WIKA Australia Pty Ltd

Site

WIKA Australia Pty Ltd

Accreditation No.

410

Site No.

403

Date of Accreditation

28 Feb 1963

Address

Unit K , 10-16 South Street, Rydalmere Business Park
Rydalmere, NSW 2116
Australia

wika.com.au

Contact

Dr Yayang Tan P: +61 (02) 88455222

Availability

Services available to external clients

Scope of Accreditation

WIKA Australia Pty Ltd

ISO/IEC 17025 (2017)

Calibration

  • The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
Service Product Determinant Technique Procedure Limitations
Pressure metrology - Pressure and vacuum measuring equipment Manometers Gauge pressure Comparison with reference instrument

Capability

Mercury, digital and other liquid manometers

with Calibration and Measurement Capability of -

0.1% of reading or 0.8 Pa (whichever is greater) from - 100 kPa to 210 kPa

Pressure metrology - Pressure and vacuum measuring equipmentincluding on site calibrations Digital pressure gauges; Mechanical pressure gauges; Pressure recorders; Pressure transducers; Vacuum gauges Absolute pressure; Gauge pressure Comparison with dead weight tester; Comparison with reference instrument Pressure gauges and vacuum gauges including test gauges to the methods of AS 1349, and MSA Test Method 2;
Pressure transducers - Self indicating and electrical transducers with dc voltage, dc current or resistance outputs by the methods of MSA Test Method 1;
Pressure Calibrators and Pressure recorders by the methods of MSA Test Method 1 or 2

Capability

with Calibration and Measurement Capability of -

Absolute pressure
0.008% of reading or 6 Pa (whichever is greater) from 40 Pa to 120 MPa

Gauge pressure
0.008% of reading or 0.8 Pa (whichever is greater) from -100 kPa to 100 kPa
0.006% of reading above 100 kPa to 5 000 kPa
0.008% of reading above 5 MPa to 20 MPa
0.007% of reading above 20 MPa to 120 MPa
0.08% of reading above 120 MPa to 260 MPa
0.1% of reading above 260 MPa to 350 MPa
On-site calibrations
0.1 % of reading or 0.8 Pa (whichever is greater) from -100 kPa to 120 MPa
0.2 % of reading above 120 MPa to 350 MPa


Pressure metrology - Pressure standards Pressure calibrators - Non-dead weight Absolute pressure; Gauge pressure Comparison with dead weight tester; Comparison with reference instrument

Capability

with Calibration and Measurement Capability of -

Absolute pressure
0.008% of reading or 6 Pa (whichever is greater) from 40 Pa to 120 MPa

Gauge pressure
0.008% of reading or 0.8 Pa (whichever is greater) from -100 kPa to 100 kPa
0.006% of reading above 100 kPa to 5 000 kPa
0.008% of reading above 5 MPa to 20 MPa
0.007% of reading above 20 MPa to 120 MPa
0.08% of reading above 120 MPa to 260 MPa
0.1% of reading above 260 MPa to 350 MPa
0.2% of reading for on-site calibration in the range -100 kPa to 350 MPa


Temperature metrology - Temperature measuring equipment Bimetallic systems; Digital temperature measuring systems; Filled metal systems; Liquid in glass (LIG) thermometers; Metallic resistance thermometers; Temperature loggers - Single and multi-channel; Vapour pressure thermometers Temperature Measurement against reference standard

Capability

with Calibration and Measurement Capability of -

Metallic resistance thermometers

0.05 ºC from -50 ºC to -35 ºC 

0.03 ºC above -35 ºC to 150 ºC

0.05 ºC above 150 ºC to 200 ºC; 

Liquid-in-glass thermometers
0.15 °C from -75 ºC to -35 °C
0.03 °C above -35 ºC to 0 °C
0.01 °C at 0 °C
0.03 °C above 0 ºC to 150 °C
0.08 °C above 150 ºC to 250 °C
0.3 °C above 250 to 400 °C;

Vapour pressure thermometers, Filled metal and Bimetallic systems
1 °C from -50 ºC to -30 °C
0.5 °C above -30 ºC to 90 °C
1 °C above 90 ºC to 200 °C
1.5 °C above 200 ºC to 315 °C
3 °C above 315 ºC to 400 °C;

Digital temperature indicator systems, Temperature Loggers (external sensor)
0.05 °C from -80 ºC to -35 °C
0.03 °C above -35 ºC to 0 °C
0.01 °C at 0°C
0.03 °C above 0 ºC to 150 °C
0.05 °C above 150 ºC to 250 °C
0.2 °C above 250 ºC to 450 °C
Temperature Loggers (internal sensor)
0.35 °C from -25 ºC to 50 °C
RTD and thermistor Thermometers/loggers only- external sensor only for on-site calibration 
0.1 °C from -30 ºC to 0 °C
0.05 °C at 0 °C
0.1 °C above 0 ºC to 200 °C

Temperature metrology - Temperature standards and reference equipment Dry block calibrators Temperature Direct measurement using a reference standard

from -50 ºC to 400ºC

Capability

with Calibration and Measurement Capability of -

0.05 ºC from -50 ºC to 250 ºC

0.2 ºC from 250 ºC to 400 ºC