![NATA](https://nata.com.au/files/2021/02/nata.png)
National Association of Testing Authorities, Australia
Scope of Accreditation
EMC Technologies Pty Ltd
Site
Melbourne Laboratory – EMC Test Laboratory
Accreditation No.
5292
Site No.
21206
Date of Accreditation
22 Sep 2008
Contact
Availability
Services available to external clients
Scope of Accreditation
Withdrawn 24 May 2024
Voluntary
ISO/IEC 17025 (2017)
- The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
Calibration
|
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Service | Product | Determinant | Technique | Procedure | Limitations | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
High frequency electrical metrology - Communications, electromagnetic field strength and EMC test equipment | Active rod antennas; Biconical antennas; Horn antennas; Log periodic antennas | Antenna factor; Frequency | Direct measurement by electrical input | SAE ARP 958, version D | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability Narrowband transducers and indicators with Calibration and Measurement Capability of- Antenna Factor from -5 dB(1/m) up to 50 dB(1/m)± 1.0 dB from 9kHz to 40 GHz |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Artificial mains networks (AMN); Line impedance stabilisation networks (LISN) | Voltage division factor | Direct measurement by electrical input | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability with Calibration and Measurement Capability of - in the frequency range 9 kHz to 108 MHz± 0.6 dB for voltage division factor ± 2.1% for voltage drop ± 10.9% for impedance± 3.8º for phase |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Coupling/decoupling networks (CDN) | Insertion loss | Direct measurement by electrical input | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability with Calibration and Measurement Capability of - in the frequency range 150 kHz to 230 MHz± 10.9% for impedance |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
E&H field probes and meters; Electromagnetic field hazard meters and warning indicators | Frequency | Comparison with a reference standard | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability with Calibration and Measurement Capability of-
|
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Electrical fast transients (EFT) and surge generators | Pulse characteristics | Direct measurement by electrical input | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability Electrical fast transients (EFT) with Calibration and Measurement Capability of - in the range ± 0.5 kV to 20 kV± 5.80% for peak voltage ± 4.2% for pulse width± 1.2% for rise time ± 3.07% for repetition rate |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Pulse characteristics | Direct measurement by electrical input | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability Surge Generators with Calibration and Measurement Capability of - in the range ± 0.1 kV to 7 kV± 5.60% for open circuit voltage ± 10.53% for open circuit front time± 4.4% for open circuit duration ± 4.4% for short circuit current ± 4.29% for short circuit front time ± 4.3% for short circuit duration |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Electrostatic discharge (ESD) generators | Pulse characteristics | Direct measurement by electrical input | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability with Calibration and Measurement Capability of - in the range ± 2 kV to ± 30 kV± 14.2% for rise time ± 8.10% for peak current± 1.2% for peak voltage |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Harmonic emission compliance test systems | Harmonic currents | Comparison with a reference standard by differential measurement | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability with Calibration and Measurement Capability of - 0.15% of the fundamental of current for Harmonic currents - from 0.7 A to 16 A total rms per phase - from 1st to 40th harmonic with fundamental frequency between 50 to 60 Hz and voltage between 120 to 230 V |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Line impedance stabilisation networks (LISN) | Voltage division factor | Direct measurement by electrical input | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability Balanced Telecom ISN (TLISN) with Calibration and Measurement Capability of - in the frequency range 150 kHz to 30 MHz± 4.3 Ω for common mode impedance ± 2.4º for common mode phase± 0.97 dB for voltage division factor ± 1.12 dB for isolation / DCMD ± 2.28 dB for Longitudinal conversion loss (LCL) |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Voltage fluctuation and flicker compliance test systems | Internal reference impedance resistance; Short term flicker; Voltage deviation | Comparison with a reference standard by substitution measurement | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability with Calibration and Measurement Capability of - 0.5% of Pst in the range from 0.5 Ps to 1.5 Ps for Short term flicker perceptibility 0.07% of fundamental of the voltage in the range from 0.5 to 2.0% for Voltage deviation, Dmax, with fundamental frequency between 50 Hz to 60 Hz and voltage between 1 V to 230 V 1.8 MΩ for Internal reference impedance resistance |
The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.