Kenelec Scientific Pty Ltd

Site

Particle Counter and Anemometer Calibration Laboratory

Accreditation No.

15814

Site No.

17066

Date of Accreditation

01 Sep 2008

Address

23 Redland Drive
Mitcham, VIC 3132
Australia

kenelec.com.au

Contact

Mr Anthony Busuttil P: +61 (03) 98731022

Availability

Services available to external clients

Scope of Accreditation

ISO/IEC 17025 (2017)

Calibration

  • The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
Service Product Determinant Technique Procedure Limitations
Flow metrology - Environmental flow measuring devices and systems Anemometers Velocity Comparison measurement in wind tunnel Calibration of TSI5725 instruments or equivalent vanes using an open jet wind tunnel.

Capability

with Calibration and Measurement Capability of-

0.050 m/s + 1.10% of indicated reading in the range 0.30 m/s to 30.0 m/s

Optical metrology - Particle measuring devices Particle counters - Air; Particle counters - Liquid Particle count Comparison with a reference standardNULL

ISO 21501-3 ( Determination of Particle Size Distribution - Single particle light interaction methods Part 3. Light extinction  liquid-borne  particle counter)

ISO 21501-4 (Determination of Particle Size Distribution - Single particle light interaction methods Part 4. Light scattering airborne particle counter for clean spaces)


Capability

with  Calibration and Measurement Capability of-

Polystyrene latex (PSL) spheres of nominal sizes 0.30 μm to 30 μm

3.5 % of particle count for liquid borne particle counters

5.9 % of particle count for air borne particle counters

The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.