Victorian Laboratory

ISO/IEC 17025 (2017)

Calibration

  • The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
Service Product Determinant Technique Procedure Limitations
DC and low frequency electrical metrology - Electrical instrument calibrators Instrument calibrators AC current; AC voltage; DC current; DC voltage; Resistance Direct measurement against a reference standard

Capability

with Calibration and Measurement Capability of -

D.C. voltage

3.9 µV/V + 380 nV from 0  to 200 mV

3.3 µV/V + 820 nV from 200 mV to 2 V

3.5 µV/V + 4. 7µV from 2 V to 20 V

5.3 µV/V + 61 µV from 20 V to 200 V

5.2 µV/V + 740 µV from 200 V to 1000 V

A.C. voltage

From 2.0 mV to 200 mV       

130 µV/V + 11 µV from 10 Hz to 40 Hz

140 µV/V + 5.7µV from 40 Hz to 100 Hz

130 µV/V + 4.8 µV from 100 Hz to 2 kHz

150 µV/V + 5.9 µV from 2 kHz to 10 kHz

320 µV/V + 9 µV from 10 kHz to 30 kHz

870 µV/V + 20 µV from 30 kHz to 100 kHz

From 200 mV to 2 V      

140 µV/V + 35 µV from 10 Hz to 40 Hz

120 µV/V + 36µV from 40 Hz to 100 Hz

120 µV/V + 20 µV from 100 Hz to 2 kHz

150 µV/V + 22 µV from 2 kHz to 10 kHz

230 µV/V + 40 µV from 10 kHz to 30 kHz

530µV/V + 190 µV from 30 kHz to 100 kHz

2.4mV/V + 1.9 mV from 100 kHz to 300 kHz

8.6 mV/V + 19 mV from 300 kHz to 1 MHz

From 2 V to 20 V

96 µV/V + 410 µV from 10 Hz to 40 Hz

81 µV/V + 300 µV from 40 Hz to 100 Hz

67 µV/V + 280 µV from 100 Hz to 2 kHz

110 µV/V + 260 µV from 2 kHz to 10 kHz

200 µV/V + 430 µV from 10 kHz to 30 kHz

510 µV/V + 1.9 mV from 30 kHz to 100 kHz

2.4 mV/V + 19 mV from 100 kHz to 300 kHz

8.5 mV/V + 190 mV from 300 kHz to 1 MHz

From 20 V to 200 V  

94 µV/V + 4.5 mV from 10 Hz to 40 Hz

78 µV/V + 3.7 mV from 40 Hz to 100 Hz

65 µV/V + 3.2 mV from 100 Hz to 2 kHz

98 µV/V + 3.5 mV from 2 kHz to 10 kHz

200 µV/V + 4.8 mV from 10 kHz to 30 kHz

510 µV/V + 19 mV from 30 kHz to 100 kHz

2.4 mV/V + 190 mV from 100 kHz to 300 kHz

From 200 V to 1 kV 

110 µV/V + 31 mV from 10 Hz to 40 Hz

150 µV/V + 23 mV from 40 Hz to 10 kHz


D.C. current

12 µA/A + 0.55 nA from 10 µA to 200 µA

13 µA/A + 4.8 nA from 200 µA to 2 mA

14 µA/A + 44 nA from 2 mA to 20 mA

45 µA/A + 1 µA from 20 mA to 200 mA

180 µA/A + 15 µA from 200 mA to 2 A

390 µA/A + 690 µA from 2 A to 20 A


A.C. current

From 100 μA to 200 µA
480 µA/A + 21 nA from 10 Hz to 10 kHz

From 200 µA to 2 mA       
280 µA/A + 210 nA from 10 Hz to 10 kHz

From 2 mA to 20 mA        
290 µA/A + 2.1 µA from 10 Hz to 10 kHz
From 20 mA to 200 mA 
270 µA/A + 21 µA from 10 Hz to 10 kHz 

From 200 mA to 2 A 

560 µA/A + 210 µA from 10 Hz to 2 kHz

560 µA/A + 500 µA from 2 kHz to 10 kHz

From 2 A to 20 A       

700 µA/A + 2.3 mA from 10 Hz to 2 kHz

1.9 mA/A + 4.3 mA from 2 kHz to 5 kHz


Resistance

2.1 µΩ/Ω + 60 µΩ from 0 Ω up to 2 Ω

7.6 µΩ/Ω + 60 µΩ from 2 Ω to 20 Ω

7.3 µΩ/Ω + 170 µΩ from 20 Ω to 200 Ω

7.5 µΩ/Ω + 1.2 mΩ from 200 Ω to 2000 Ω

7.5 µΩ/Ω + 10 mΩ from 2000 Ω to 20 kΩ

7.3 µΩ/Ω + 160 mΩ from 20 kΩ to 200 kΩ

8.7 µΩ/Ω + 3.1 Ω from 200 kΩ to 2 MΩ

18 µΩ/Ω + 140 Ω from 2 MΩ to 20 MΩ

130 µΩ/Ω + 9.8 kΩ from 20 MΩ to 200 MΩ

1.4 mΩ/Ω + 540 kΩ from 200 MΩ to 1GΩ





DC and low frequency electrical metrology - Electrical measurement and test equipment Ammeters; Digital multimeters (DMM); Ohm meters; Voltmeters AC current; AC voltage; DC current; DC voltage; Resistance Direct measurement against a reference standard

Capability

with Calibration and Measurement Capability of -

D.C. voltage

5.5 µV/V + 720 nV from 25 μV to 22 mV

6.9 µV/V + 690 nV from 22 mV to 220 mV

6.2 µV/V + 980 nV from 220 mV to 2.2 V

6.3 µV/V + 5.5 µV from 2.2 V to 22 V

7 µV/V + 78 µV from 22 V to 220 V

8.7 µV/V + 440 µV from 220 V to 1000 V

A.C. voltage

From 220 μV to 2.2 mV

520 µV/V + 4 µV from 10 Hz to 20 Hz

310 µV/V + 3.9 µV from 20 Hz to 40 Hz

240 µV/V + 3.9 µV from 40 Hz to 20 kHz

64 0µV/V + 3.8 µV from 20 kHz to 50 kHz

920 µV/V + 6.2 µV from 50 kHz to 100 kHz

From 2.2 mV to 22 mV       

470 µV/V + 4.9 µV from 10 Hz to 20 Hz

200 µV/V + 4.9 µV from 20 Hz to 40 Hz

120 µV/V + 4.9 µV from 40 Hz to 20 kHz

340 µV/V + 4.9 µV from 20 kHz to 50 kHz

750 µV/V + 6.6 µV from 50 kHz to 100 kHz

From 22 mV to 220 mV      

480 µV/V + 13 µV from 10 Hz to 20 Hz

210 µV/V + 7.5 µV from 20 Hz to 40 Hz

130 µV/V + 7.2 µV from 40 Hz to 20 kHz

300 µV/V + 7.8 µV from 20 kHz to 50 kHz

710µV/V + 24µV from 50 kHz to 100 kHz

910µV/V + 23µV from 100 kHz to 300 kHz

From 220 mV to 2.2 V

470 µV/V + 78 µV from 10 Hz to 20 Hz

150 µV/V + 24 µV from 20 Hz to 40 Hz

71 µV/V + 5.5 µV from 40 Hz to 20 kHz

120 µV/V + 16 µV from 20 kHz to 50 kHz

220 µV/V + 62 µV from 50 kHz to 100 kHz

390 µV/V + 120 µV from 100 kHz to 300 kHz

990 µV/V + 300 µV from 300 kHz to 500 kHz

2.1 mV/V + 730 µV from 500 kHz to 1 MHz

From 2.2 V to 22 V  

470 µV/V + 770µV from 10 Hz to 20 Hz

150 µV/V + 240 µV from 20 Hz to 40 Hz

71 µV/V + 53 µV from 40 Hz to 20 kHz

120 µV/V + 160 µV from 20 kHz to 50 kHz

220 µV/V + 310 µV from 50 kHz to 100 kHz

470 µV/V + 1.3 mV from 100 kHz to 300 kHz

1.1 mV/V + 3.9 mV from 300 kHz to 500 kHz

2.4 mV/V + 6.9 mV from 500 kHz to 1 MHz

From 22 V to 220 V 

470 µV/V + 7.8 uV/V from 10 Hz to 20 Hz

150 µV/V + 2.4 mV from 20 Hz to 40 Hz

72 µV/V + 830 µV from 40 Hz to 20 kHz

200 µV/V + 3.1 mV from 20 kHz to 50 kHz

470 µV/V + 7.7 mV from 50 kHz to 100 kHz

From 220 V to 1 kV 

77 µV/V + 2.7 mV from 50 Hz to 1 kHz

From 220 V to 330 V          

150 µV/V + 3.9 mV at 45 Hz

From 330 V to 1000 V        

240 µV/V + 7.5 mV at 45 Hz

D.C. current
47 µA/A + 7.7 nA from 10 μA to 220 μA

47 µA/A + 7.7 nA from 220 μA to 2.2 mA

47 µA/A + 77 nA from 2.2 mA to 22 mA

55 µA/A + 770 nA from 22 mA to 220 mA

77µA/A + 23 µA from 220 mA to 2.2 A

390 µA/A + 400 µA from 2.2 A to 11 A

800 µA/A + 370 µA from 11 A to 20 A

A.C. current

From 100 μA to 220 µA      

630 µA/A + 23 nA from 10 Hz to 20 Hz

330 µA/A + 19 nA from 20 Hz to 40 Hz

140 µA/A + 15 nA from 40 Hz to 1 kHz

570 µA/A + 36 nA from 1 kHz to 5 kHz

1.5 mA/A + 74 nA from 5 kHz to 10 kHz

From 220 µA to 2.2 mA       

620 µA/A + 39 nA from 10 Hz to 20 Hz

330 µA/A + 32 nA from 20 Hz to 40 Hz

130 µA/A + 31 nA from 40 Hz to 1 kHz

550 µA/A + 390 nA from 1 kHz to 5 kHz

1.4 mA/A + 770 nA from 5 kHz to 10 kHz

From 2.2 mA to 22 mA        

620 µA/A + 390 nA from 10 Hz to 20 Hz

330 µA/A + 310 nA from 20 Hz to 40 Hz

130 µA/A + 310 nA from 40 Hz to 1 kHz

550 µA/A + 3.9 µA from 1 kHz to 5 kHz

1.4 mA/A + 7.7 µA from 5 kHz to 10 kHz

From 22 mA to 220 mA       

620 µA/A + 4 µA from 10 Hz to 20 Hz

330 µA/A + 3.2 µA from 20 Hz to 40 Hz

150 µA/A + 3 µA from 40 Hz to 1 kHz

550 µA/A + 39 µA from 1 kHz to 5 kHz

1.4 mA/A + 77 µA from 5 kHz to 10 kHz

From 220 mA to 2.2 A        

590µA/A + 30µA from 40 Hz to 1 kHz

690µA/A + 72µA from 1 kHz to 5 kHz

7.9mA/A + 130µA from 5 kHz to 10 kHz

From 2.2 A  to 11 A 

520 µA/A + 1.5 mA from 45 Hz to 100 Hz

810 µA/A + 1.5 mA from 100 Hz to 1 kHz

24 mA/A + 1.6 mA from 1 kHz to 5 kHz

From 11 A to 20 A   

960 µA/A + 3.8 mA from 45 Hz to 100 Hz

1.2 mA/A + 3.8 mA from 100 Hz to 1 kHz

24 mA/A + 3.9 mA from 1 kHz to 5 kHz

Resistance

Within the ranges

33 µΩ/Ω + 780 µΩ 0 Ω to 11 Ω

24 µΩ/Ω + 1.2 mΩ from 11 Ω to 33 Ω

23 µΩ/Ω + 1.1 mΩ from 33 Ω to 110 Ω

24 µΩ/Ω + 1.4 mΩ from 110 Ω to 330 Ω

22 µΩ/Ω + 1.9 mΩ from 330 Ω to 1.1 kΩ

24 µΩ/Ω + 15 mΩ from 1.1 kΩ to 3.3 kΩ

23 µΩ/Ω + 19 mΩ from 3.3 kΩ to 11 kΩ

23 µΩ/Ω + 160 mΩ from 11 kΩ to 33 kΩ

23 µΩ/Ω + 160 mΩ from 33 kΩ to 110 kΩ

26 µΩ/Ω + 1.6 Ω from 110 kΩ to 330 kΩ

26 µΩ/Ω + 1.7 Ω from 330 kΩ to 1.1 MΩ

49 µΩ/Ω + 22 Ω from 1.1 MΩ to 3.3 MΩ

110 µΩ/Ω + 37 Ω from 3.3 MΩ to 11 MΩ

210 µΩ/Ω + 1.9 kΩ from 11 MΩ to 33 MΩ

400 µΩ/Ω + 2.6 kΩ from 33 MΩ t0 110 MΩ

2.4 mΩ/Ω + 73 kΩ from 110 MΩ to 330 MΩ

12 mΩ/Ω + 370 kΩ from 330 MΩ to 1090 MΩ

With fixed values

87 µΩ/Ω at 1 Ω

27 µΩ/Ω at 10 Ω

16 µΩ/Ω at 100 Ω

12 µΩ/Ω at 1 kΩ

11 µΩ/Ω at 10 kΩ

13 µΩ/Ω at 100 kΩ

18 µΩ/Ω at 1 MΩ

38 µΩ/Ω at 10 MΩ

108 µΩ/Ω at 100 MΩ




Dimensional metrology - Engineering equipment and precision instruments Bevel protractors Angle (arc) Comparison with a reference standard

Including compliance with AS B139

Capability

with Calibration and Measurement Capability of -
3.6 minutes of arc from 150 mm to 300 mm

Bore gauges Length measurements Comparison with a reference standard

Capability

with Calibration and Measurement Capability of -
2.5 μm from 2 mm to 50 mm

3.5 μm above 50 mm to 200 mm

Depth and height micrometers Length measurements Comparison with a reference standard

Capability

with Calibration and Measurement Capability of -

2 μm from 1 mm to 125 mm
(1.1+L/180) μm above 125 mm to 600 mm where L is the indicated depth in mm

Dial gauges Length measurements Comparison with a reference standard

Dial gauges and other displacement measuring devices


Capability

with Calibration and Measurement Capability of -

Dial Gauges
(0.8 + 0.032L) μm where L is the range in mm from 1 mm to 100 mm

Digital Indicators
(0.8 + 0.013L) μm where L is the range in mm from 1 mm to 100 mm
Dial Test Indicators

(0.8 + 100R) μm where R is the resolution in mm

Dial gauges; Thickness gauges Length measurements Comparison with a reference standard

Including compliance with AS 2103

Capability

with Calibration and Measurement Capability of -
(0.8 + 0.032L) μm from 1 mm to 100 mm where L is the range in mm

Eccentric mandrels Length measurements Comparison with a reference standard

Capability

with Calibration and Measurement Capability of -
0.21 μm from 1 to 10 μm eccentricity
0.43 μm from 10 to 30 μm eccentricity

External micrometers Length measurements Comparison with a reference standard

Including compliance with AS 2102

Capability

with Calibration and Measurement Capability of -
1.8 μm up to 25 mm
2.6 μm above 25 mm to 100 mm
(1.8+0.01L) μm where L is the length in mm
above 100 mm to 1200 mm

Feeler gauges Length measurements Comparison with a reference standard

Including compliance with 1655

Capability

with Calibration and Measurement Capability of -
(6L + 0.7) μm from 0.04 mm to 1 mm where L is thickness in mm

Internal micrometers Length measurements Comparison with a reference standard

Including compliance with AS 2101

Capability

with Calibration and Measurement Capability of -
2.5 μm from 25 mm to 150 mm
4 μm above 150 mm to 300 mm
(1.5 + 0.008L) μm above 300 mm to 900 mm
where L is the length in mm

Micrometer heads Length measurements Comparison with a reference standard

Including compliance with AS 2328

Capability

with Calibration and Measurement Capability of -
(0.4+0.6R) μm where R is the resolution in μm

Micrometer setting gauges Length measurements Comparison with a reference standard

Up to 1200 mm

Capability

with Calibration and Measurement Capability of -
(1 + 0.007L) μm up to 1200 mm where L is the length in mm

Precision spirit levels Angle (arc) Comparison with a reference standard

Including compliance with AS 2054

Capability

with Calibration and Measurement Capability of -
2 seconds of arc limited to  a sensitivity of 10 seconds of arc

Squares Length measurements Comparison with a reference standard

Including compliance with BS 939

Capability

with Calibration and Measurement Capability of -
2 μm from 100 mm to 200 mm

2.6 μm above 200 mm to 450 mm

4.3 μm above 450 mm to 600 mm

Steel rulers and measuring tapes Length measurements Comparison with a reference standard

Including compliance to AS 1290.4 (only to retractable steel pocket rules)

Capability

with Calibration and Measurement Capability of -
Steel rules
5.4 μm or (3.5+0.007L) μm (whichever is greater) up to 2 m where L is the length in mm
Retractable steel pocket rules
0.47 mm from 0.5 mm to 10 m

Straight edges Length measurements Comparison with a reference standard

Including compliance with AS 1003 and AS B72

Capability

with Calibration and Measurement Capability of -
2 μm from 50 mm to 300 mm

3 μm above 300 to 500 mm

5 μm above 500 to 1000 mm

8 μm above 1000 to 1500 mm

Surface plates Flatness Comparison with a reference standard

Including compliance with AS 1004

Excluding clause 3.4 rigidity

Capability

with Calibration and Measurement Capability of -
(3.6 L + 2) μm where L is length of longest side of the surface/table

Vernier calipers Length measurements Comparison with a reference standard

Including compliance with AS 1984, BS 887 and JIS 7507

Capability

with Calibration and Measurement Capability of -
(10 + 0.025L) μm up to 1.5 m for indication errors where L is the scale value in mm

Vernier height and depth gauges Length measurements Comparison with a reference standard

Including compliance with AS B138

Capability

with Calibration and Measurement Capability of -
Height gauges
 2.5 μm from 10 mm to 300 mm
 (6 + 0.006L) μm from 300 mm to 1000 mm where L is the indicated height in mm
 Depth gauges
 6 μm from 1 to 100 mm
 (5.5 + 0.006L) μm above 100 mm to 1000 mm where L is the indicated depth in mm

Dimensional metrology - Engineering equipment and precision instrumentsincluding on site calibrations Extensometers Length measurements Comparison with a reference standard

in laboratory and in situ

Including compliance to AS 1545

Capability

including on site calibrations

with Calibration and Measurement Capability of -
(0.89 + 0.79L) μm from 0.01 mm to 250 mm where L is the indicated extension in mm

Dimensional metrology - Jigs, fixtures, cutting tools, machine tools, gears, splines and serrations Components and QC standards; Crimp tools; Cutting tools; Jigs and fixtures Length measurements Comparison with a reference standard by differential measurement

Capability

with Calibration and Measurement Capability of -
(1 + 0.007 L) μm up to 1200 mm where L is the length in mm

Dimensional metrology - Length and angle standards Dial gauge calibrators; Extensometer calibrators Length measurements Comparison with a reference standard

Capability

with Calibration and Measurement Capability of -
(0.4+0.6R) μm where R is the resolution in μm

Precision graticules; Precision linear scales Length measurements; Microscope graticules Comparison with a reference standard

Capability

with Calibration and Measurement Capability of -
Precision graticules

1.5 μm with a minimum scale separation of 0.01 mm

Precision linear scales

2 μm up to 500 mm 

Dimensional metrology - Limit gauges and reference standards Indicator gauges; Radius gauges Radius Comparison with a reference standard

Capability

with Calibration and Measurement Capability of -
(8+1.1R) μm where R is the radius in mm

Plain plug gauges; Plain ring gauges Diameter Comparison with a reference standard

Capability

with Calibration and Measurement Capability of -
Plain plug gauges

1 μm from 2 mm to 50 mm
3 μm above 50 mm to 250 mm

Plain ring gauges

1 μm from 2 mm to 50 mm
3 μm above 50 mm to 250 mm

Pressure metrology - Pressure and vacuum measuring equipmentincluding on site calibrations Pressure control devices; Pressure gauges; Pressure recorders; Pressure transducers; Vacuum gauges Gauge pressure Comparison with dead weight tester; Comparison with reference instrument

Including test gauges as defined in AS 1349

-100 kPa to 120 MPa

Capability

with Calibration and Measurement Capability of -
0.05% of reading or 0.05 kPa (whichever is greater) from 100 kPa to 120 MPa
0.23% of reading from - 100 kPa to - 20 kPa and 20 kPa to 100 kPa
0.46% of reading or 80 Pa (whichever is greater) from - 20 kPa to 20 kPa


For on site calibrations
with Calibration and Measurement Capability of -
0.23% of reading from -100 kPa to - 20 kPa and 20 kPa to 120 MPa
0.46% of reading or 80 Pa (whichever is greater) from - 20 kPa to 20 kPa
Torque - Torque measuring and testing equipmentincluding on site calibrations Torque wrenches Torque Comparison with a reference standard

Capability

with Calibration and Measurement Capability of - 

Including onsite calibrations

0.5% of reading In the range from 0.1 Nm and up to 1500 Nm


The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.

Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.