National Association of Testing Authorities, Australia
Scope of Accreditation
APR Laboratory Pty Ltd
Site
Australian Photometry and Radiometry Laboratory
Accreditation No.
15508
Site No.
16453
Date of Accreditation
25 Jun 2007
Contact
Availability
Services available to external clients
Scope of Accreditation
ISO/IEC 17025 (2017)
Calibration
- The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
Service | Product | Determinant | Technique | Procedure | Limitations | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Optical metrology - Optical measuring equipment | Blue light radiometers; UVA meters; UVC meters; Ultraviolet radiometers | Cosine correction; Irradiance; Linearity; Responsivity | Comparison with a reference meter | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability With Calibration and measurement capability of- Irradiance 6.5% from 10 to 500 μW/cm2 for UVA; Broadband responsivity 1.5% Response linearity 1.5% in the range 0.5 to 10 lux; Cosine correction 1.0% in range 0 to 30 º; |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Illuminance meters; Light meters; Luxmeters; Photometers | Cosine correction; Illuminance; Linearity; Responsivity | Comparison with a reference meter | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability With Calibration and measurement capability of- Illuminance; 0.6% from 0.1 to 5 lux; Broadband responsivity; 1.5% Response linearity; 1.5% in the range 0.5 to 10 lux; Cosine correction; 1.0% in range 0 to 30 º; |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Luminance meters | Luminance | Direct comparison against a reference meter |
from 1 to 12 000 cd/m2 ; |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability With Calibration and measurement capability of- 1.5% from 1 to 4000 cd/m2 ; |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Photodiodes; Photometric detectors | Linearity; Responsivity | Comparison with a reference meter |
Response linearity from 0.5 to 150 000 lux |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability With Calibration and measurement capability of- Broadband responsivity 1.5% Response linearity |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Tricolorimeters | Chromaticity; Luminance | Direct measurement using a reference standard | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability Calibration of incident light tristimulus colorimeters; With Calibration and measurement capability of- 1.5% from 1 to 4000 cd/m2 ; |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Optical metrology - Optical measuring equipmentincluding on site calibrations | Spectroradiometers | Responsivity; Wavelength | Direct comparison against a reference spectrometer; Direct measurement using a reference standard |
from 360 nm to 830 nm; |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability Calibration in the laboratory and on-site; With a Calibration and Measurement Capability of- |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Optical metrology - Optical standards and reference equipment | Diffuse reflectance plates | Reflectance | Direct measurement using a reference meter; Indirect measurement using a spectrophotometer and software integration | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability With Calibration and measurement capability of- Spectral reflectance; 2.5% of measured value or 0.0025 (whichever is greater) from 380 to 410 nm; |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Discharge sources - Lamps; Incandescent lamps; LED sources; Solid state lighting (SSL) sources | Chromaticity | Direct measurement using a reference colorimeter; Direct measurement using a reference spectrophotometer | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability With Calibration and Measurement Capability of- 0.001 in x; |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Correlated colour temperature | Direct measurement using a reference meter | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability With Calibration and Measurement Capability of- for incandescent sources; |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Spectral irradiance | Direct measurement using a reference spectrophotometer |
from 250 nm to 1000 nm; |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability With Calibration and measurement capability of- 5.5% of measured value from 250 to 280 nm; |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Discharge sources - Lamps; LED sources; Solid state lighting (SSL) sources; UVC sources | Broadband irradiance | Direct measurement using a reference meter |
from 50 to 2000 μW/cm2 for UVC sources |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability With Calibration and measurement capability of- 4.0% from 50 to 500 μW/cm2 for UVC sources; |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Filters | Chromaticity; Transmittance | Comparison with a reference meter | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability With Calibration and measurement capability of- Broadband Transmittance; 0.5% of measured value; Spectral Transmittance; 3.0% of measured value in the range 0.200 to 1.000 from 280 to 315 nm; Chromaticity of filters and surfaces: 0.001 in x |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Incandescent lamps; LED sources; Solid state lighting (SSL) sources | Luminous intensity | Direct measurement by goniophotometer; Direct measurement using a reference meter | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability With Calibration and Measurement Capability of- Incandescent lamps; 0.7%; LED and other Solid State Lighting sources; 1.2%; |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Luminous flux | Direct measurement by goniophotometer; Direct measurement by integrating sphere | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability With Calibration and Measurement Capability of- Incandescent lamps; 1.5% above 10 lm; LED and other Solid State Lighting sources; 1.5% above 10 lm; |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Illuminance | Direct measurement using a reference meter |
From 0.5 to 150 000 lux; |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability With Calibration and measurement capability of- 0.7% in the range 0.5 to 3000 lux; |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Integrating sphere sources; Luminance standard sources | Luminance | Direct measurement using a reference meter |
from 1 to 4000 cd/m2 |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability With Calibration and measurement capability of- 2% from 1 to 4000 cd/m2 |