National Association of Testing Authorities, Australia
Scope of Accreditation
Noise and Vibration Measurement Systems Pty Ltd
Scope of Accreditation
ISO/IEC 17025 (2017)
Calibration
- The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
| Service | Product | Determinant | Technique | Procedure | Limitations | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Acoustical metrology - Acoustic equipment | Band pass filters | Attenuation | Direct measurement by electrical input | IEC 61260-3 | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Capability With Calibration and Measurement Capability of- 0.06 dB equal to and less than 16 dB0.08 dB from16 dB to 40 dB 0.12 dB above 40 dB to 60 dB 0.50 dB above 60 dB to 80 dB |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| Acoustical metrology - Acoustic equipment | Signal conditioning amplifiers | Electrical response | Direct measurement by electrical input | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Capability With Calibration and Measurement Capability of -0.03 dB at 1 kHz for Gain from -20 dB to 70 dB (with Low-pass filter, High-pass filter, and A- filter check(s) to manufacturer specification of a measurement uncertainty of 0.1 dB). |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| Acoustical metrology - Acoustic equipment | Sound level meters | Electrical response; Sound pressure level | Direct measurement by electrical input; Direct measurement by reference sound source |
IEC 61672 part 3 AS 1259.1-1990 and AS 1259.2-1990, IEC 60651, IEC 60804 according to OIML R-88Statistical analyser by DIN 45657 |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Capability With Calibration and Measurement Capability of- Electrical tests 0.09 dB Acoustic tests |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| Acoustical metrology - Acoustic equipment | Dose meters for sound | Noise exposure/dose; Sound pressure level | Direct measurement by electrical input; Direct measurement by reference sound source | AS/NZS 2399-1998 (IEC 61252 ) Clauses 6, 7, 8, 9, 10 and 11 | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Capability With Calibration and Measurement Capability of- Electrical tests 0.09 dB 1% Pa2H Acoustic tests 0.12 dB from 31.5 Hz to 1 kHz 0.31 dB from above 1 kHz to 2 kHz 0.51 dB from above 2 kHz to 4 kHz 1.20 dB from above 4 kHz to 8 kHz |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| Acoustical metrology - Acoustic standards and reference equipment | Acoustic calibrators - Multi frequency; Acoustic calibrators - Single frequency | Distortion; Frequency; Sound pressure level | Comparison with a reference standard; Direct measurement by frequency analyser | AS/IEC 60942 - Annex B Periodic tests | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Capability With Calibration and Measurement Capability of- Acoustic0.09 dB at nominal frequencies of 250 and 1000 Hz Distortion 0.23% at 94 dB SPL 0.13% at 114 dB SPL Frequency 0.03 Hz at 94 and 114 dB |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| Acoustical metrology - Acoustic standards and reference equipment | Reference microphones | Frequency response | Direct measurement by electrostatic actuator | IEC 61094-6 | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Capability With Calibration and Measurement Capability of- Frequency response by the method of IEC 61094 part 6 relative to 250 Hz using an electrostatic actuator for 1" and 1/2" microphones 0.12 dB from 20 Hz to 100 Hz reducing to 0 dB at 250 Hz linearly as a function of the logarithm of the frequency from 0 dB at 250 Hz increasing to 0.09 dB at 500 Hz linearly as a function of the logarithm of the frequency 0.09 dB from 500 Hz to 4 kHz increasing to 0.12 dB at 8 kHz linearly as a function of the logarithm of the frequency from 0.12 dB at 8 kHz increasing to 0.24 dB at 31.6 kHz linearly as a function of the logarithm of the frequency from 0.24 dB at 31.6 kHz increasing to 0.30 dB at 40 kHz linearly as a function of the logarithm of the frequency |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| Acoustical metrology - Acoustic standards and reference equipment | Reference microphones | Pressure sensitivity | Pressure field comparison measurement | IEC 61094-5 | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Capability With Calibration and Measurement Capability of- Sensitivity by pressure-field comparison by the method of IEC 61094 part 5 0.07 dB at nominal frequencies of 250 and 1000 Hz |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| Vibration metrology - Vibration equipment | Accelerometers | Acceleration sensitivity | Comparison measurement against reference transducer |
By the methods of ISO 16063 part 21 |
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Capability Level response from 0.1 m/s² for transducers with a sensitivity above 0.1 pC/m/s² and 0.4 mV/m/s² 1 % from 5 Hz to 10 Hz 0.8% at greater than 10 Hz to 2 kHz 1.2% at greater than 2 kHz to 7 kHz 2 % at greater than 7 kHz to 10 kHz |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| Vibration metrology - Vibration equipment | Vibration measuring systems | Acceleration sensitivity; Frequency response | Comparison measurement against reference transducer | By the methods of ISO 16063 part 21 | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Capability With Calibration and Measurement Capability of- Conditioning amplifiers 0.001 V/ms 2 0.8% at greater than 10 Hz to 2 kHz 1.2% at greater than 2 kHz to 7 kHz 2 % at greater than 7 kHz to 10 kHz |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| Vibration metrology - Vibration standards and reference equipment | Vibration calibrators | Acceleration; Distortion; Frequency | Direct measurement by frequency analyser |
By the methods of ISO 16063-21 |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Capability with Calibration and Measurement Capability of - At a nominal frequency of 160 Hz Acceleration0.8% Distortion 0.25% Frequency 0.01 Hz |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
The only data displayed is that deemed relevant and necessary for the clear description of the Activities and Services covered by the Scope of Accreditation (SoA).
Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.
The Product column may also describe an Item or Material.
The Limitations column may also describe the Range.