 1000 A

Data recorders; Digital multimeters (DMM); Voltmeters AC voltage Comparison with a reference standard

Capability

With a Calibration and Measurement Capability of -

At 45 Hz to 10 kHz
At 20 μV to 1020 V

0.6% reading + 20 μV from 1 to 33 mV
0.2% reading + 20 μV from 33 to 330 mV
0.12% reading + 60 μV from 0.330 mV to 3.3 V
0.16% reading + 600 μV from 3.3 to 33 V
0.2% reading + 6.6 mV from 33 to 330 V
0.2% reading + 80 mV from 330 to 1020 V

Digital multimeters (DMM); Ohm meters Resistance Comparison with a reference standard

Capability

With a Calibration and Measurement Capability of -
0.048% reading + 0.008 Ω from 0 to 11 Ω
0.048% reading + 0.015 Ω from 11 to 33 Ω
0.036% reading + 0.015 Ω from 33 to 330 Ω
0.036% reading + 0.06 Ω from 330 Ω to 3.3 kΩ
0.036% reading + 0.6 Ω from 3.3 to 33 kΩ
0.048% reading + 6 Ω from 33 to 330 kΩ
0.06% reading + 55 Ω from 330 kΩ to 3.3 MΩ
0.4% reading + 550 Ω from 3.3 to 33 MΩ
2.0% reading + 5.5 kΩ from 33 to 110 MΩ
2.0% reading + 16.5 kΩ from 110 to 330 MΩ

Digital multimeters (DMM); Voltmeters DC voltage Comparison with a reference standard

Capability

With a Calibration and Measurement Capability of -

At 3 μV to 1020 V
0.024% reading + 3 μV up to 330 mV
0.02% reading + 5 μV from 300 mV to 3 V
0.02% reading + 50 μV from 3 V to 33 V
0.022% reading + 0.5 mV from 33 V to 330 V
0.022% reading +1.5 mV from 330 V to 1020 V

Insulation resistance test equipment Resistance Comparison with a reference standard

Capability

with Calibration and Measurement Capability of -
0.048% reading + 0.008 Ω from 0 to 11 Ω up to 500 V dc
0.048% reading + 0.015 Ω from 11 to 33 Ω  up to 500 V dc
0.036% reading + 0.015 Ω from 33 to 330 Ω  up to 500 V dc
0.036% reading + 0.06 Ω from 330 Ω to 3.3 kΩ  up to 500 V dc
0.036% reading + 0.6 Ω from 3.3 to 33 kΩ  up to 500 V dc
0.048% reading + 6 Ω from 33 to 330 kΩ  up to 500 V dc
0.06% reading + 55 Ω from 330 kΩ to 3.3 MΩ  up to 500 V dc
0.4% reading + 550 Ω from 3.3 to 33 MΩ  up to 500 V dc
2.0% reading + 5.5 kΩ from 33 to 100 MΩ  up to 500 V dc

Time and frequency metrology - Frequency, time and waveform measuring equipment Counters; Frequency analysers; Frequency meters; Waveform measuring instruments Frequency Measurement against reference standard

Capability

With a Calibration and Measurement Capability of -
At 0.01 Hz to 2 MHz
100 μHz/Hz + 1 mHz from 0.01 Hz to 10 kHz
100 μHz/Hz + 15 mHz above 10 kHz to 2 MHz
Waveform measuring instruments Current Measurement against reference standard; To be determined

Capability

With a Calibration and Measurement Capability of - 

A, C. Current input at 50 Hz

At 0.25 μA to 1000 A
 0.5% reading + 0.25 μA up to 0.33 mA
 0.4% reading  + 0.3 μA from 0.33 mA to 3.3 mA
 0.36% reading + 3 μA from 3.3 mA to 33 mA
 0.36% reading + 30 μA from 33 mA to 330 mA
 0.4% reading + 300 μA from 330 mA to 2.2 A
 0.24% reading + 2 mA from 2.2 A to 11 A
 0.8% reading from 11 A to 100 A
 2.0% reading from 100 A to 1000 A

Voltage Measurement against reference standard

Capability

With a Calibration and Measurement Capability of -
A.C. Voltage input at 45 Hz to 10 kHz

At 1 mV to 1020 V
0.6% reading + 20 μV from 1 mV  to 33 mV
 0.2% reading + 20 μV from 33 to 330 mV
 0.12% reading + 60 μV from 0.330 mV to 3.3 V
 0.16% reading + 600 μV from 3.3 to 33 V
 0.2% reading + 6.6 mV from 33 to 330 V
 0.2% reading + 80 mV from 330 to 1020 V

ISO/IEC 17025 (2017)

Infrastructure and Asset Integrity

Service Product Determinant Technique Procedure Limitations
Evaluation of high voltage/power equipment and components - Performance Clothing materials, gloves and tools for use in high voltage/power applications Workplace electrical safety Not applicable

AS 2225 (1994);

ASTM F496;

IEC 60903; IEC 60984;

IEEE 978 (1984)

Electricity grid and industrial power supply equipment and systems AC voltage Not applicable in-house procedures CTMW-101 to 110 inclusive
Insulated systems Workplace electrical safety Not applicable

AS/NZS 2978 (1995)

IEC 61111; IEC 61112; IEC 61229;

Modiewark high voltage proximity indicators to procedure CTMW-006

Platform access equipment for use in high voltage/power applications Workplace electrical safety Not applicable

AS/NZS 1418.10; 

ANSI/SIA A92.2