National Association of Testing Authorities, Australia
Scope of Accreditation
Trescal (Australia) Pty Ltd
Site
New South Wales Laboratory
Accreditation No.
19397
Site No.
22182
Date of Accreditation
22 Jan 2014
Contact
Availability
Services available to external clients
Scope of Accreditation
ISO/IEC 17025 (2017)
Calibration
- The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
Service | Product | Determinant | Technique | Procedure | Limitations | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
DC and low frequency electrical metrology - Electrical instrument calibrators | Instrument calibrators | Resistance | Direct measurement against a reference standard |
0 Ω to 100 MΩ |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability with Calibration and Measurement Capability of - 57 μΩ/Ω + 50 μΩ up to 1.9 Ω |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
AC voltage; DC voltage | Direct measurement against a reference standard |
D.C. voltage From 0 mV to 1000 V A.C. voltage From 10 mV to 700 V at 10 Hz to 1 MHz |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability with Calibration and Measurement Capability of - D.C. voltage 14 μV/V + 0.3 μV up to 100 mV 8 μV/V from 100 mV to 1 V A.C. voltage 0.04% from 10 mV to 100 mV and 10 Hz to 1 kHz |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
AC current; DC current | Direct measurement against a reference standard |
D.C. current From 0 nA to 20 A A.C.current From 10 μA to 10 A at 1 kHz From 10 A to 20 A at 50 Hz to 5 kHz |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability with Calibration and Measurement Capability of - D.C. current 770 μA/A + 5 pA up to 100 nA A.C. current 0.13% from 10 μA to 100 μA at 1 kHz |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
DC and low frequency electrical metrology - Electrical measurement and test equipment | Ammeters; Digital multimeters (DMM) | AC current; DC current | Direct measurement against a reference standard |
D.C. current From 0 μA to 1500 A A.C.current From 20 μA to 10 A at 10 Hz to 10 kHz From 10 A to 20 A at 40 Hz to 1 kHz From 20 A to 1500 A at 56 Hz |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability with Calibration and Measurement Capability of - D.C. current 0.02% + 10 nA up to 220 μA A.C. current 0.04% from 20 μA to 220 μA and 10 Hz to 1 kHz |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Digital multimeters (DMM); Ohm meters | Resistance | Direct measurement against a reference standard |
From 0 Ω to 1 GΩ |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability with Calibration and Measurement Capability of - 0.013% + 50 μΩ up to 1 Ω |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Digital multimeters (DMM); Voltmeters | AC voltage; DC voltage | Direct measurement against a reference standard |
D.C. voltage From 0 mV to 1000 V From 2000 V to 10000 V A.C. voltage From 1 mV to 1000V at 10 Hz to 1 MHz From 2000 V at 5000 V at 50 Hz |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability with Calibration and Measurement Capability of - D.C. voltage 21 μV/V + 0.8 μV up to 220 mV A.C. voltage 0.31% from 1 mV to 2.2 mV and 40 Hz to 20 kHz |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Insulation resistance test equipment | Resistance | Direct measurement against a reference standard | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability with Calibration and Measurement Capability of -0.3% of reading from 10 kΩ to 39.99 kΩ up to 55 Vdc 0.3% of reading from 40 kΩ to 99.99 kΩ up to 400 Vdc 0.4% of reading from 100 kΩ to 199.99 kΩ up to 800 Vdc 0.6% of reading from 200 kΩ to 999.99 kΩ up to 1000 Vdc 0.7% of reading from 1 MΩ to 1.999 MΩ up to 1000 Vdc 0.8% of reading from 2 MΩ to 9.999 MΩ up to 1000 Vdc 1.2% of reading from 10 MΩ to 199.9 MΩ up to 1000 Vdc 1.3% of reading from 200 MΩ to 999.9 MΩ up to 1000 Vdc 1.8% of reading at 1 GΩ up to 1000 Vdc 1.9% of reading at 1 GΩ using Multiplier up to 1000 Vdc |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
LCR meters | Capacitance; Inductance; Resistance | Comparison with a reference standard by substitution measurement |
Two terminal capacitors 0.7 μF to 100 μF at 100 Hz 0.4 nF to 300 nF at 1 kHz DC Resistance and Inductance 0.1 mH to 100 mH at 1kHz 0 Ω to 1 GΩ |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability with Calibration and Measurement Capability of - Two terminal capacitors 0.70% from 0.7 μF to 1.2 μF at 100 Hz 0.52% from 1.2 μF to 10.9 μF at 100 Hz 0.92% from 10.9 μF to 33 μF at 100 Hz 0.70% from 33 μF to 100 μF at 100Hz 0.69% from 0.35 nF to 1.5 nF at 1 kHz 1.3% from 1.5 nF to 25 nF at 1 kHz 0.41% from 25 nF to 100 nF at 1 kHz 0.42% from 100 nF to 300 nF at 1 kHz DC Resistance and Inductance 0.06% from 0.1 to 100 mH at 1 kHz 0.013% + 50 μΩ up to 1 Ω |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
DC and low frequency electrical metrology - Electrical standards | Inductors | Inductance | Comparison with a reference standard |
0.1 mH to 100 mH at 1kHz |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability with Calibration and Measurement Capability of - 0.06% from 0.1 to 100 mH at 1kHz |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Precision capacitors | Capacitance | Comparison with a reference standard |
Two terminal capacitors at 1 kHz 10 pF to 1000 nF |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability with Calibration and Measurement Capability of - 0.11% from 10 pF to 100 pF at 1 kHz 0.06% from 101 pF to 1000 nF at 1 kHz |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Pressure metrology - Pressure and vacuum measuring equipment | Digital pressure gauges; Manometers; Mechanical pressure gauges; Pressure recorders; Pressure transducers; Vacuum gauges | Gauge pressure | Comparison with dead weight tester; Comparison with reference instrument |
AS 1349 and MSA Test Method 2 MSA Test Method 1 |
Electrical transducers with dc voltage or dc current output |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability with Calibration and Measurement Capability of- Calibration using a pressure calibrator 0.1 kPa from -100 kPa to 0 kPa 0.4kPa from 200kPa to 2000 kPa Calibration using a Dead Weight Tester 0.02% or 0.05 kPa (whichever is greater) from 300 kPa to 85 MPa |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Pressure metrology - Pressure and vacuum measuring equipmentincluding on site calibrations | Mechanical pressure gauges | Gauge pressure | Comparison with reference instrument |
AS 1349 and MSA Test Method 2 Calibration of industrial gauges |
-80 kPa to 35 MPa | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability with Calibration and Measurement Capability of- 0.2 kPa from -80 kPa to 200 kPa 0.4kPa from 200kPa to 2000 kPa On-site calibration 0.2 kPa from -80 kPa to 200 kPa 1.4 kPa from 200 kPa to 7 MPa 6 kPa from 7 MPa to 35 MPa |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Pressure metrology - Pressure standards | Pressure calibrators - Non-dead weight | Gauge pressure | Comparison with dead weight tester; Comparison with reference instrument | MSA Test Method 1 |
Self indicating and electrical transducers with dc voltage or dc current output |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability with Calibration and Measurement Capability of- Calibration using a pressure calibrator 0.1 kPa from -100 kPa to 0 kPa Calibration using a Dead Weight Tester 0.02% or 0.05 kPa (whichever is greater) from 300 kPa to 85 MPa |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Temperature metrology - Ancillary temperature measuring equipment | Process calibrators | Resistance to temperature conversion | Direct measurement by electrical input |
RTD calibrators (source mode) and RTD indicators (including calibrators in measure mode) from -200 °C to 800 °C |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability with Calibration and Measurement Capability of - RTD calibrators (source mode) 0.06 °C from -200 °C to 0 °C 0.08 °C from 0 °C to 100 °C 0.07 °C from 100 °C to 400 °C 0.08 °C from 400 °C to 630 °C 0.07 °C from 630 °C to 800 °C RTD indicators (including calibrators in measure mode) 0.09 °C from -200 °C to 0 °C 0.11 °C from 0 °C to 100 °C 0.14 °C from 100 °C to 400 °C 0.17 °C from 400 °C to 630 °C 0.31 °C from 630 °C to 800 °C |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Volt to temperature conversion | Direct measurement by electrical input |
Thermocouple calibrators (source mode) Type J Thermocouple from -210 °C to 1 200 °C Type K Thermocouple from -200 °C to 1 372 °C Type N Thermocouple from -200 °C to 1 300 °C Type R Thermocouple from 0 °C to 1 767 °C Type S Thermocouple from 0 °C to 1 767 °C Type T Thermocouple from -250 °C to 400 °C Thermocouple indicators (including calibrators in measure mode) Type J Thermocouple from -210 °C to 1 200 °C Type K Thermocouple from -200 °C to 1 372 °C Type N Thermocouple from -200 °C to 1 300 °C Type R Thermocouple Type S Thermocouple Type T Thermocouple from -250 °C to 400 °C |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability with Calibration and Measurement Capability of - Thermocouple calibrators (source mode) Type J Thermocouple 0.49 °C from -210 °C to -100 °C 0.42 °C from -100 °C to 760 °C 0.46 °C from 760 °C to 1 200 °C Type K Thermocouple 0.54 °C from -200 °C to -100 °C 0.42 °C above -100 °C to 120 °C 0.48 °C above 120 °C to 1 000 °C 0.61 °C above 1 000 °C to 1 372 °C Type N Thermocouple 0.61 °C from -200 °C to -100 °C Type R Thermocouple 0.80 °C from 0 °C to 250 °C Type S Thermocouple 0.70 °C from 0 °C to 250 °C Type T Thermocouple 0.87 °C from -250 °C to -150 °C 0.46 °C from -150 °C to 0 °C 0.41 °C above 0 °C to 400 °C Thermocouple indicators (including calibrators in measure mode)Type J Thermocouple 0.49 °C from -210 °C to -100 °C 0.42 °C from -100 °C to 760 °C 0.47 °C from 760 °C to 1 200 °C Type K Thermocouple 0.55°C from -200 °C to -100 °C 0.43 °C above -100 °C to 120 °C 0.48 °C above 120 °C to 1 000 °C 0.62 °C from 1 000 °C to 1 372 °C Type N Thermocouple 0.61 °C from -200 °C to -100 °C Type R Thermocouple 0.80 °C from 0 °C to 250 °C Type S Thermocouple Type T Thermocouple |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Time and frequency metrology - Frequency and time standards | Frequency standards | Frequency | Measurement against reference standard |
100 mHz to 2 GHz |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability with Calibration and Measurement Capability of - 6.0 in 106 from 100 mHz to 1 kHz 6.0 in 107 from 1 kHz to 50 MHz 2.0 in 107 from 50 MHz to 100 MHz 6.0 in 108 from 100 MHz to 225 MHz 3.0 in 108 from 225 MHz to 1 GHz 6.0 in 109 from 1 GHz to 2 GHz |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Signal sources | Frequency | Measurement against reference standard |
100 mHz to 2 GHz |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability with Calibration and Measurement Capability of - 6.0 in 106 from 100 mHz to 1 kHz 6.0 in 107 from 1 kHz to 50 MHz 2.0 in 107 from 50 MHz to 100 MHz 6.0 in 108 from 100 MHz to 225 MHz 3.0 in 108 from 225 MHz to 1 GHz 6.0 in 109 from 1 GHz to 2 GHz |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Time and frequency metrology - Frequency, time and waveform measuring equipment | Counters | Frequency | Measurement against reference standard |
1 Hz to 2.1 GHz |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability with Calibration and Measurement Capability of - 1.1 in 106 from 1 Hz to 10 Hz 0.6 in 106 from 10 Hz to 2.1 GHz |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Frequency analysers; Waveform measuring instruments | Time interval | Measurement against reference standard |
2 ns to 20 ms 50 ms to 5 s |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability with Calibration and Measurement Capability of - 85 μs/s from 50 ms to 5 s 28 μs/s from 2 ns to 20 ms |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Frequency | Measurement against reference standard |
Frequency response flatness With internal 50 Ω 5 mVp-p to 5 Vp-p and 100 kHz to 1000 MHz With external 50 Ω 5 mVp-p to 5 Vp-p and 100 kHz to 600 MHz |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability with Calibration and Measurement Capability of - With internal 50 Ω 4.5% at 5 mVp-p from 100 kHz to 100 MHz 5.0% at 5 mVp-p from 100 to 300 MHz 7.2% at 5 mVp-p from 300 to 600 MHz 2.8% from 10 mVp-p tp 5.5 Vp-p and 100 kHz to 100 MHz 3.2% from 10 mVp-p to 5.5 Vp-p and 100 to 300 MHz 5.0% from 10 mVp-p to 5.5 Vp-p and 300 to 600 MHz 8.2% from 5 mVp-p to 5.5 Vp-p from 600 to 1000 MHz With external 50 Ω 5.7% at 5 mVp-p from 100 kHz to 100 MHz |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Voltage | Measurement against reference standard |
For DC voltage with 1 MΩ input resistance 5 mV to 100 V For DC voltage with 50 Ω input resistance 5 mV to 5 V For square wave with 1 MΩ input resistance 5 mV to 100 V For square wave voltage with 50 Ω input resistance |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability with Calibration and Measurement Capability of - For DC voltage with 1 MΩ input resistance 0.06% + 40 μV at 5 mV 0.15% from 10 to 50 mV 0.07% from 50 mV to 0.5 V 0.06% from 0.5 V to 100 V For DC voltage with 50 Ω input resistance For square wave with 1 MΩ input resistance For square wave voltage with 50 Ω input resistance |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Frequency meters | Frequency | Measurement against reference standard |
1 Hz to 2.1 GHz |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability with Calibration and Measurement Capability of - 1.1 in 106 from 1 Hz to 10 Hz 0.6 in 106 from 10 Hz to 2.1 GHz |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Tachometers | Rotational speed | Measurement against reference standard |
Optical Tachometers |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability with Calibration and Measurement Capability of - 1.1 rpm from 240 rpm to 999 rpm 1.4 rpm from 1000 rpm to 19997 rpm 2.4 rpm above 19998 rpm to 60000 rpm |
The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.
Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.