PLUS ES

Site

Measurement Services

Accreditation No.

62

Site No.

710

Date of Accreditation

14 Jul 1952

Address

Ausgrid Building, 48 - 50 Holker Street
Silverwater, NSW 2128
Australia

pluses.com.au

Contact

Mr Scott Murray P: +61 (02) 87451600

Availability

Services available to external clients

Scope of Accreditation

Measurement Services

ISO/IEC 17025 (2017)

Calibration

Service Product Determinant Technique Procedure Limitations
DC and low frequency electrical metrology - Electrical instrument calibrators Instrument calibrators AC current; AC voltage; Capacitance; DC current; DC voltage; Phase angle; Resistance Comparison with a reference standard; Comparison with a reference standard by differential measurement; Comparison with a reference standard by null measurement; Indirect measurement

Capability

DC Voltage
with a Calibration and Measurement Capability of -
5 μV/V + 200 nV from 0 V to 1100 V

AC Voltage
10 Hz to 40 Hz
with a Calibration and Measurement Capability of -
800 μV/V from 2 to 10 mV
300 μV/V from 10 to 100 mV
150 μV/V from 100 to 600 mV
220 μV/V from 600 mV to 300 V at 10 Hz
300 μV/V from 300 to 1 000 V at 10 Hz
70 μV/V from 600 mV to 300 V and 20 Hz to 40 Hz
100 μV/V from 300 to 1 000 V and 20 Hz to 40 Hz

40 Hz to 30 kHz
with a Calibration and Measurement Capability of -
600 μV/V from 2 to 10 mV
220 μV/V from 10 to 20 mV
120 μV/V from 20 to 100 mV
80 μV/V from 100 to 200 mV
50 μV/V from 200 to 600 mV
20 μV/V from 600 mV to 20 V
25 μV/V from 20 to 300 V
40 μV/V from 300 to 1 000 V and 40 Hz to 20 kHz

30 kHz to 100 kHz
with a Calibration and Measurement Capability of -
1100 μV/V from 2 to 10 mV
500 μV/V from 10 to 20 mV
300 μV/V from 20 to 100 mV
200 μV/V from 100 to 200 mV
100 μV/V from 200 to 600 mV
40 μV/V from 600 mV to 20 V
40 μV/V from 20 to 300 V
120 μV/V from 300 to 1 000 V and 20 kHz to 100 kHz

100 kHz to 1 MHz
with a Calibration and Measurement Capability of -
2000 μV/V from 2 to 10 mV
1000 μV/V from 10 to 100 mV
400 μV/V from 100 to 600 mV
250 μV/V from 600 mV to 20 V
200 μV/V from 20 to 60 V and 100 kHz to 500 kHz

DC Current
with a Calibration and Measurement Capability of-
10 μA/A from 10 μA to 10 mA
15 μA/A from 10 mA to 1 A
30 μA/A from 1 A to 10 A
70 μA/A from 10 A to 100 A

AC Current
10 Hz to 50 Hz
with a Calibration and Measurement Capability of -
100 μA/A from 30 μA to 100 μA at 50 Hz
100 μA/A from 100 μA to 10 mA
50 μA/A from 10 mA to 1 A
90 μA/A from 1 A to 10 A
100 μA/A from 10 A to 20 A
130 μA/A from 20 A to 100 A at 50 Hz

50 Hz to 1 kHz
with a Calibration and Measurement Capability of -
70 μA/A from 100 μA to 10 mA
45 μA/A from 10 mA to 100 mA
50 μA/A from 100 mA to 1 A
90 μA/A from 1 A to 10 A
100 μA/A from 10 A to 20 A

1 kHz to 5 kHz
with a Calibration and Measurement Capability of -
100 μA/A from 100 μA to 10 mA
50 μA/A from 10 mA to 1 A
90 μA/A from 1 A to 10 A
200 μA/A from 10 A to 20 A

5 kHz to 10 kHz
with a Calibration and Measurement Capability of -
150 μA/A from 100 μA to 10 mA
50 μA/A from 10 mA to 100 mA
60 μA/A from 100 mA to 1 A
90 μA/A from 1 A to 10 A

10 kHz to 30 kHz
with a Calibration and Measurement Capability of -
80 μA/A from 10 mA to 100 mA
100 μA/A from 100 mA to 1 A
200 μA/A from 1 A to 10 A

DC Resistance
with a Calibration and Measurement Capability of -
10 nΩ at 0 Ω
25 μΩ/Ω + 1 nΩ from 10 μΩ to 1 mΩ
10 μΩ/Ω from 1 mΩ to 0.1 Ω
3 μΩ/Ω from 0.1 Ω to 1 Ω
1 μΩ/Ω from 1 Ω to 100 kΩ
2 μΩ/Ω from 100 kΩ to 1 MΩ
3.5 μΩ/Ω from 1 MΩ to 10 MΩ
12 μΩ/Ω from 10 MΩ to 100 MΩ
35 μΩ/Ω from 100 MΩ to 1 GΩ
0.1% above 1 GΩ to 100 GΩ

Capacitance
at 50 Hz
with a Calibration and Measurement Capability of -
0.11% from 70 µF to 300 μF

at 100 Hz
with a Calibration and Measurement Capability of -
0.065% from 300 nF to 70 µF
0.11% from 70 µF to 300 μF
1% from 300 μF to 1.1 mF

at 1 000 Hz
with a Calibration and Measurement Capability of -
0.1% + 1.1 pF from 200 pF to 1 nF
0.1% + 0.6 pF from 1 nF to 7 nF
0.065% from 7 nF to 300 nF

DC Ramp
with a Calibration and Measurement Capability of -
0.06% from 200 µF to 110 mF

Phase Angle
Voltage to Voltage (10 mV - 500 V)
with a Calibration and Measurement Capability of -
0.03° for 50 Hz to 500 Hz
0.06° for 500 Hz to 2 kHz
0.09° for 2 kHz to 5 kHz
0.15° for 5 kHz to 10 kHz

Voltage to Current (10 mV - 500 V and 100 mA – 20 A)
with a Calibration and Measurement Capability of -
0.03° for 50 Hz to 500 Hz
0.06° for 500 Hz to 2 kHz
0.16° for 2 kHz to 5 kHz
0.22° for 5 kHz to 10 kHz
DC and low frequency electrical metrology - Electrical measurement and test equipment Ammeters; Appliance testers; Capacitance meters; Current clamps; Data recorders; Digital multimeters (DMM); Electrical energy meters; Electricity meters; Galvanometers and null detectors; Insulation resistance test equipment; LCR meters; Ohm meters; Phase angle meters; Power analysers; Power factor meters; Power supplies; Residual current circuit breaker testers; Residual current devices (RCD) testers; Varmeters; Voltmeters; Watt meters AC current; AC voltage; Capacitance; DC current; DC voltage; Resistance Comparison with a reference standard; Comparison with a reference standard by differential measurement

Capability

DC voltage
with a Calibration and Measurement Capability of -
5 μV/V + 200 nV from 0 V to 1100 V

AC current
10 Hz to 50 Hz
with a Calibration and Measurement Capability of -
100 μA/A from 30 μA to 100 μA at 50 Hz
100 μA/A from 100 μA to 10 mA
50 μA/A from 10 mA to 1 A
90 μA/A from 1 A to 10 A
100 μA/A from 10 A to 20 A
130 μA/A from 20 A to 100 A at 50 Hz

50 Hz to 1 kHz
with a Calibration and Measurement Capability of -
300 μA/A + 20 nA from 30 μA to 100 μA
70 μA/A from 100 μA to 10 mA
45 μA/A from 10 mA to 100 mA
50 μA/A from 100 mA to 1 A
90 μA/A from 1 A to 10 A
100 μA/A from 10 A to 20 A
800 μA/A from 20 A to 100 A

1 kHz to 5 kHz
with a Calibration and Measurement Capability of -
100 μA/A from 100 μA to 10 mA
50 μA/A from 10 mA to 1 A
90 μA/A from 1 A to 10 A
200 μA/A from 10 A to 20 A

5 kHz to 10 kHz
with a Calibration and Measurement Capability of -
150 μA/A from 100 μA to 10 mA
50 μA/A from 10 mA to 100 mA
60 μA/A from 100 mA to 1 A
90 μA/A from 1 A to 10 A
0.4% + 200 μA from 10 A to 11 A

10 kHz to 30 kHz
with a Calibration and Measurement Capability of -
80 μA/A from 10 mA to 100 mA
100 μA/A from 100 mA to 1 A
200 μA/A from 1 A to 10 A

Multi-turn technique
with a Calibration and Measurement Capability of -
0.15% up to 100 A at 50 Hz
0.35% from 100 A to 1 000 A at 50 Hz
0.6% from 1 000 A to 3 000 A at 50 Hz

DC current
with a Calibration and Measurement Capability of -
10 μA/A from 10 μA to 10 mA
15 μA/A from 10 mA to 1 A
30 μA/A from 1 A to 10 A
70 μA/A from 10 A to 100 A
200 μA/A from 100 A to 500 A

AC voltage
10 Hz to 40 Hz
with a Calibration and Measurement Capability of -
800 μV/V from 2 to 10 mV
300 μV/V from 10 to 100 mV
150 μV/V from 100 to 600 mV
220 μV/V from 600 mV to 300 V at 10 Hz
300 μV/V from 300 to 1 000 V at 10 Hz
70 μV/V from 600 mV to 300 V and 20 Hz to 40 Hz
100 μV/V from 300 to 1 000 V and 20 Hz to 40 Hz

40 Hz to 30 kHz
with a Calibration and Measurement Capability of -
320 μV/V from 2 to 10 mV
220 μV/V from 10 to 20 mV
120 μV/V from 20 to 100 mV
80 μV/V from 100 to 200 mV
50 μV/V from 200 to 600 mV
20 μV/V from 600 mV to 20 V
25 μV/V from 20 to 300 V
40 μV/V from 300 to 1 000 V and 40 Hz to 20 kHz

30 kHz to 100 kHz
with a Calibration and Measurement Capability of -
1 100 μV/V from 2 to 10 mV
500 μV/V from 10 to 20 mV
300 μV/V from 20 to 100 mV
200 μV/V from 100 to 200 mV
100 μV/V from 200 to 600 mV
40 μV/V from 600 mV to 20 V
40 μV/V from 20 to 300 V
120 μV/V from 300 to 1 000 V and 20 kHz to 100 kHz

100 kHz to 1 MHz
with a Calibration and Measurement Capability of -
2 000 μV/V from 2 to 10 mV
1 000 μV/V from 10 to 100 mV
400 μV/V from 100 to 600 mV
250 μV/V from 600 mV to 20 V
200 μV/V from 20 to 60 V and 100 kHz to 500 kHz

Capacitance
at 1 000 Hz

with a Calibration and Measurement Capability of -
(0.01% + 1 pF) from 1 pF to 1 μF
0.05% from 1.1 to 11 μF
0.1% from 11 to 100 μF

at 100 Hz
with a Calibration and Measurement Capability of -
0.2% from 300 nF to 300 μF
1% from 300 μF to 1.1 mF

with a Calibration and Measurement Capability of -
0.3% from 100 pF to 100 μF
1% from 10 pF to 100 pF and 100 μF to 1.1 mF

Resistance
with a Calibration and Measurement Capability of -
10 nΩ at 0 Ω
150 μΩ/Ω from 10 μΩ to 100 µΩ
40 μΩ/Ω from 100 µΩ to 1 mΩ
20 μΩ/Ω from 1 mΩ to 10 mΩ
15 μΩ/Ω from 10 mΩ to 100 mΩ
4 μΩ/Ω from 100 mΩ to 1 Ω
3 μΩ/Ω from 1 Ω to 10 Ω
2 μΩ/Ω from 10 Ω to 100 kΩ
2.5 μΩ/Ω from 100 kΩ to 1 MΩ
5 μΩ/Ω from 1 MΩ to 10 MΩ
20 μΩ/Ω from 10 MΩ to 100 MΩ
100 μΩ/Ω from 100 MΩ to 1 GΩ
1000 μΩ/Ω from 1 GΩ to 100 GΩ

Insulation Resistance Test Equipment
0.5% from 100 kΩ to 100 GΩ up to 10 kV
Appliance testers; Capacitance meters; Current clamps; Digital multimeters (DMM); Insulation resistance test equipment; LCR meters; Ohm meters; Phase angle meters; Power factor meters; Residual current circuit breaker testers; Residual current devices (RCD) testers Inductance Comparison with a reference standard

Capability

with a Calibration and Measurement Capability of -
0.1% or 1 μH (whichever is greater) from 100 μH to 100 H
Appliance testers; Current clamps; Data recorders; Electrical energy meters; Electricity meters; Ohm meters; Phase angle meters; Power analysers; Power factor meters; Power supplies; Varmeters; Watt meters Active power Comparison with a reference standard

Capability

with a Calibration and Measurement Capability of -
50 μW/W up to 1 000 V and 20 A on d.c.
200 μW/VA from 1 V to 1 000 V and 0.001 A to 120 A at 50 Hz

Multi-turn technique for 0.5 lead to 0.5 lagging power factors
with a Calibration and Measurement Capability of -
0.2% from 20 A to 100 A at 50 Hz
0.3% from 100 A to 1 000 A at 50 Hz
0.6% up from 1 000 A to 3 000 A at 50 Hz
Capacitance meters; Current clamps; Digital multimeters (DMM); Electrical energy meters; Electricity meters; LCR meters; Phase angle meters; Power analysers; Power factor meters; Varmeters; Watt meters Phase angle Indirect measurement against a reference standard

Capability

with a Calibration and Measurement Capability of -
0.02º in the range 50 to 500 Hz
0.05º in the range 500 Hz to 10 kHz
Power factor Comparison with a reference standard; Indirect measurement against a reference standard

Capability

with a Calibration and Measurement Capability of -
0.001 power factor from 50 Hz to 10 kHz
Electrical energy meters; Electricity meters; LCR meters; Phase angle meters; Power analysers; Power factor meters; Varmeters; Watt meters Active energy; Reactive energy Comparison with a reference standard

Capability

Watthour meters
with a Calibration and Measurement Capability of -
250 μWh/VAh from 0.01 A to 0.02 A and 63.5 V to 415 V at 50 Hz
100 μWh/VAh from 0.02 A to 100 A and 63.5 V to 415 V at 50 Hz

Varhour meters
with a Calibration and Measurement Capability of -
320 μvarh/VAh from 0.05 A to 100 A and 63.5 V to 415 V at 50 Hz
Electrical energy meters; Electricity meters; Phase angle meters; Power analysers; Power factor meters; Varmeters; Watt meters Reactive power Comparison with a reference standard

Capability

with a Calibration and Measurement Capability of -
0.05% at zero power factor
DC and low frequency electrical metrology - Electrical standards AC bridges Capacitance; Resistance; Voltage ratio Comparison with a reference standard by differential measurement

Capability

DC Resistance
with a Calibration and Measurement Capability of -
150 μΩ/Ω from 10 μΩ to 100 µΩ
40 μΩ/Ω from 100 µΩ to 1 mΩ
20 μΩ/Ω from 1 mΩ to 10 mΩ
15 μΩ/Ω from 10 mΩ to 100 mΩ
3.5 μΩ/Ω from 100 mΩ to 1 Ω
2 μΩ/Ω from 1 Ω to 100 kΩ
2.5 μΩ/Ω from 100 kΩ to 1 MΩ
8 μΩ/Ω from 1 MΩ to 10 MΩ
30 μΩ/Ω from 10 MΩ to 100 MΩ
100 μΩ/Ω from 100 MΩ to 1 GΩ
1000 μΩ/Ω from 1 GΩ to 100 GΩ

Capacitance
at 100 Hz
with a Calibration and Measurement Capability of -
0.2% from 300 nF to 300 μF
1% from 300 μF to 1.1 mF

at 1 000 Hz
with a Calibration and Measurement Capability of -
(0.01% + 1 pF) from 1 pF to 1 μF
0.05% from 1.1 to 11 μF
0.1% from 11 to 100 μF

Voltage Ratio
with a Calibration and Measurement Capability of -
0.03%
Conductance boxes; Precision resistors; Resistance boxes Resistance Comparison with a reference standard; Comparison with a reference standard by null measurement

Capability

with a Calibration and Measurement Capability of -
25 μΩ/Ω + 1 nΩ from 10 μΩ to 1 mΩ
10 μΩ/Ω from 1 mΩ to 0.1 Ω
3 μΩ/Ω from 0.1 Ω to 1 Ω
1 μΩ/Ω from 1 Ω to 100 kΩ
2 μΩ/Ω from 100 kΩ to 1 MΩ
3.5 μΩ/Ω from 1 MΩ to 10 MΩ
12 μΩ/Ω from 10 MΩ to 100 MΩ
35 μΩ/Ω from 100 MΩ to 1 GΩ
0.1% above 1 GΩ to 100 GΩ
Current shunts Current AC-DC difference; Resistance Comparison with a reference standard by null measurement for loads up to 500 A

Capability

Resistance
with a Calibration and Measurement Capability of -
25 μΩ/Ω + 1 nΩ from 10 μΩ to 1 mΩ
10 μΩ/Ω from 1 mΩ to 0.1 Ω
3 μΩ/Ω from 0.1 Ω to 10 kΩ

AC-DC difference
with a Calibration and Measurement Capability of -
50 μA/A from 10 mA to 20 A and 50 Hz to 1 kHz
150 μA/A from 10 mA to 20 A and 1 kHz to 10 kHz
1 000 μA/A from 20 A to 100 A and 50 Hz to 1 kHz
DC bridges Voltage ratio Direct measurement

Capability

with a Calibration and Measurement Capability of -
0.005%
DC potentiometers Voltage Direct measurement

Capability

with a Calibration and Measurement Capability of -
5 μV/V or 0.01 μV (whichever is greater) up to 1.6 V
Inductors Inductance Direct measurement Self inductors

Capability

with a Calibration and Measurement Capability of -
0.1% or 1 μH (whichever is greater) from 100 μH to 100 H at 1 000 Hz
Potential dividers; Voltage ratio boxes Voltage ratio Comparison with a reference standard by differential measurement

Capability

with a Calibration and Measurement Capability of -
1 μV/V to 5 μV/V up to 1 100 V d.c.
Precision capacitors Capacitance Comparison with a reference standard by resonance measurement

Capability

Two terminal capacitors at 1 000 Hz
with a Calibration and Measurement Capability of -
(0.01% + 1 pF) from 1 pF to 1 μF
0.05% from 1.1 to 11 μF
0.1% from 11 to 100
0.25% from 100 μF to 400 μF
1% from 400 μF to 1.1 mF

Two terminal capacitors at 100 Hz
With a Calibration and Measurement Capability of -
0.2% from 300 nF to 300 μF
1% from 300 μF to 1.1 mF

Three terminal capacitors at 1 000 Hz
with a Calibration and Measurement Capability of -
(0.01% + 0.03 fF) from 0.1 pF to 10 pF
0.01% from 10 pF to 1.1 μF
0.05% from 1.1 μF to 11 μF


Ratiometers Voltage ratio Comparison with a reference standard

Capability

with a Calibration and Measurement Capability of -
0.03%
Voltage standards - E.M.F. reference devices Voltage Comparison with a reference standard by differential measurement

Capability

with a Calibration and Measurement Capability of -
0.5 μV/V at 10 V
DC and low frequency electrical metrology - High-voltage/high-current standards and equipment Current transformers Current error; Phase displacement Comparison with a reference standard by differential measurement AS 1675-1986 Withdrawn
AS 60044.1-2007 Superseded
IEC 60044.1-2003  Superseded
IEC 61869-1,-2
 AS 61869-1, -2

Capability

Accuracy tests at 50 Hz, up to 5000 A applied primary current;

for CTs with rated secondaries of 1 A or 5 A;

with Calibration and Measurement Capability of -;

0.025% for current error;

0.030 crad for phase displacement;
High voltage dividers Voltage ratio Comparison with a reference standard

Capability

with a Calibration and Measurement Capability of -
0.5% from 1 kV to 100 kV at 50 Hz (AC voltage)
0.5% from 1 kV to 100 kV (DC voltage)
High voltage dividers; High voltage test sets Current Comparison with a reference standard

Capability

with a Calibration and Measurement Capability of -
0.3% + 60 nA from 30 μA to 200 mA at 50 Hz (AC Current)
0.2% + 50 nA from 10 μA to 200 mA (DC Current)
Voltage transformers Phase displacement; Voltage error Comparison with a reference standard by differential measurement AS 1243-1982 Superseded
AS 60044.2-2007 Superseded
IEC 60044.2-2003 Superseded
IEC 61869-1,-3
AS 61869-1,-3
PSS Method TM02 (1:1 ratio method)

Capability

Accuracy tests and calibration of two-winding VTs with windings of 1:1 ratio, at rated voltages of up to 110 V, at 50 Hz
With Calibration and Measurement Capability of:
±0.007% for voltage error ;
±0.009 crad for phase displacement

Accuracy tests at 50 Hz, up to rated primary voltage of 132 kV/√3;

Single phase measurements of VTs with rated secondaries of 63.5 V or 110 V;

with Calibration and Measurement Capability of -;

0.025% for voltage error;

0.025 crad for phase displacement;

DC and low frequency electrical metrology - High-voltage/high-current standards and equipmentIncluding onsite calibrations High voltage dividers; High voltage test sets Voltage Comparison with a reference standard

Capability

Measurements performed in the laboratory
with a Calibration and Measurement Capability of -
0.5% from 1 kV to 100 kV at 50 Hz (AC voltage)
0.5% from 1 kV to 100 kV (DC voltage)

Measurements performed in the field;
with Calibration and Measurement Capability of;
a.c. test sets
2.5% of reading from 1 kVrms to 100 kVrms at 50 Hz;
2.5% of reading from ±1 kVpeak/√2 to ±100 kVpeak/√2, at 50 Hz;
2.5% of reading from ±1 kVpeak to ±140 kVpeak, at 50 Hz;
 d.c. test sets
2.5% of reading from -90 kV to 90 kV
Temperature metrology - Ancillary temperature measuring equipment Digital voltmeters (DVM) and digital multimeters (DMM); Multi-channel thermocouple data recorders; Process calibrators Resistance to temperature conversion; Volt to temperature conversion Direct measurement by electrical input

Capability

Including portable temperature measuring instruments
by Electrical Simulation

Type T Thermocouple
with a Calibration and Measurement Capability of -
0.7 ºC from -250 ºC to -150 ºC
0.3 ºC from -150 ºC to 0 ºC
0.2 ºC from 0 ºC to 40 0ºC

Type K Thermocouple
with a Calibration and Measurement Capability of -
0.4 ºC from -200 ºC to -100 ºC
0.2 ºC from -100 ºC to 120 ºC
0.3 ºC from 120 ºC to 1000 ºC
0.5 ºC from 1000 ºC to 1372 ºC

RTD Pt 385, 100Ω
Measure Mode
with a Calibration and Measurement Capability of -
0.1 ºC from -200 ºC to 0 ºC
0.2 ºC from 0 ºC to 630 ºC
0.3 ºC from 630 ºC to 800 ºC
Source Mode
with a Calibration and Measurement Capability of -
0.1 ºC from -200 ºC to 800 ºC
Temperature metrology - Temperature measuring equipment Digital temperature measuring systems; Temperature loggers - Single and multi-channel Temperature Measurement against reference standard

Capability

with a Calibration and Measurement capability of -
0.03 ºC at 0 ºC and 20 ºC
0.4 ºC + 0.1% T from 50 ºC to 155 ºC
Time and frequency metrology - Frequency and time standards Frequency standards; Rubidium standard; Signal sources Frequency; Time interval; Voltage Measurement against reference standard

Capability

Frequency
with a Calibration and Measurement Capability of -
1 in 1010 + 1 count from 0.01 Hz to 1.1 GHz

Time interval
with a Calibration and Measurement Capability of -
1 in 1010 + 500 ps from 4 ns  to 1 in 103 s

Period
with a Calibration and Measurement Capability of -
1 in 1010 + 1 count from 1 ns to 100 s

Rise Time
With a Calibration and Measurement Capability of –
 20 ps

Voltage – Level Flatness
Measurement of the difference between the same nominal peak to peak voltage at test frequency and the (same nominal) peak to peak voltage at reference frequency (50 kHz) divided by the peak to peak voltage at reference frequency e.g. Flatness(%)=(Vtf-Vref)/Vref*100%
with a Calibration and Measurement Capability of –
0.81% of relative voltage reading from 5 mVpk-pk to 5.5 Vpk-pk and 50 kHz to 550 MHz and
1.0% of relative voltage reading from 5 mVpk-pk to 5.5 Vpk-pk and 550 MHz to 1.1 GHz
Time and frequency metrology - Frequency, time and waveform measuring equipment Clocks and timers; Counters; Distortion analysers; Frequency analysers; Frequency meters; Oscilloscopes; Time interval meters; Waveform measuring instruments Frequency Measurement against reference standard

Capability

with a Calibration and Measurement Capability of -
1 in 1010 + 1 count from 0.01 Hz to 1.1 GHz

Oscilloscopes Frequency characteristics
with a Calibration and Measurement capability of -
5% of frequency up to 300 MHz (bandwidth) at voltages of 5 mV to 5.5 Vp-p
Counters; Distortion analysers; Frequency analysers; Frequency meters; Oscilloscopes; Time interval meters Time interval Measurement against reference standard

Capability

with a Calibration and Measurement Capability of -
1 in 1010 + 500 ps from 4 ns to 1 in 103 s

Oscilloscopes Timing characteristics
with a Calibration and Measurement capability of -
0.2% from 10 ns to 10 s
Counters; Distortion analysers; Frequency analysers; Frequency meters; Oscilloscopes; Time interval meters; Waveform measuring instruments Voltage Comparison with a reference standard

Capability

Oscilloscopes input characteristics
with a Calibration and Measurement capability of -
0.5% from 1 mV to 130 Vp-p

Voltage Swells & Sags
with a Calibration and Measurement capability of -
0.1% + 1 digit of reading from 1 V to 1000 V and 45 Hz to 65 Hz

Oscillatory Transient - comparison purposes only
Fast Transient 1.2 / 50 μs comparison purposes only
Distortion analysers; Frequency analysers; Waveform measuring instruments Current harmonics; Distortion; Voltage flicker; Voltage harmonics Measurement against reference standard

Capability


Current Harmonics
with a Calibration and Measurement Capability of -
(with individual harmonic amplitudes from 5 µA)
0.1% of fundamental up to the 99ᵗʰ from 0.01 A to 120 A and 45 Hz to 65 Hz

Multi-turn technique
with a Calibration and Measurement Capability of –
(with individual harmonic amplitudes from 150 mA)
0.1% of fundamental up to the 8ᵗʰ from 120 A to 500 A and 45 Hz to 65 Hz
(with individual harmonic amplitudes from 850 mA)
 0.1% of fundamental up to the 8ᵗʰ from 500 A to 3000 A and 45 Hz to 65 Hz

Voltage Harmonics
with a Calibration and Measurement Capability of -
(with individual harmonic amplitudes from 2 mV)
0.1% of fundamental up to the 99th from 10 V to 1 000 V and 45 to 65 Hz

Total Harmonic Distortion
up to 99 sup th harmonic from 10 V to 1 000 V, 0.01 A to 3 000 A and between 0.01% to 40%
with a Calibration and Measurement Capability of -
0.05% of fundamental for THD from 0.01% to 0.3%
0.1% of fundamental for THD from 0.3% to 40%

Voltage Flicker
from 10 V to 1 000 V and 47 Hz to 53 Hz
with a Calibration and Measurement Capability of -
0.1% of fundamental

ISO/IEC 17025 (2017)

Infrastructure and Asset Integrity

Service Product Determinant Technique Procedure Limitations
Evaluation of high voltage/power equipment and components - Performance Bushings; Cable feeders; Cables; Clothing materials, gloves and tools for use in high voltage/power applications; Insulated systems; Insulation products; Insulator and conductor fittings; Insulators; Three phase cable installations; Transformers and inductors Insulation resistance Insulation resistance

By the method of PSS WI-10

Tests performed in the laboratory and in the field;

From 1 MΩ to 100 GΩ at voltages of 50 Vdc to 5,000 Vdc

Clothing materials, gloves and tools for use in high voltage/power applicationsHigh voltage operating sticks AC voltage; Applied high voltage withstand; Electrical performance of cables and components; Workplace electrical safety High voltage; Surface leakage

 ISSC14 Guide to electrical workers' safety equipment chapter 13;

 PSS WI-12

At 50 Hz up to 45 kV

       

Transformers and inductors Accuracy tests Phase displacement; Voltage error IEC 60044-2:2003 Superseded;
AS 60044.2: 2007 Superseded;
AS 1243:1982 Superseded;
IEC 61869-1; and IEC 61869-3; Instrument transformers - Voltage transformers
AS 61869-1,-3
PSS Method TM02 (1:1 ratio method)

Accuracy tests at 50 Hz up to rated primary voltage of 132 kV/√3;

Single phase measurements of voltage transformers with rated secondaries of 63.5 V or 110 V 
Accuracy tests of two-winding VTs with windings of 1:1 ratio, at rated voltages of up to 110 V, at 50 Hz
Accuracy tests Current error; Phase displacement IEC 60044-1:2003 Supersded;
AS 60044.1: 2007 Superseded;
Instrument Transformers-Current Transformers;
IEC 61869-1 and IEC 61869-1 -2: Instrument Transformers-Current Transformers;
AS 61869-1,-2
 AS 1675: 1986 Withdrawn Instrument Transformers-Measurement & Protection

Accuracy tests at 50 Hz up to 5000 A applied primary current

for current transformers with rated secondaries of 1 A or 5 A

The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.

Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.