Main Roads Western Australia

Site

JJG Punch Materials Engineering Laboratory

Accreditation No.

1989

Site No.

2039

Date of Accreditation

29 Apr 1986

Address

5-9 Colin Jamieson Drive
Welshpool, WA 6106
Australia

mainroads.wa.gov.au

Contact

Mr Anthony Cullinane P: +61 (08) 93234739

Availability

Services conditionally available to external clients

Scope of Accreditation

JJG Punch Materials Engineering Laboratory

ISO/IEC 17025 (2017)

Calibration

  • The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
Service Product Determinant Technique Procedure Limitations
Dimensional metrology - Engineering equipment and precision instruments Dial gauges; Electronic length gauges; LVDTs Length measurements Comparison with a reference standard

Dial gauges, electronic length gauges

To comply with tolerances in AS 2103 (1978)

LVDTs

By the method of -

in-house WA 2250.1

Capability

with Calibration and Measurement Capability of -

1.0 μm from 0 mm up to 25 mm
1.5 μm above 25 mm up to 50 mm
On-site calibration only for LVDTs






Dimensional metrology - Engineering equipment and precision instruments Test sieves Length measurements Comparison with a reference standard

By the methods of -

AS 1152 (1993), BS 410-1, BS 410-2, ISO 3310-1, ISO 3310-2

Capability

with Calibration and Measurement Capability of -

2 μm from 0.038 mm to 125 mm


Force metrology - Force measuring and testing equipment Compression and universal machines in compression; Elastic force measuring devices; Load cells; Load measuring rings Force in compression Comparison measurement with reference load cell

Capability

Including on-site calibration

with Calibration and Measurement Capability of -

0.14% of reading from 0.15 kN to 2000 kN

Force metrology - Force measuring and testing equipment Falling weight deflectometer Load cell and sensor gains (deflection) Comparison with a reference standard

Capability

with Calibration and Measurement Capability of -
1 kN in the range 0-80 kN 
6μm for deflection and 8μm for curvature  in the range 0-2000μm
Mass - Determination of mass and calibration of weighing devices Laboratory weighing devices; Precision laboratory balances Mass Gravimetric measurement against reference mass

Capability

Including on site calibrations

with Calibration and Measurement Capability of -

4 in 106 or 40 μg (whichever is greater) from 1 mg to 1 kg;

1 in 10⁵ or 1 mg (whichever is greater) above 1 kg to 40 kg

Pressure metrology - Pressure and vacuum measuring equipment Digital pressure gauges; Mechanical pressure gauges; Pressure transducers Gauge pressure Comparison with reference instrument Self-indicating pressure transducers only

Capability

with Calibration and Measurement Capability of -

0.2 kPa from -95 kPa to 200 kPa

0.55 kPa above 200 kPa to 1000 kPa

Temperature metrology - Temperature measuring equipment Digital temperature measuring systems; Temperature loggers - Single and multi-channel Temperature Measurement against reference standard

By the methods of -

in-house method Main Roads WA 1130.3 and WA 2240.1

RTD or Thermistor type only


Capability

with Calibration and Measurement Capability of -

0.11°C from 0°C to 200 °C

Temperature metrology - Verification of controlled enclosures Baths; Ovens Spatial uniformity; Temperature Direct temperature measurement

Ovens

By the methods of -

Main Roads WA 2080.1, WA 2130.1 and AS 2853 (1986)

Baths

By the methods of -

Main Roads WA 1130.6 and AS 2853 (1986)

Capability

Including on-site calibration
with Calibration and Measurement Capability of -

0.3°C from 0°C to 190°C




Volume and density - Determination of density Standard nuclear gauge density blocks kg/m3 Gravimetric AS 1289.5.8.5,
MRWA Method WA 2040.1

Capability


with Calibration and Measurement Capability of -
0.3% of reading from 1366  kg/m3 to 3082 kg/m3   


Volume and density - Nuclear density gauges Nuclear moisture/density gauges Density Comparison with a reference standard AS 1289.5.8.4
MRWA Method WA135.1

Capability

Calibration of nuclear density gauges using Type C standard blocks

with Calibration and Measurement Capability of-

0.007 t/m3 density in the range 1.378 to 3.082 t/m3

The only data displayed is that deemed relevant and necessary for the clear description of the Activities and Services covered by the Scope of Accreditation (SoA).

Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.

The Product column may also describe an Item or Material.

The Limitations column may also describe the Range.