Flinders Biomedical Engineering - Flinders Medical Centre

Site

Biomedical Engineering Lab, Imaging and Rehab Support team – – Balance and Pipette Calibration Laboratory

Accreditation No.

13967

Site No.

13960

Date of Accreditation

26 Oct 1998

Address

Flinders Medical Centre, Flinders Drive
Bedford Park, SA 5042
Australia

fbe.org.au

Contact

Mr Noel Kitto P: +61 (08) 82044052

Availability

Services conditionally available to external clients

Scope of Accreditation

ISO/IEC 17025 (2017)

Calibration

  • The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
Service Product Determinant Technique Procedure Limitations
Mass - Determination of mass and calibration of weighing devicesincluding on site calibrations Precision laboratory balances Mass Gravimetric measurement against reference mass

Capability

Including on site calibrations

with Calibration and Measurement Capability of-

12 μg from 1 mg to 2 g
35 μg above 2 g to 20 g
2 in 10⁶ above 20 g to 2 kg

Volume and density - Laboratory volumetric glassware and measures Piston operated volumetric apparatus (POVA) Volume Gravimetric method

Calibration of piston operated burettes

by the method of ISO8655-6

Capability

with Calibration and Measurement Capability of -
0.03 μL from 0.5 μL to 5 μL

0.04 μL at 10 μL and 20 μL

0.10 μL at 50 μL and 100 μL

0.27 μL at 200 μL

0.7 μL at 500 μL and 1 mL

2.7 μL at 2 mL and 5 mL

6.7 μL at 10 mL

14 μL at 20 mL

27 μL at 50 mL

67 μL at 100 mL

Volume Gravimetric method

Calibration of piston operated pipettes;

By the method of  ISO8655-6

Capability

with Calibration and Measurement Capability of -
0.03 μL from 0.5 μL to 5 μL;

0.04 μL at  10 μL and 20 μL;

0.10 μL at 50 μL and 100 μL;

0.27 μL at 200 μL;

0.7 μL at 500 μL and 1 mL;

2.7 μL at 2 mL and 5 mL;

6.7 μL at 10 mL;

14 μL at 20 mL

The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.