National Association of Testing Authorities, Australia
Scope of Accreditation
PCS Measurement Pty Limited
Site
PCS Measurement Pty Limited
Accreditation No.
2143
Site No.
2136
Date of Accreditation
18 Dec 1986
Contact
Availability
Services available to external clients
Scope of Accreditation
ISO/IEC 17025 (2017)
Calibration
- The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
Service | Product | Determinant | Technique | Procedure | Limitations | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Dimensional metrology - Engineering equipment and precision instruments | Depth and height micrometers | Length measurements | Direct measurement against a reference standard | BS 6468 | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability With calibration and measurement capability of - (2 + L/250) μm where L is the indicated depth in mm from 0 to 300 mm | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Electronic calipers; Vernier calipers | Length measurements | Direct measurement against a reference standard | AS 1984 JIS B 7507 | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability With calibration and measurement capability of - 13 μm for linear measurements from 2.5 mm up to 150 mm; 18 µm in the range 0 mm and up to 1000 mm in accordance with JIS B 7507 | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Extensometers | Length measurements | Direct measurement | AS 1545 | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability With calibration and measurement capability of - (1.5 + 0.5 L) μm where L is the extension in mm 0 to 25 mm | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
External micrometers | Length measurements | Direct measurement against a reference standard | AS 2102 | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability With calibration and measurement capability of - 1.4 μm for linear measurements up to 25 mm; | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Feeler gauges | Length measurements | Direct measurement | AS1655 | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability With calibration and measurement capability of - 3 μm up to 5 mm | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Internal micrometers | Length measurements | Direct measurement against a reference standard | AS 2101 | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability With calibration and measurement capability of - 2 μm from 12.5 to 100 mm; | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Micrometer setting gauges | Length measurements | Direct measurement against a reference standard | BS870 | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability With calibration and measurement capability of - 2 μm from 0.5 to 100 mm | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Straight edges | Length measurements | Direct measurement | AS 1003 | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability With calibration and measurement capability of - 4.5 μm from 50 mm to 1200 mm | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Dimensional metrology - Engineering equipment and precision instrumentsIncluding on site calibrations | Dial gauges; Electronic length gauges | Length measurements | Direct measurement | AS 2103-1978 In house developed method for onsite calibration Work instruction 94 | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability With calibration and measurement capability of - 1 μm up to 100 mm by AS 2103 For on-site calibrations1.7 µm in the range 0 mm and up to 50 mm using in house method work instruction 94 | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Dimensional metrology - Limit gauges and reference standards | Parallel screw plug gauges | Simple pitch diameter only | Comparison with a reference standard by differential measurement | BS 919, AS 3635, AS 1014 and AS 1721 | With the exception of setting plugs and check plugs Screw plug limits and tolerances includes compliance with BS 919-1/07 From 2 mm to 64 mm | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability with calibration and measurement capability of - 7 μm from 2 mm to 64 mm | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Plain plug gauges; Plain ring gauges | Diameter | Direct measurement; Direct measurement against a reference standard | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability Plain plug gauges with calibration and measurement capability of - 3 μm from 2 mm to 50 mm Plain ring gauges with calibration and measurement capability of - 4 μm from 2 mm to 100 mm | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Force metrology - Force measuring and testing equipmentIncluding on-site calibrations | Compression and universal machines in compression; Elastic force measuring devices; Load cells; Load measuring rings | Force in compression | Comparison measurement with reference load cell | AS 2193 | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability Including on-site calibrations With calibration and measurement capability of - 0.1% of reading from 1 N to 2 MN to Class A requirements of AS 2193 | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Load cells; Tension and universal machines in tension | Force in tension | Comparison measurement with reference load cell | AS 2193 | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability Including on-site calibrations with calibration and measurement capability of - 0.1% of reading from 1 N to 500 kN to Class A requirements of AS 2193 | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Force metrology - Force standards | Force standards and load cells used as force standards | Force in compression; Force in tension | Comparison measurement with reference load cell; Gravimetric measurement with reference mass | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability With calibration and measurement capability of - 0.02 % of reading from 1 N to 50 kN in tension and compression to Class 1 and Class 2 requirements of AS 2193 | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Force metrology - Hardness standards and equipment | Rubber bases; Rubber hardness meters (durometers) | Rubber hardness | Comparison with a reference standard; Direct force and length measurement | AS 1683.15.2 and ASTM D2240 | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability With a Calibration and Measurement Capability of - Shore A and D durometers for compliance with AS 1683.15.2 and ASTM D2240 0.2% of reading Testing of rubber bases for liquid limit devices for compliance with AS 1289.3.1.1 0.5% for rubber hardness 0.6% for dimensional measurements | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Force metrology - Hardness standards and equipmentIncluding on-site calibrations | Brinell hardness machines; Portable Brinell measuring microscopes | Brinell hardness | Comparison with a reference standard; Direct force and length measurement | AS 1816 ASTM E110 | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability Including on-site calibrations Direct and Indirect verification to AS 1816 From 39.2 N to 29.4 With calibration and measurement capability of - | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Rockwell hardness machines; Rockwell superficial hardness machines | Rockwell hardness | Comparison with a reference standard; Direct force and length measurement | AS 1815 | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability Including on-site calibrationsDirect and Indirect verification excluding indenter dimensions and measuring apparatus to AS 1815 With calibration and measurement capability of - 0.1% of applied load for verification of test force 0.4 Rockwell units in ranges as defined in AS1815 | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Vickers hardness machines; Vickers micro-hardness machines | Vickers hardness | Comparison with a reference standard; Direct force and length measurement | AS 1817 | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability Direct and Indirect verification to AS 1817 excluding indenter dimensions Including on-site calibrations from 24.5 N to 980 N With calibration and measurement capability of -0.032% for verification of length measuring devices 0.1% of applied load for verification of test force 0.5% of Vickers units in ranges as defined in AS1817 | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Mass - Determination of mass and calibration of weighing devices | Industrial mass standards; Mass standards | Mass | Direct measurement; Gravimetric measurement against reference mass | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability with calibration and measurement capability of - Mass standards by gravimetric measurement against a reference mass 30 μg from 10 mg up to 2 g 2 in 106 or 30 μg (whichever is greater) above 2 g to 200 g; Industrial mass standards by direct measurement using a reference load cell | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Mass - Determination of mass and calibration of weighing devicesincluding on site calibrations | Industrial weighing devices; Laboratory weighing devices; Precision laboratory balances | Mass | Gravimetric measurement against reference mass | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability Including on site calibrations With Calibration and Measurement Capability of- Precision laboratory balances 2 in 106 or 20 μg (whichever is greater) from 10 mg to 200 g Industrial balances 5 in 106 from 200 g to 5 kg; Industrial weighing appliances 1 in 104 from 30 kg to 2 t | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Temperature metrology - Temperature measuring equipmentIncluding on-site calibrations | Digital temperature measuring systems | Temperature | Measurement against reference standard | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability Including on-site calibrations With calibration and measurement capability of - 1.1 ºC from -20 ºC to 100 ºC 1.4 ºC from 100 ºC to 300 ºC 2.1 ºC from 300 ºC to 600 ºC | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Temperature metrology - Verification of controlled enclosuresincluding on-site calibrations | Baths; Furnaces; Ovens | Temperature | Direct temperature measurement | AS 2853-1986 | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability Including on-site calibrations With calibration and measurement capability of - 1.2 ºC from 0 ºC to 140 ºC 3.8 ºC from 140 to 1000 ºC | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Time and frequency metrology - Frequency, time and waveform measuring equipment | Tachometers | Rotational speed | Direct measurement | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability With calibration and measurement capability of - 1.5 rpm from 100 to 10000 rpm | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Torque - Torque measuring and testing equipment | Torque gauges; Torque transducers; Torque wrenches | Torque | Comparison with a reference standard; Gravimetric measurement with reference beam | Torque wrenches, clockwise and anti-clockwise at torques Calibration of torque wrenches to ISO 6789-2 | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability With calibration and measurement capability of - Torque wrenches 0.34% from 1 N.m to 22 N.m 0.12% above 22 N.m to 678 N.m 0.77% above 678 N.m to 1500 N.m Torque transducers and Analogue torque meters 0.4 % from 1 N.m to 4067 N.m |
The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.
Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.
Accreditation No.
2143
Site No.
2136
Print date
08 Dec 2024