Scope of Accreditation

JST Biosafety

ISO/IEC 17025 (2017)

Infrastructure and Asset Integrity

Service Product Determinant Technique Procedure Limitations
Evaluation of air control equipment Biological safety cabinets Illuminance Not applicable AS 1807.15 (2000), AS 1807 clause 4.5, NSF ANSI 49 N5.9 for compliance with: AS 2252.1, AS 2252.2, NSF ANSI 49
Sound level Not applicable AS 1807.20 (2000), AS 1807 clause 4.7, NSF ANSI 49 N5.11 for compliance with: AS 2252.1, AS 2252.2, NSF ANSI 49
Integrity of HEPA filter installations Cold DOP AS 1807.6 (2000), AS 1807.7, AS 1807 clause 4.4, NSF ANSI 49 N5.5 for compliance with: AS 2252.1, AS 2252.2, NSF ANSI 49
Air flow; Air flow direction and visualization Not applicable NSF ANSI 49 N5.2, N5.3, N5.4 for compliance with: NSF ANSI 49
Ultraviolet intensity Not applicable AS 1807.23 (2000), AS 1807 clause 4.10 for compliance with: AS 2252.1, AS 2252.2
Air velocity and uniformity of air velocity Cold DOP AS 1807.1 (2000), AS 1807 clause 4.1, NSF ANSI 49 N5.2 for compliance with: AS 2252.1, AS 2252.2, NSF ANSI 49
Recovery times Not applicable AS 1807.24 (2000), AS/NZS 14644.3 Annex B4 for compliance with: AS 2252.1, AS 2252.2
Work zone integrity Not applicable AS 1807.5 (2000), AS 1807 clause 4.3 for compliance with: AS 2252.1, AS 2252.2
Particle count in work zone Automatic particle counter AS 1807.8 (2000), AS ISO 14644.1 for compliance with: AS 2252.1, AS 2252.2
Air barrier containment Cold DOP AS 1807.22 (2000), AS 1807 clause 4.9 for compliance with: AS 2252.1, AS 2252.2
Gas tightness Not applicable AS 1807.25 (2004), AS 1807 clause 4.11 for compliance with: AS 2252.1, AS 2252.2
Inward air velocity Cold DOP AS 1807.21 (2000), AS 1807 clause 4.8, NSF ANSI 49 N5.3 for compliance with: AS 2252.1, AS 2252.2, NSF ANSI 49
Clean rooms, pharmaceutical isolators and work stations Illuminance Not applicable AS 1807.15 (2000), AS 1807 clause 4.5 for compliance with: AS 2252.6 (Clean Workstation), AS/NZS 14644.3 (Clean Room), AS 4273
Inward air velocity Not applicable AS 1807.21 (2000), AS 1807 clause 4.8 for compliance with: AS 2252.6, AS ISO 14644.1 (Clean Room)
Air barrier containment Not applicable AS 1807.22 (2000), AS 1807 clause 4.9 for compliance with: AS 2252.6, AS/NZS 14644.3 (Clean Room), AS 4273
Sound level Not applicable AS 1807.7 (2000), AS 1807 clause 4.7 for compliance with: AS/NZS 14644.3 (Clean Room), AS 4273
Recovery times Not applicable AS 1807.24 (2000), AS/NZS 14644.3 Annex B4 for compliance with: AS 2252.6, AS/NZS 14664.3
Air velocity and uniformity of air velocity Cold DOP AS 1807.3 (2000), AS ISO 14644.3 clause 4.2.2, 4.2.3, Methods B.2 for compliance with: AS 2252.6 (Clean Workstation), AS/NZS 14644.3 (Clean Room), AS 4273
Air velocity and uniformity of air velocity Cold DOP AS 1807.1 (2000), AS 1807 clause 4.1 for compliance with: AS 2252.6 (Clean Workstation), AS/NZS 14644.3 (Clean Room), AS 4273
Ultraviolet intensity Not applicable AS 1807.23 (2000), AS 1807 clause 4.10 for compliance with: AS 2252.6 (Clean Workstation), AS/NZS 14644.3 (Clean Room), AS 4273
Work zone integrity Cold DOP AS 1807.5 (2000), AS 1807 clause 4.3 for compliance with: AS 2252.6 (Clean Workstation), AS/NZS 14644.3 (Clean Room), AS 4273
Integrity of HEPA filter installations Cold DOP AS 1807.6 (2000), AS 1807.7, AS 1807 clause 4.4, AS/NZS 14644.3 Clause 4.2.7, Method B7 for compliance with: AS 2252.1, AS 2252.2, AS 2252.6 (Clean Workstation), AS/NZS 14644.3 (Clean Room), AS 4273
Sound level Not applicable AS 1807.16 (2000) for compliance with: AS 2252.6 (Clean Workstation), AS/NZS 14644.3 (Clean Room)
Air pressure Not applicable AS 1807.10 (2000), AS/NZS 14644.3 Annex B1 for compliance with: AS 2252.6 (Clean Workstation), AS/NZS 14644.3 (Clean Room), AS 4273
Particle count in work zone Automatic particle counter AS 1807.8 (2000), AS ISO 14644.1 Appendix A for compliance with: AS 2252.6 (Clean Workstation), AS ISO 14644.1 (Clean Room), AS 4273
Cytotoxic drug safety cabinets Air pressure Not applicable AS 1807.10 (2000), AS/NZS 14644.3 Annex B1 for compliance with: AS 2252.5

Integrity of HEPA filter installations Cold DOP AS 1807.6 (2000), AS 1807.7, AS 1807 clause 4.4 for compliance with: AS 2252.1, AS 2252.2, AS 2252.5

Air barrier containment Cold DOP AS 1807.22 (2000), AS 1807 clause 4.9 for compliance with: AS 2252.5

Sound level Not applicable AS 1807.20 (2000), AS 1807 clause 4.7 for compliance with: AS 2252.5

Sound level Not applicable AS 1807.16 (2000) for compliance with: AS 2252.5

Gas tightness Not applicable AS 1807.25 (2004), AS 1807 clause 4.11 for compliance with: AS 2252.5

Particle count in work zone Automatic particle counter AS 1807.8 (2000),
AS ISO 14644.1
for compliance with: AS 2252.5

Illuminance Not applicable AS 1807.15 (2000), AS 1807 clause 4.5 for compliance with: AS 2252.5

Ultraviolet intensity Not applicable AS 1807.23 (2000), AS 1807 clause 4.10 for compliance with: AS 2252.5

Work zone integrity Not applicable AS 1807.5 (2000), AS 1807 clause 4.3 for compliance with: AS 2252.5

Air velocity and uniformity of air velocity Cold DOP AS 1807.1 (2000), AS 1807 clause 4.1 for compliance with: AS 2252.5

Fume cupboards and recirculating fume cabinets Extraction performance Smoke visualisation AS/NZS 2243.8 Appendix A
AS/NZS 2243.9 Appendix A
for compliance with: AS/NZS 2243.8, AS/NZS 2243.9
Illuminance Not applicable AS 1807.15 (2000), AS 1807 clause 4.5 for compliance with: AS/NZS 2243.8, AS/NZS 2243.9
Extraction performance Face velocity AS/NZS 2243.8 Appendix B
AS/NZS 2243.9 Appendix B
for compliance with: AS/NZS 2243.8, AS/NZS 2243.9
Sound level Not applicable AS 1807.20 (2000), AS 1807 clause 4.7 for compliance with: AS/NZS 2243.8, AS/NZS 2243.9