National Association of Testing Authorities, Australia
Scope of Accreditation
Thermo Fisher Scientific Australia Pty Ltd
Site
National Service Department
Accreditation No.
2145
Site No.
2138
Date of Accreditation
01 Feb 1983
Contact
Availability
Services available to external clients
Scope of Accreditation
ISO/IEC 17025 (2017)
Calibration
- The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
Service | Product | Determinant | Technique | Procedure | Limitations | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
DC and low frequency electrical metrology - Electrical instrument calibrators | Instrument calibrators | AC voltage; DC current; DC voltage; Resistance | Direct measurement against a reference standard | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability with Calibration and Measurement Capability of -DC voltage 0.0013% + 3 μV up to 0.1 V 0.0007% above 0.1 V to 1 V 0.0006% above 1 V to 10 V 0.0009% above 10 V to 100 V 0.0025% above 100 V to 1000 V AC voltage 0.04% + 1.1 μV from 0 V to 0.01 V and from 10 Hz to 1 kHz 0.06% + 1.1 μV from 0 V to 0.01 V and from 1 kHz to 20 kHz 0.15% from 0.01 V to 0.1 V and from 10 Hz to 1 kHz 0.03% from 0.01 V to 0.1 V and from 1 kHz to 20 kHz 0.01% from 0.1 V to 1 V and from 10 Hz to 1 kHz 0.02% from 0.1 V to 1 V and from 1 kHz to 10 kHz 0.034% from 0.1 V to 1 V and from 10 kHz to 50 kHz 0.09% from 0.1 V to 1 V and from 50 kHz to 100 kHz 0.35% from 0.1 V to 1 V and from 100 kHz to 300 kHz 0.01% from 1 V to 10.0 V and from 10 Hz to 1 kHz 0.02% from 1 V to 10.0 V and from 1 kHz to 20 kHz 0.033% from 1 V to 10.0 V and from 20 kHz to 50 kHz 0.09% from 1 V to 10.0 V and from 50 kHz to 100 kHz 0.022% from 10 V to 100.0 V and from 10 Hz to 1 kHz 0.023% from 10 V to 100.0 V and from 1 kHz to 20 kHz 0.045% from 100 V to 1000 V and from 10 Hz to 1 kHz DC current 0.007% + 0.3 nA from 0 μA to 100 μA 0.005% above 100 μA to 1 mA 0.004% above 1 mA to 10 mA 0.007% above 10 mA to 100 mA 0.017% above 100 mA to 1 A Resistance 0.003% + 60 μΩ from 0 Ω to 10 Ω 0.002% above 10 Ω to 100 Ω 0.0013% above 100 Ω to 10 kΩ 0.0015% above 10 kΩ to 100 kΩ 0.003% above 100 kΩ to 1 MΩ 0.007% above 1 MΩ to 10 MΩ | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
DC and low frequency electrical metrology - Electrical measurement and test equipment | Ammeters; Data recorders; Digital multimeters (DMM); Ohm meters; Voltmeters | AC voltage; DC current; DC voltage; Resistance | Comparison with a reference standard | | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability with Calibration and Measurement Capability of -DC voltage 0.0013% + 3μV from 0 V to 0.1 V 0.0013% + 5μV above 0.1 V to 1 V 0.0020% + 100μV above 1 V to 10 V 0.0030% + 2mV above 10 V to 100 V 0.0022% + 2mV above 100 V to 1000 V AC voltage 0.040% + 15 μV from 0 V to 0.01 V (10 Hz to 1 kHz) 0.055% + 15 μV from 0 V to 0.01 V (1 kHz to 20 kHz) 0.10% + 5 μV from 0.01 V to 0.1 V (10 Hz to 1kHz) 0.050% + 25 μV from 0.01 V to 0.1 V (1 kHz to 20 kHz) 0.010% + 300 μV from 0.1 V to 1 V (10 Hz to 1 kHz) 0.015% + 300 μV from 0.1 V to 1 V (1 kHz to 10 kHz) 0.033% + 800 μV from 0.1 V to 1 V (10 kHz to 50 kHz) 0.085% + 3 mV from 0.1 V to 1 V (50 kHz to 100 kHz) 0.55% + 10 mV from 0.1 V to 1 V (100 kHz to 300 kHz) 0.013% + 3 mV from 1 V to 10 V (10 Hz to 1 kHz) 0.015% + 3 mV from 1 V to 10 V (1 kHz to 20 kHz) 0.033% + 8 mV from 1 V to 10 V (20 kHz to 50 kHz) 0.080% + 30 mV from 1 V to 10 V (50 kHz to 100 kHz) 0.022% + 20 mV from 10 V to 100 V (10 Hz to 1 kHz) 0.022% + 35 mV from 10 V to 100 V (1 kHz to 20 kHz) 0.040% + 0.6 V from 100 V to 1000 V (10 Hz to 1 kHz) DC current 0.020% + 0.05 μA from 0 μA to 100 μA 0.020% + 0.05 μA above 100 μA to 1 mA 0.020% + 0.05 μA above 1 mA to 10 mA 0.020% + 0.5 μA above 10 mA to 100 mA 0.030% + 3 μA above 100mA to 1A Resistance 0.0050% + 3 mΩ from 0 Ω to 10 Ω 0.0050% + 70 mΩ above 10 Ω to 100 Ω 0.0050% + 70 mΩ above 100 Ω to 10 kΩ 0.0050% above 10 kΩ to 100 kΩ 0.0050% above 100 kΩ to 1 MΩ 0.019% above 1MΩ to 10 MΩ | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
DC and low frequency electrical metrology - Electrical standards | Conductance boxes; Precision resistors; Resistance boxes | Resistance | Comparison with a reference standard | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability with Calibration and Measurement Capability of -0.003% + 60 μΩ from 0 Ω to 10 Ω 0.002% above 10 Ω to 100 Ω 0.0013% above 100 Ω to 10 kΩ 0.0015% above 10 kΩ to 100 kΩ 0.003% above 100 kΩ to 1 MΩ 0.007% above 1 MΩ to 10 MΩ | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Dimensional metrology - Engineering equipment and precision instrumentsincluding on-site calibrations | Centrifuges | Rotational speed | Comparison with a reference standard by differential measurement | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability including on-site calibrationswith Calibration and Measurement Capability of - 3 rpm from 300 rpm to 15000 rpm | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Mass - Determination of mass and calibration of weighing devicesincluding on site calibrations | Laboratory weighing devices; Precision laboratory balances | Mass | Gravimetric measurement against reference mass | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability Including on site calibrations with Calibration and Measurement Capability of- Precision laboratory balances Industrial balances | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Pressure metrology - Pressure and vacuum measuring equipment | Air data test sets; Altimeters | Absolute pressure; Gauge pressure | Comparison with dead weight tester; Comparison with reference instrument | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability with Calibration and Measurement Capability of -Pneumatic Gauge 0.0065% of reading or 0.003kPa (whichever is greater) from -95kPa to -2kPa: 0.0035% of reading or 0.006 kPa (whichever is greater) above 2kPa to 7Mpa: Pneumatic Absolute 0.015kPa from 10kPa to below 50kPa 0.0035% or 0.006 kPa (whichever is greater) from 50kPa to 7Mpa: | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Barometers; Digital pressure gauges; Manometers; Mechanical pressure gauges; Pressure control devices; Pressure recorders; Pressure transducers; Vacuum gauges | Absolute pressure; Gauge pressure | By pressure calibrator; Comparison with dead weight tester | AS 1349 | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability Self-indicating types and transducers with voltage, current and frequency outputs Digital manometers With Calibration and Measurement Capability of- Pneumatic Gauge pressure 0.0065% of reading or 0.003kPa (whichever is greater) from -95kPa to -2kPa: Pneumatic Absolute pressure 0.015kPa from 10kPa to below 50kPa Hydraulic Gauge pressure 0.007% or 0.08 kPa (whichever is greater) in the range from 500 kPa to 70 MPa Hydraulic Absolute pressure 0.007% or 0.15 kPa (whichever is greater) in the range from 600 kPa to 70 MPa | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Temperature metrology - Ancillary temperature measuring equipment | Digital voltmeters (DVM) and digital multimeters (DMM); Multi-channel thermocouple data recorders; Process calibrators | Resistance to temperature conversion; Volt to temperature conversion | Direct measurement by electrical input | by the methods of Manual E0211 | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability with Calibration and Measurement Capability of -0.1% of range from -200 ºC to 2360 ºC | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Temperature metrology - Temperature measuring equipment | Digital temperature measuring systems | Temperature | Measurement against reference standard | by the methods of Manual E0211 | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability including on-site calibrationswith Calibration and Measurement Capability of - 0.05 oC from -40 oC to 300 oC 0.25 oC above 300 oC to 650 oC On-site calibrations 0.1 ºC from -40 ºC to 300 ºC 0.5 ºC above 300 ºC to 650 ºC | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Temperature metrology - Temperature standards and reference equipment | Dry block calibrators | Temperature | Direct measurement using a reference standard | EA 10/13 | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability with Calibration and Measurement Capability of -0.5 oC In the range from -40 oC to 650 oC | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Temperature metrology - Verification of controlled enclosuresincluding on-site calibrations | Environmental chambers - Temperature; Freezers; Incubators; Ovens | Spatial uniformity; Temperature | Direct temperature measurement | AS2853 | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability including on-site calibrationswith Calibration and Measurement Capability of 1.0 ºC from -80 ºC to 0 ºC 0.5 ºC above 0 ºC to 150 ºC | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Time and frequency metrology - Frequency and time standards | Signal sources | Frequency | Comparison with a reference standard | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability with Calibration and Measurement Capability of0.05% of reading from 1 Hz to 40 Hz 0.01% of reading above 40 Hz to 10 MHz | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Time and frequency metrology - Frequency, time and waveform measuring equipment | Counters; Time interval meters | Frequency | Measurement against reference standard | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability with Calibration and Measurement Capability of -Counters 0.05% of reading from 1 Hz to 40 Hz 0.01% of reading above 40 Hz to 10 MHz Time interval meters 0.01% of reading from 100 ns to 1 s | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Frequency meters | Frequency | Measurement against reference standard | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability with Calibration and Measurement Capability of -0.05% of reading from 1 Hz to 40 Hz 0.01% of reading above 40 Hz to 10 MHz | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Volume and density - Laboratory volumetric glassware and measures | Pipettes; Piston operated volumetric apparatus (POVA) | Volume | Gravimetric | ISO 8655-6 | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability with Calibration and Measurement Capability of -0.070 μL from 0.20 μL to 10 μL 0.080 μL above 10 μL up to 50 μL 0.10 μL above 50 μL up to 100 μL 0.14 μL above 100 μL up to 200 μL 0.20 μL above 200 μL up to 300 μL 0.36 μL above 300 μL up to 500 μL 0.62 μL above 500 μL up to 1 mL |
The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.
Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.
Accreditation No.
2145
Site No.
2138
Print date
22 Nov 2024