Pipette Laboratory

Accreditation No.

4283

Site No.

4276

Date of Accreditation

21 Nov 1994

Address

61-63 Victoria Avenue
Chatswood, NSW 2067
Australia

johnmorrisgroup.com

Contact

Miss Trang Diep P: +61 (02) 94964249

Availability

Services available to external clients

Scope of Accreditation

Pipette Laboratory

ISO/IEC 17025 (2017)

Calibration

  • The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
Service Product Determinant Technique Procedure Limitations
Volume and density - Laboratory volumetric glassware and measures Piston operated volumetric apparatus (POVA) Volume Gravimetric

Calibration and verification of piston operated pipetters by the method of ISO 8655-6

From 0.2 μL up to 10060 μL

Capability

With Calibration and Measurement Capability of - 

0.012 μL from 0.2 μL up to 2 μL;

0.017 μL above 2 μL up to 10 μL;

0.025 μL above 10 μL up to 20 μL;

0.12 μL above 20 μL up to 100.8 μL;

0.14 μL above 100.8 μL up to 201 μL;

0.43 μL above 201 μL up to 508 μL;

1.3 μL above 508 μL up to 1008 μL;

2.8 μL above 1008 μL up to 5040 μL;

12 μL above 5040 μL up to 10060 μL.