Simtars

Site

Training, Testing and Certification Centre

Accreditation No.

2681

Site No.

2672

Date of Accreditation

30 Jun 1989

Address

2 Robert Smith Street
Redbank, QLD 4301
Australia

simtars.com.au

Contact

Mr Aaron Heck P: +61 0475970736

Availability

Services available to external clients

Scope of Accreditation

Training, Testing and Certification Centre

ISO/IEC 17025 (2017)

Calibration

  • The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
Service Product Determinant Technique Procedure Limitations
Chemical metrology - Gas analysers Gas analysers Concentration Comparison measurement against a certified reference material; Direct measurement by certified reference material

in house method LP0022

Carbon monoxide : NDIR; 0.0001-10%

Carbon monoxide : Electrochemical; 0.0001-1%

Carbon dioxide : NDIR; 0.001-100%

Methane : NDIR; 0.01-100%

Methane : Pellistor; 0.01-5%

Methane : Methanometer; 0.01-100%

Oxygen : Electrochemical; 0.01-25%

Oxygen : Paramagnetic; 0.01-25%

Hydrogen Sulphide : Electrochemical; 0.1-20000ppm

Sulfur dioxide : Electrochemical; 0.1-20000ppm

Sulfur dioxide : NDIR; 0.00001-50%

Sulfur dioxide : UV fluorescence; 0.1-50ppm

Nitric oxide : Chemiluminescent; 0.1-5000ppm

Nitrogen dioxide : Chemiluminescent; 0.1-5000ppm

Nitrogen dioxide : Electrochemical; 0.1-10000ppm

Oxides of nitrogen : Chemiluminescent; 0.1-10000ppm

Ammonia : Electrochemical; 0.0001-1%

Propane : Pellistor; 0.01-3%

Hexane/Methane/Propane: Pellistor; 0.01-100% LEL

Methane : FID; 0.01-100%

Methane/Hydrogen : Thermal conductivity; 0.01 - 100% LEL

Capability

with least uncertainty of measurement of 5.0 % relative

Flow metrology - Environmental flow measuring devices and systems Air sampling pumps and meters Flow rate Comparison with a reference meter; Volumetric transfer EP0143 - Procedure for the Calibration of Secondary Flow Meters

Capability

Calibration of gas flowmeters using atmospheric air as the test medium with a minimum test volume of 100 mL
With a Calibration and Measurement Capability of -
Bubble flow meter as reference
2% of reading for flow rates from 0.2 L/min to 5 L/min
Mass flow meter as reference
(1% of reading + 0.006 L/min) for flow rates from 0.1 L/min to 15 L/min
Flow rate Comparison with a reference meter EP0140

Capability


Calibration of air sampling pumps using air at atmospheric conditions
with Calibration and Measurement Capability of -
Sampling pump flow rate: 1.2% of flow from 1 L/min to 3 L/min
Sampling pump timer: (0.05% reading + 0.5 minutes) over a minimum of 2 hours

Temperature metrology - Verification of controlled enclosures Baths; Furnaces; Ovens Spatial uniformity; Temperature; Temporal uniformity Direct temperature measurement

AS 2853- 1986

From -40 ºC to 850 ºC
 
 

Capability

Including on site calibrations.

With Calibration and Measurement Capability of -
1.3 ºC  from -40 ºC to 300 ºC
5.4 ºC  above 300 ºC to 850 ºC

Environmental chambers - Humidity Fluctuation; Gradient; Humidity By temperature and humidity measurement AS 60068.3.6; IEC 60068.3.6

Capability

Including on site calibrations.

With Calibration and Measurement Capability of -
Humidity and Relative humidity gradient tests
3.0 %RH in the range 10% to 95%RH  and from 23 °C to 55 °C
5.0 %RH in the range 10% to 93% RH and above 55 °C to 80 °C

Environmental chambers - Temperature Fluctuation; Gradient; Temperature Direct temperature measurement

AS 60068.3.5, IEC 60068.3.5

From -40 ºC to 300 ºC
 
 

Capability

Including on site calibrations.

With Calibration and Measurement Capability of -
1.3 ºC from -40 ºC to 300 ºC

The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.

Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.